Carotid CTA: radiation exposure and image quality with the use of attenuation-based, automated kilovolt selection.

A Eller1, W Wuest, M Kramer, M May, A Schmid, M Uder, M M Lell.   

Abstract

BACKGROUND AND
PURPOSE: CTA is considered the imaging modality of choice in evaluating the supraaortic vessels in many institutions, but radiation exposure remains a matter of concern. The objective of the study was to evaluate a fully automated, attenuation-based kilovolt selection algorithm in carotid CTA in respect to radiation dose and image quality compared with a standard 120-kV protocol.
MATERIALS AND METHODS: Ninety-eight patients were included: 53 examinations (patient age, 66 ± 12 years) were performed by use of automated adaption of tube potential (80-140 kV) on the basis of the attenuation profile of the scout scan (study group), and 45 examinations (patient age, 67 ± 11 years) were performed by use of a standard 120-kV protocol (control group). CT dose index volume and dose-length product were recorded from the examination protocol. Image quality was assessed by ROI measurements and calculations of SNR and contrast-to-noise ratio. Subjective image quality was evaluated by 2 observers with the use of a 4-point scale (3, excellent; 0, not diagnostic).
RESULTS: Subjective image quality was rated as "excellent" or "good" in all examinations (study group, 2.8; control group, 2.8). The algorithm automatically selected 100 kV in 47% and 80 kV in 34%; 120 kV was retained in 19%. An elevation to 140 kV did not occur. Compared with the control group, overall CT dose index volume reduction was 33.7%; overall dose-length product reduction was 31.5%. In the low-kilovolt scans, image noise and mean attenuation of ROIs inside the carotid arteries were significantly higher than in 120-kV scans, resulting in a constant or increased (80-kV group) contrast-to-noise ratio.
CONCLUSIONS: The attenuation-based, kilovolt selection algorithm enables a dose reduction of >30% in carotid artery CTA while maintaining contrast-to-noise ratio and subjective image quality at adequate levels.

Entities:  

Mesh:

Year:  2013        PMID: 23907241      PMCID: PMC7965763          DOI: 10.3174/ajnr.A3659

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  36 in total

Review 1.  Radiation dose of cardiac computed tomography - what has been achieved and what needs to be done.

Authors:  Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2010-10-19       Impact factor: 5.315

Review 2.  Technical approaches to the optimisation of CT.

Authors:  Willi A Kalender; Stefanie Buchenau; Paul Deak; Markus Kellermeier; Oliver Langner; Marcel van Straten; Sabrina Vollmar; Sylvia Wilharm
Journal:  Phys Med       Date:  2008-03-10       Impact factor: 2.685

3.  Application- and patient size-dependent optimization of x-ray spectra for CT.

Authors:  Willi A Kalender; Paul Deak; Markus Kellermeier; Marcel van Straten; Sabrina V Vollmar
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

4.  Contrast-enhanced MR angiography is not more accurate than unenhanced 2D time-of-flight MR angiography for determining > or = 70% internal carotid artery stenosis.

Authors:  L S Babiarz; J M Romero; E K Murphy; B Brobeck; P W Schaefer; R G González; M H Lev
Journal:  AJNR Am J Neuroradiol       Date:  2009-01-22       Impact factor: 3.825

Review 5.  Iterative reconstruction for coronary CT angiography: finding its way.

Authors:  Jonathon Leipsic; Brett G Heilbron; Cameron Hague
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-27       Impact factor: 2.357

Review 6.  2011 ASA/ACCF/AHA/AANN/AANS/ACR/ASNR/CNS/SAIP/SCAI/SIR/SNIS/SVM/SVS guideline on the management of patients with extracranial carotid and vertebral artery disease: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American Stroke Association, American Association of Neuroscience Nurses, American Association of Neurological Surgeons, American College of Radiology, American Society of Neuroradiology, Congress of Neurological Surgeons, Society of Atherosclerosis Imaging and Prevention, Society for Cardiovascular Angiography and Interventions, Society of Interventional Radiology, Society of NeuroInterventional Surgery, Society for Vascular Medicine, and Society for Vascular Surgery.

Authors:  Thomas G Brott; Jonathan L Halperin; Suhny Abbara; J Michael Bacharach; John D Barr; Ruth L Bush; Christopher U Cates; Mark A Creager; Susan B Fowler; Gary Friday; Vicki S Hertzberg; E Bruce McIff; Wesley S Moore; Peter D Panagos; Thomas S Riles; Robert H Rosenwasser; Allen J Taylor
Journal:  J Am Coll Cardiol       Date:  2011-02-01       Impact factor: 24.094

7.  Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness.

