Clinical evaluation of a new continuous intraarterial blood gas monitoring system in the intensive care setting.

S Nunomiya1, T Tsujimoto, M Tanno, N Matsuyama, K Ohtake, T Kubota.   

Abstract

The present study was designed to evaluate a new continuous intraarterial blood gas monitoring system under routine clinical intensive care conditions. Nine mechanically ventilated adult patients were enrolled in this study. A multiparameter intravascular sensor was inserted into the radial or dorsalis pedis artery through a 20-gauge cannula in each patient. The accuracy of the sensor for pH, Pco2, and Po2 values was evaluated by comparing the data simultaneously obtained from the monitoring system and from conventional blood gas analysis. Measurements were performed for 3 days for each sensor. A total of 62 blood samples were obtained for comparison. The ranges of measured variables were: pH 7.185-7.602, Pco2, 28.8-68.5 mmHg, and Po2 45.2-542.4 mmHg. The overall bias ±precision values were 0.002±0.018 for pH units, 0.53±2.04mmHg for Pco2, and -1.62±20.00 mmHg for Po2. In clinically important ranges of Po2, less than 200 mmHg in particular, the bias and precision values were -2.25±6.48 mmHg in the range of less than 100mmHg, and 0.98±14.38 mmHg in the range of 100-200 mmHg. Variations of sensor accuracy as a function of elapsed time were within the clinically acceptable range throughout the study period. These findings suggest that this new device is sufficiently useful for routine clinical settings.

Entities:  

Year:  1996        PMID: 23839619     DOI: 10.1007/BF02471384

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  4 in total

1.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

2.  Continuous intra-arterial blood gas and pH monitoring in critically ill patients with severe respiratory failure: a prospective, criterion standard study.

Authors:  M Haller; E Kilger; J Briegel; H Forst; K Peter
Journal:  Crit Care Med       Date:  1994-04       Impact factor: 7.598

3.  Initial evaluation of a new intra-arterial blood gas system in humans.

Authors:  J L Zimmerman; R P Dellinger
Journal:  Crit Care Med       Date:  1993-04       Impact factor: 7.598

4.  A multiparameter sensor for continuous intra-arterial blood gas monitoring: a prospective evaluation.

Authors:  B Venkatesh; T H Clutton Brock; S P Hendry
Journal:  Crit Care Med       Date:  1994-04       Impact factor: 7.598

  4 in total
  1 in total

1.  Oxygen reserve index, a new method of monitoring oxygenation status: what do we need to know?

Authors:  Shu-Ting Chen; Su Min
Journal:  Chin Med J (Engl)       Date:  2020-01-20       Impact factor: 2.628

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.