Validity of the Water Hammer Formula for Determining Regional Aortic Pulse Wave Velocity: Comparison of One-Point and Two-Point (Foot-to-Foot) Measurements Using a Multisensor Catheter in Human.

Shizuo Hanya1.   

Abstract

BACKGROUND: Lack of high-fidelity simultaneous measurements of pressure and flow velocity in the aorta has impeded the direct validation of the water-hammer formula for estimating regional aortic pulse wave velocity (AO-PWV1) and has restricted the study of the change of beat-to-beat AO-PWV1 under varying physiological conditions in man.
METHODS: Aortic pulse wave velocity was derived using two methods in 15 normotensive subjects: 1) the conventional two-point (foot-to-foot) method (AO-PWV2) and 2) a one-point method (AO-PWV1) in which the pressure velocity-loop (PV-loop) was analyzed based on the water hammer formula using simultaneous measurements of flow velocity (Vm) and pressure (Pm) at the same site in the proximal aorta using a multisensor catheter. AO-PWV1 was calculated from the slope of the linear regression line between Pm and Vm where wave reflection (Pb) was at a minimum in early systole in the PV-loop using the water hammer formula, PWV1 = (Pm/Vm)/ρ, where ρ is the blood density. AO-PWV2 was calculated using the conventional two-point measurement method as the distance/traveling time of the wave between 2 sites for measuring P in the proximal aorta. Beat-to-beat alterations of AO-PWV1 in relationship to aortic pressure and linearity of the initial part of the PV-loop during a Valsalva maneuver were also assessed in one subject.
RESULTS: The initial part of the loop became steeper in association with the beat-to-beat increase in diastolic pressure in phase 4 during the Valsalva maneuver. The linearity of the initial part of the PV-loop was maintained consistently during the maneuver. Flow velocity vs. pressure in the proximal aorta was highly linear during early systole, with Pearson's coefficients ranging from 0.9954 to 0.9998. The average values of AO-PWV1 and AO-PWV2 were 6.3 ± 1.2 and 6.7 ± 1.3 m/s, respectively. The regression line of AO-PWV1 on AO-PWV2 was y = 0.95x + 0.68 (r = 0.93, p <0.001).
CONCLUSION: This study concluded that the water-hammer formula (one-point method) provides a reliable and conventional estimate of beat-to-beat aortic regional pulse wave velocity consistently regardless of the changes in physiological states in human clinically. (English Translation of J Jpn Coll Angiol 2011; 51: 215-221).

Entities:  

Keywords:  aortic elasticity; aortic wave reflection; multisensor catheter; regional pulse wave velocity of the proximal aorta; water-hammer formula

Year:  2013        PMID: 23825494      PMCID: PMC3692983          DOI: 10.3400/avd.oa.13-00046

Source DB:  PubMed          Journal:  Ann Vasc Dis        ISSN: 1881-641X


  20 in total

1.  Validity, reproducibility, and clinical significance of noninvasive brachial-ankle pulse wave velocity measurement.

Authors:  Akira Yamashina; Hirofumi Tomiyama; Kazuhiro Takeda; Hideichi Tsuda; Tomio Arai; Kenichi Hirose; Yutaka Koji; Saburoh Hori; Yoshio Yamamoto
Journal:  Hypertens Res       Date:  2002-05       Impact factor: 3.872

Review 2.  Pulse wave analysis and pulse wave velocity: a critical review of their strengths and weaknesses.

Authors:  Justine Ina Davies; Allan D Struthers
Journal:  J Hypertens       Date:  2003-03       Impact factor: 4.844

3.  Pulsatile diameter and elastic modulus of the aortic arch in essential hypertension: a noninvasive study.

Authors:  R N Isnard; B M Pannier; S Laurent; G M London; B Diebold; M E Safar
Journal:  J Am Coll Cardiol       Date:  1989-02       Impact factor: 24.094

4.  Estimation of total arterial compliance: an improved method and evaluation of current methods.

Authors:  Z Liu; K P Brin; F C Yin
Journal:  Am J Physiol       Date:  1986-09

5.  Estimation of total systemic arterial compliance in humans.

Authors:  W K Laskey; H G Parker; V A Ferrari; W G Kussmaul; A Noordergraaf
Journal:  J Appl Physiol (1985)       Date:  1990-07

6.  Characteristic impedance of the proximal aorta determined in the time and frequency domain: a comparison.

Authors:  J P Dujardin; D N Stone
Journal:  Med Biol Eng Comput       Date:  1981-09       Impact factor: 2.602

7.  Alterations with age in the viscoelastic properties of human arterial walls.

Authors:  B M Learoyd; M G Taylor
Journal:  Circ Res       Date:  1966-03       Impact factor: 17.367

8.  Aortic pulse-wave velocity and its relationship to mortality in diabetes and glucose intolerance: an integrated index of vascular function?

Authors:  Kennedy Cruickshank; Lisa Riste; Simon G Anderson; John S Wright; Graham Dunn; Ray G Gosling
Journal:  Circulation       Date:  2002-10-15       Impact factor: 29.690

9.  Use of simultaneous pressure and velocity measurements to estimate arterial wave speed at a single site in humans.

Authors:  Justin E Davies; Zachary I Whinnett; Darrel P Francis; Keith Willson; Rodney A Foale; Iqbal S Malik; Alun D Hughes; Kim H Parker; Jamil Mayet
Journal:  Am J Physiol Heart Circ Physiol       Date:  2005-08-26       Impact factor: 4.733

10.  On-line noninvasive one-point measurements of pulse wave velocity.

Authors:  Akimitsu Harada; Takashi Okada; Kiyomi Niki; Dehua Chang; Motoaki Sugawara
Journal:  Heart Vessels       Date:  2002-12       Impact factor: 2.037

View more
  4 in total

Review 1.  A review of wave mechanics in the pulmonary artery with an emphasis on wave intensity analysis.

Authors:  J Su; O Hilberg; L Howard; U Simonsen; A D Hughes
Journal:  Acta Physiol (Oxf)       Date:  2016-09-29       Impact factor: 6.311

2.  Effect of acute resistance exercise on carotid artery stiffness and cerebral blood flow pulsatility.

Authors:  Wesley K Lefferts; Jacqueline A Augustine; Kevin S Heffernan
Journal:  Front Physiol       Date:  2014-03-19       Impact factor: 4.566

3.  Improved pressure contour analysis for estimating cardiac stroke volume using pulse wave velocity measurement.

Authors:  Shun Kamoi; Christopher Pretty; Joel Balmer; Shaun Davidson; Antoine Pironet; Thomas Desaive; Geoffrey M Shaw; J Geoffrey Chase
Journal:  Biomed Eng Online       Date:  2017-04-24       Impact factor: 2.819

4.  Photoplethysmography and ultrasonic-measurement-integrated simulation to clarify the relation between two-dimensional unsteady blood flow field and forward and backward waves in a carotid artery.

Authors:  Shusaku Sone; Toshiyuki Hayase; Kenichi Funamoto; Atsushi Shirai
Journal:  Med Biol Eng Comput       Date:  2016-07-27       Impact factor: 2.602

  4 in total