Slotted surface coil with reduced g-factor for SENSE imaging.

K Ocegueda1, A O Rodriguez.   

Abstract

A new coil design inspired on the slot-and-hole magnetron tube is proposed for SENSE imaging. To investigate its g-factor behaviour: an SNR formula was derived using the quasi-static approach, and combined with the ultimate g-factor formula to compute the ultimate-factor-g-vs-depth plots. A g-factor expression was derived for the circular coil using the same approach for comparison purposes. SNR-vs-depth profiles of an 4-slot coil showed an important improvement over the circular coil. The 4-slot coil g-factor can be up to 58.32% lower than that of a single circular-shaped coil. This improvement makes the slotted surface coil a good choice for SENSE imaging.

Mesh:

Year:  2006        PMID: 17946922     DOI: 10.1109/IEMBS.2006.259736

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X