Comparative power law analysis of structured breast phantom and patient images in digital mammography and breast tomosynthesis.

L Cockmartin1, H Bosmans, N W Marshall.   

Abstract

PURPOSE: This work characterizes three candidate mammography phantoms with structured background in terms of power law analysis in the low frequency region of the power spectrum for 2D (planar) mammography and digital breast tomosynthesis (DBT).
METHODS: The study was performed using three phantoms (spheres in water, Voxmam, and BR3D CIRS phantoms) on two DBT systems from two different vendors (Siemens Inspiration and Hologic Selenia Dimensions). Power spectra (PS) were calculated for planar projection, DBT projection, and reconstructed images and curve fitted in the low frequency region from 0.2 to 0.7 mm(-1) with a power law function characterized by an exponent β and magnitude κ. The influence of acquisition dose and tube voltage on the power law parameters was first explored. Then power law parameters were calculated from images acquired with the same anode∕filter combination and tube voltage for the three test objects, and compared with each other. Finally, PS curves for automatic exposure controlled acquisitions (anode∕filter combination and tube voltages selected by the systems based on the breast equivalent thickness of the test objects) were compared against PS analysis performed on patient data (for Siemens 80 and for Hologic 48 mammograms and DBT series). Dosimetric aspects of the three test objects were also examined.
RESULTS: The power law exponent (β) was found to be independent of acquisition dose for planar mammography but varied more for DBT projections of the sphere-phantom. Systematic increase of tube voltage did not affect β but decreased κ, both in planar and DBT projection phantom images. Power spectra of the BR3D phantom were closer to those of the patients than these of the Voxmam phantom; the Voxmam phantom gave high values of κ compared to the other phantoms and the patient series. The magnitude of the PS curves of the BR3D phantom was within the patient range but β was lower than the average patient value. Finally, PS magnitude for the sphere-phantom coincided with the patient curves for Siemens but was lower for the Hologic system. Close agreement of doses for all three phantoms with patient doses was found.
CONCLUSIONS: Power law parameters of the phantoms were close to those of the patients but no single phantom matched in terms of both magnitude (κ) and texture (β) for the x-ray systems in this work. PS analysis of structured phantoms is feasible and this methodology can be used to suggest improvements in phantom design.

Entities:  

Mesh:

Year:  2013        PMID: 23927334     DOI: 10.1118/1.4816309

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  A Statistical Model for Rigid Image Registration Performance: The Influence of Soft-Tissue Deformation as a Confounding Noise Source.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2019-03-27       Impact factor: 10.048

2.  DukeSim: A Realistic, Rapid, and Scanner-Specific Simulation Framework in Computed Tomography.

Authors:  Ehsan Abadi; Brian Harrawood; Shobhit Sharma; Anuj Kapadia; William P Segars; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2018-12-12       Impact factor: 10.048

3.  Quantification of breast lesion compositions using low-dose spectral mammography: A feasibility study.

Authors:  Huanjun Ding; David Sennung; Hyo-Min Cho; Sabee Molloi
Journal:  Med Phys       Date:  2016-10       Impact factor: 4.071

4.  Three-dimensionally-printed anthropomorphic physical phantom for mammography and digital breast tomosynthesis with custom materials, lesions, and uniform quality control region.

Authors:  Andrea H Rossman; Matthew Catenacci; Christine Zhao; Dhiraj Sikaria; John E Knudsen; Danielle Dawes; Michael E Gehm; Ehsan Samei; Benjamin J Wiley; Joseph Y Lo
Journal:  J Med Imaging (Bellingham)       Date:  2019-03-22

5.  Quantitative contrast-enhanced spectral mammography based on photon-counting detectors: A feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

6.  Synthetic breast phantoms from patient based eigenbreasts.

Authors:  Gregory M Sturgeon; Subok Park; William Paul Segars; Joseph Y Lo
Journal:  Med Phys       Date:  2017-10-19       Impact factor: 4.071

7.  Human observer performance on in-plane digital breast tomosynthesis images: Effects of reconstruction filters and data acquisition angles on signal detection.

Authors:  Changwoo Lee; Minah Han; Jongduk Baek
Journal:  PLoS One       Date:  2020-03-12       Impact factor: 3.240

8.  Effect of optical blurring of X-ray source on breast tomosynthesis image quality: Modulation transfer function, anatomical noise power spectrum, and signal detectability perspectives.

Authors:  Changwoo Lee; Jongduk Baek
Journal:  PLoS One       Date:  2022-05-19       Impact factor: 3.240

9.  Computational Breast Anatomy Simulation Using Multi-Scale Perlin Noise.

Authors:  Bruno Barufaldi; Craig K Abbey; Miguel A Lago; Trevor L Vent; Raymond J Acciavatti; Predrag R Bakic; Andrew D A Maidment
Journal:  IEEE Trans Med Imaging       Date:  2021-11-30       Impact factor: 10.048

10.  Evaluation of Digital Breast Tomosynthesis as Replacement of Full-Field Digital Mammography Using an In Silico Imaging Trial.

Authors:  Aldo Badano; Christian G Graff; Andreu Badal; Diksha Sharma; Rongping Zeng; Frank W Samuelson; Stephen J Glick; Kyle J Myers
Journal:  JAMA Netw Open       Date:  2018-11-02
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.