Literature DB >> 8143468

A multiparameter sensor for continuous intra-arterial blood gas monitoring: a prospective evaluation.

B Venkatesh1, T H Clutton Brock, S P Hendry.   

Abstract

OBJECTIVE: To compare measurements of arterial blood gases made by a new continuous intra-arterial blood gas monitor with measurements in a standard blood gas analyzer in patients in the general and cardiac intensive care units.
DESIGN: Criterion standard study.
SETTING: The cardiac surgical and the general medical intensive care units of a tertiary referral center. PATIENTS: Thirteen consecutive patients requiring mechanical ventilation and blood gas monitoring.
INTERVENTIONS: All patients had a blood gas sensor placed through a 20-gauge cannula inserted into the radial artery. The duration of monitoring ranged from 9.42 to 117.45 hrs.
MEASUREMENTS AND MAIN RESULTS: A total of 158 simultaneous measurements of pH, PCO2, and PO2 were made from the sensor and the blood gas analyzer, and the bias and precision were calculated on all measured parameters. The overall bias +/- precision values were 0.01 +/- 0.06 for pH, 1.4 +/- 4.8 torr (0.2 +/- 0.7 kPa) for PCO2, and 2.8 +/- 25.6 torr (0.4 +/- 3.4 kPa) for PO2. The bias and precision for PO2 measurements that were < 150 torr (< 20 kPa) were 0.45 +/- 20.7 torr (0.1 +/- 2.8 kPa). The bias and precision for values of PO2 that were > 150 torr (> 20 kPa) were -8.1 +/- 28 torr (-1.1 +/- 3.8 kPa). The mean in vitro 90% response times of pH, PCO2, and PO2 sensors were found to be 78, 143, and 70 secs, respectively. There were no instances of any complications attributable to the sensor.
CONCLUSIONS: The continuous intra-arterial blood gas monitor tested in this study measures and trends arterial blood gases with an acceptable level of clinical accuracy. Longevity and safety of sensor function have also been demonstrated.

Entities:  

Mesh:

Year:  1994        PMID: 8143468     DOI: 10.1097/00003246-199404000-00013

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  15 in total

1.  [Continuous measurement of jugular venous blood gas. A case of subarachnoid bleeding].

Authors:  M Menzel; J Soukup; A Rieger; S Roth; J Radke; W Burkert
Journal:  Anaesthesist       Date:  1997-04       Impact factor: 1.041

Review 2.  Continuous intra-arterial blood gas monitoring.

Authors:  B Venkatesh; S P Hendry
Journal:  Intensive Care Med       Date:  1996-08       Impact factor: 17.440

3.  Optical Sensing and Imaging of pH Values: Spectroscopies, Materials, and Applications.

Authors:  Andreas Steinegger; Otto S Wolfbeis; Sergey M Borisov
Journal:  Chem Rev       Date:  2020-11-04       Impact factor: 60.622

4.  Misleading information from pulse oximetry and the usefulness of continuous blood gas monitoring in a post cardiac surgery patient.

Authors:  J P Benson; B Venkatesh; V Patla
Journal:  Intensive Care Med       Date:  1995-05       Impact factor: 17.440

5.  Clinical evaluation of a new continuous intraarterial blood gas monitoring system in the intensive care setting.

Authors:  S Nunomiya; T Tsujimoto; M Tanno; N Matsuyama; K Ohtake; T Kubota
Journal:  J Anesth       Date:  1996-09       Impact factor: 2.078

Review 6.  Continuous arterial gas monitoring: initial experience with the Paratrend 7 in children.

Authors:  I K Weiss; S Fink; S Edmunds; R Harrison; K Donnelly
Journal:  Intensive Care Med       Date:  1996-12       Impact factor: 17.440

7.  Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor.

Authors:  C Morgan; S J Newell; D A Ducker; J Hodgkinson; D K White; C J Morley; J M Church
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-03       Impact factor: 5.747

8.  Intra-arterial blood gas monitoring system: more accurate values can be obtained.

Authors:  K Kurahashi; Y Hirose; H Yamada; M Toyoshima; Y Usuda
Journal:  J Clin Monit       Date:  1996-03

9.  Subcutaneous oxygen pressure in spontaneously breathing lean and obese volunteers: a pilot study.

Authors:  Luzius B Hiltebrand; Heiko A Kaiser; Dagmar J Niedhart; Gunther Pestel; Andrea Kurz
Journal:  Obes Surg       Date:  2007-12-07       Impact factor: 4.129

10.  Clinical evaluation of a multiparameter intra-arterial blood-gas sensor.

Authors:  E Abraham; T J Gallagher; S Fink
Journal:  Intensive Care Med       Date:  1996-05       Impact factor: 17.440