Authors:  Anna Winklehner; Robert Goetti; Stephan Baumueller; Christoph Karlo; Bernhard Schmidt; Rainer Raupach; Thomas Flohr; Thomas Frauenfelder; Hatem Alkadhi
Journal:  Invest Radiol       Date:  2011-12       Impact factor: 6.016

8.  Carotid atherosclerotic plaque progression and change in plaque composition over time: a 5-year follow-up study using serial CT angiography.

Authors:  M J van Gils; D Vukadinovic; A C van Dijk; D W J Dippel; W J Niessen; A van der Lugt
Journal:  AJNR Am J Neuroradiol       Date:  2012-02-16       Impact factor: 3.825

9.  Vascular enhancement and image quality of MDCT pulmonary angiography in 400 cases: comparison of standard and low kilovoltage settings.

Authors:  Shin Matsuoka; Andetta R Hunsaker; Ritu R Gill; Isabel B Oliva; Beatrice Trotman-Dickenson; Francine L Jacobson; Hiroto Hatabu
Journal:  AJR Am J Roentgenol       Date:  2009-06       Impact factor: 3.959

10.  The current role of carotid duplex ultrasonography in the management of carotid atherosclerosis: foundations and advances.

Authors:  Kelly R Byrnes; Charles B Ross
Journal:  Int J Vasc Med       Date:  2012-03-07
View more
  10 in total

1.  Carotid CTA at the Lowest Tube Voltage (70 kV) in Comparison with Automated Tube Voltage Adaption.

Authors:  A Eller; M Wiesmüller; W Wüst; R Heiss; M Kopp; M Saake; M Brand; M Uder; M M May
Journal:  AJNR Am J Neuroradiol       Date:  2019-07-11       Impact factor: 3.825

2.  Imaging the Parasinus Region with a Third-Generation Dual-Source CT and the Effect of Tin Filtration on Image Quality and Radiation Dose.

Authors:  M M Lell; M S May; M Brand; A Eller; T Buder; E Hofmann; M Uder; W Wuest
Journal:  AJNR Am J Neuroradiol       Date:  2015-03-26       Impact factor: 3.825

Review 3.  [Imaging of the head and neck region].

Authors:  M Lell; K Mantsopoulos; M Uder; W Wuest
Journal:  HNO       Date:  2016-03       Impact factor: 1.284

4.  Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose.

Authors:  Matthias S May; Manuel R Kramer; Achim Eller; Wolfgang Wuest; Michael Scharf; Michael Brand; Marc Saake; Bernhard Schmidt; Michael Uder; Michael M Lell
Journal:  Neuroradiology       Date:  2014-06-25       Impact factor: 2.804

5.  Comparison of virtual monoenergetic and polyenergetic images reconstructed from dual-layer detector CT angiography of the head and neck.

Authors:  Victor Neuhaus; Nils Große Hokamp; Nuran Abdullayev; Volker Maus; Christoph Kabbasch; Anastasios Mpotsaris; David Maintz; Jan Borggrefe
Journal:  Eur Radiol       Date:  2017-10-10       Impact factor: 5.315

6.  Automatic detection of carotid arteries in computed tomography angiography: a proof of concept protocol.

Authors:  Florentino Luciano Caetano Dos Santos; Atte Joutsen; Michelangelo Paci; Juha Salenius; Hannu Eskola
Journal:  Int J Cardiovasc Imaging       Date:  2016-05-03       Impact factor: 2.357

Review 7.  [Imaging of the head and neck region].

Authors:  M Lell; K Mantsopoulos; M Uder; W Wuest
Journal:  Radiologe       Date:  2016-02       Impact factor: 0.635

8.  Comparison of the image quality and radiation dose of different scanning modes in head-neck CT angiography.

Authors:  Tianle Zhang; Xue Geng; Dongxue Li; Yize Xu; Yongxia Zhao
Journal:  Dentomaxillofac Radiol       Date:  2021-01-06       Impact factor: 3.525

9.  Computed Tomography Angiography of Carotid Arteries and Vertebrobasilar System: A Simulation Study for Radiation Dose Reduction.

Authors:  Manuel Kramer; Stephan Ellmann; Thomas Allmendinger; Achim Eller; Ferdinand Kammerer; Matthias S May; João F Baigger; Michael Uder; Michael M Lell
Journal:  Medicine (Baltimore)       Date:  2015-07       Impact factor: 1.889

10.  Interleaving cerebral CT perfusion with neck CT angiography. Part II: clinical implementation and image quality.

Authors:  Marcel T H Oei; Frederick J A Meijer; Willem-Jan van der Woude; Ewoud J Smit; Bram van Ginneken; Rashindra Manniesing; Mathias Prokop
Journal:  Eur Radiol       Date:  2016-09-21       Impact factor: 5.315

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.