Preliminary investigations on the determination of three-dimensional dose distributions using scintillator blocks and optical tomography.

Florian Kroll1, Jörg Pawelke, Leonhard Karsch.   

Abstract

PURPOSE: Clinical QA in teletherapy as well as the characterization of experimental radiation sources for future medical applications requires effective methods for measuring three-dimensional (3D) dose distributions generated in a water-equivalent medium. Current dosimeters based on ionization chambers, diodes, thermoluminescence detectors, radiochromic films, or polymer gels exhibit various drawbacks: High quality 3D dose determination is either very sophisticated and expensive or requires high amounts of effort and time for the preparation or read out. New detectors based on scintillator blocks in combination with optical tomography are studied, since they have the potential to facilitate the desired cost-effective, transportable, and long-term stable dosimetry system that is able to determine 3D dose distributions with high spatial resolution in a short time.
METHODS: A portable detector prototype was set up based on a plastic scintillator block and four digital cameras. During irradiation the scintillator emits light, which is detected by the fixed cameras. The light distribution is then reconstructed by optical tomography, using maximum-likelihood expectation maximization. The result of the reconstruction approximates the 3D dose distribution. First performance tests of the prototype using laser light were carried out. Irradiation experiments were performed with ionizing radiation, i.e., bremsstrahlung (6 to 21 MV), electrons (6 to 21 MeV), and protons (68 MeV), provided by clinical and research accelerators.
RESULTS: Laser experiments show that the current imaging properties differ from the design specifications: The imaging scale of the optical systems is position dependent, ranging from 0.185 mm/pixel to 0.225 mm/pixel. Nevertheless, the developed dosimetry method is proven to be functional for electron and proton beams. Induced radiation doses of 50 mGy or more made 3D dose reconstructions possible. Taking the imaging properties into account, determined dose profiles are in agreement with reference measurements. An inherent drawback of the scintillator is the nonlinear light output for high stopping-power radiation due to the quenching effect. It impacts the depth dose curves measured with the dosimeter. For single Bragg peak distributions this leads to a peak to plateau ratio of 2.8 instead of 4.5 for the reference ionization chamber measurement. Furthermore, the transmission of the clinical bremsstrahlung beams through the scintillator leads to the saturation of one camera, making dose reconstructions in that case presently not feasible.
CONCLUSIONS: It is shown that distributions of scintillation light generated by proton or electron beams can be reconstructed by the dosimetry system within minutes. The quenching apparent for proton irradiation, and the yet not precisely determined position dependency of the imaging scale, require further investigation and corrections. Upgrading the prototype with larger or inorganic scintillators would increase the detectable proton and electron energy range. The presented results show that the determination of 3D dose distributions using scintillator blocks and optical tomography is a promising dosimetry method.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23927341     DOI: 10.1118/1.4813898

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Optical cone beam tomography of Cherenkov-mediated signals for fast 3D dosimetry of x-ray photon beams in water.

Authors:  Adam K Glaser; Jacqueline M Andreozzi; Rongxiao Zhang; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

2.  Novel, full 3D scintillation dosimetry using a static plenoptic camera.

Authors:  Mathieu Goulet; Madison Rilling; Luc Gingras; Sam Beddar; Luc Beaulieu; Louis Archambault
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

3.  Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems.

Authors:  Chinmay D Darne; Fahed Alsanea; Daniel G Robertson; Narayan Sahoo; Sam Beddar
Journal:  Phys Med Biol       Date:  2017-06-23       Impact factor: 3.609

Review 4.  Towards effective and efficient patient-specific quality assurance for spot scanning proton therapy.

Authors:  X Ronald Zhu; Yupeng Li; Dennis Mackin; Heng Li; Falk Poenisch; Andrew K Lee; Anita Mahajan; Steven J Frank; Michael T Gillin; Narayan Sahoo; Xiaodong Zhang
Journal:  Cancers (Basel)       Date:  2015-04-10       Impact factor: 6.639

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.