Literature DB >> 23925712

Using DynaCT for the assessment of ilio-femoral arterial calibre, calcification and tortuosity index in patients selected for trans-catheter aortic valve replacement.

James A Crowhurst1, Douglas Campbell, Owen C Raffel, Mark Whitby, Pavthrun Pathmanathan, Stanley Redmond, Alexander Incani, Karl Poon, Christopher James, Constantine Aroney, Andrew Clarke, Darren L Walters.   

Abstract

Adequate vascular access for femoral trans-catheter aortic valve replacement is fundamental to the success of the procedure. Assessment of vascular calibre, tortuosity and calcification is performed by angiography and multi-slice computed tomography (MSCT). Can DynaCT provide the same information as MSCT? 15 Patients underwent MSCT, angiography and DynaCT. Vessel diameter measurements were taken in three positions of the left and right ilio-femoral arteries. Tortuosity was assessed using an index of the direct distance and the distance taken by the artery between two points. Calcification was assessed in MSCT and DynaCT using a simple scoring system. Concordance correlation coefficient of arterial calibre between angiography and MSCT was 0.96 (95 % CI 0.94-0.97). DynaCT and angiography was 0.94 (95 % CI 0.91-0.96) and Dyna CT and MSCT, 0.95 (95 % CI 0.92-0.97). Bland-Altman tests demonstrate a mean difference between the angiogram and the MSCT of 0.06 mm (+0.97, -1.42), angiogram and DynaCT, 0.13 mm, (+1.00, -0.87), DynaCT and MSCT, 0.2 mm, (+1.15, -0.76). Tortuosity comparisons gave a median tortuosity index for MSCT 1.29 and DynaCT 1.23 (p = 0.472). Calcification comparisons of MSCT and DynaCT using correlation coefficients demonstrate a correlation of 0.245 (p = 0.378). Effective radiation doses were: DynaCT; 3.63 ± 0.65 mSv and angiography; 0.57 ± 0.72 mSv, MSCT; 7.15 ± 2.58 mSv. DynaCT is equal to MSCT and angiography in assessing femoral artery calibre. Like MSCT, it can assess tortuosity and can produce 3D images but is inferior in the assessment of calcification.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23925712     DOI: 10.1007/s10554-013-0221-y

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  17 in total

1.  Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery.

Authors:  Martin B Leon; Craig R Smith; Michael Mack; D Craig Miller; Jeffrey W Moses; Lars G Svensson; E Murat Tuzcu; John G Webb; Gregory P Fontana; Raj R Makkar; David L Brown; Peter C Block; Robert A Guyton; Augusto D Pichard; Joseph E Bavaria; Howard C Herrmann; Pamela S Douglas; John L Petersen; Jodi J Akin; William N Anderson; Duolao Wang; Stuart Pocock
Journal:  N Engl J Med       Date:  2010-09-22       Impact factor: 91.245

2.  Location of femoral artery puncture site and the risk of postcatheterization pseudoaneurysm formation.

Authors:  Marcin Gabriel; Katarzyna Pawlaczyk; Krzysztof Waliszewski; Zbigniew Krasiński; Wacław Majewski
Journal:  Int J Cardiol       Date:  2007-01-23       Impact factor: 4.164

3.  Transcatheter versus surgical aortic-valve replacement in high-risk patients.

Authors:  Craig R Smith; Martin B Leon; Michael J Mack; D Craig Miller; Jeffrey W Moses; Lars G Svensson; E Murat Tuzcu; John G Webb; Gregory P Fontana; Raj R Makkar; Mathew Williams; Todd Dewey; Samir Kapadia; Vasilis Babaliaros; Vinod H Thourani; Paul Corso; Augusto D Pichard; Joseph E Bavaria; Howard C Herrmann; Jodi J Akin; William N Anderson; Duolao Wang; Stuart J Pocock
Journal:  N Engl J Med       Date:  2011-06-05       Impact factor: 91.245

4.  Vessel tortuosity and brain tumor malignancy: a blinded study.

Authors:  Elizabeth Bullitt; Donglin Zeng; Guido Gerig; Stephen Aylward; Sarang Joshi; J Keith Smith; Weili Lin; Matthew G Ewend
Journal:  Acad Radiol       Date:  2005-10       Impact factor: 3.173

5.  Validation of DynaCT in the morphological assessment of abdominal aortic aneurysm for endovascular repair.

Authors:  Ian M Nordon; Robert J Hinchliffe; Amir H Malkawi; Jeremy Taylor; Peter J Holt; Robert Morgan; Ian M Loftus; Matt M Thompson
Journal:  J Endovasc Ther       Date:  2010-04       Impact factor: 3.487

6.  Thirty-day results of the SAPIEN aortic Bioprosthesis European Outcome (SOURCE) Registry: A European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve.

Authors:  Martyn Thomas; Gerhard Schymik; Thomas Walther; Dominique Himbert; Thierry Lefèvre; Hendrik Treede; Holger Eggebrecht; Paolo Rubino; Iassen Michev; Rüdiger Lange; William N Anderson; Olaf Wendler
Journal:  Circulation       Date:  2010-06-21       Impact factor: 29.690

7.  Impact of chronic kidney disease on prognosis of patients with diabetes mellitus treated with percutaneous coronary intervention.

Authors:  Eugenia Nikolsky; Roxana Mehran; Diane Turcot; Eve D Aymong; Gary S Mintz; Zoran Lasic; Alexandra J Lansky; Emmanouil Tsounias; Jeffrey W Moses; Gregg W Stone; Martin B Leon; George D Dangas
Journal:  Am J Cardiol       Date:  2004-08-01       Impact factor: 2.778

8.  Managing patient dose in multi-detector computed tomography(MDCT). ICRP Publication 102.

Authors:  J Valentin
Journal:  Ann ICRP       Date:  2007

9.  Access site complications following cardiac catheterization assessed by duplex ultrasonography.

Authors:  Ljiljana Banfić; Majda Vrkić Kirhmajer; Marina Vojković; Maja Strozzi; Anton Smalcelj; Zoran Lasić
Journal:  Coll Antropol       Date:  2008-06

10.  Influence of the timing of cardiac catheterization and the amount of contrast media on acute renal failure after cardiac surgery.

Authors:  Marco Ranucci; Andrea Ballotta; Alessia Kunkl; Donatella De Benedetti; Hassan Kandil; Daniela Conti; Nadia Mollichelli; Eduardo Bossone; Rajendra H Mehta
Journal:  Am J Cardiol       Date:  2008-03-04       Impact factor: 2.778

View more
  1 in total

1.  Imaging for approach selection of TAVI: assessment of the aorto-iliac tract diameter by computed tomography-angiography versus projection angiography.

Authors:  E M A Wiegerinck; H A Marquering; N Y Oldenburger; M A Elattar; R N Planken; B A J M De Mol; J J Piek; J Baan
Journal:  Int J Cardiovasc Imaging       Date:  2013-12-11       Impact factor: 2.357

  1 in total