Radiation exposure to operating room personnel and patients during endovascular procedures.

Abhisekh Mohapatra1, Roy K Greenberg, Tara M Mastracci, Matthew J Eagleton, Brett Thornsberry.   

Abstract

OBJECTIVE: To characterize radiation exposure to patients and operating room personnel during fluoroscopic procedures.
METHODS: Patient dose information was collected from the imaging equipment. Real-time dosimetry was used to measure doses to the operators, scrub nurse, radiologic technologist (RT), and anesthesiologist in 39 cases of endovascular thoracoabdominal aortic aneurysm repair using fenestrated endografts. Overall equivalent doses and dose rates at time points of interest were noted and compared with the corresponding patient doses.
RESULTS: The dosimeter on the anesthesia equipment received 143 μSv (38-247) more radiation per case than the average operator, and the scrub nurse and RT received 106 μSv (66-146) and 100 μSv (55-145) less, respectively. Adjusting for protective lead aprons by the Webster methodology, the average operator received an effective dose of 38 μSv. Except for the RT, personnel doses were well correlated with patient dose as measured by kerma area product (KAP) (r = .82 for average operator, r = .85 for scrub nurse, and r = .86 for anesthesia; all P < .001) but less well correlated with fluoroscopy time or cumulative air kerma (CAK). When preoperative cone beam computed tomography was performed, the equivalent dose to the RT was 1.1 μSv (0.6-1.5) when using shielding and 37 μSv (22-53) when unshielded. Digital subtraction acquisitions accounted for a large fraction of all individuals' doses. Decreasing field size (and thus, increasing magnification) was associated with decreased KAP (r = .47; P < .001) and increased CAK (r = -.56; P < .001). The square of the field size correlated strongly with the KAP/CAK ratio (r = .99; P < .001). Increased lateral angulation of the C-arm increased both CAK and KAP (at field size, 22 cm; r = .54 and r = .44; both P < .001) and the average dose rate to an operator was 1.78 (1.37-2.31) times as high in a lateral projection as in a posterior-anterior projection.
CONCLUSIONS: Personnel doses were best correlated with KAP and less well correlated with fluoroscopy time or CAK. The dosimeter on the anesthesia equipment recorded the highest doses attributable to ineffective shielding. Operators can reduce the effective dose to themselves, the patient, and other personnel by minimizing the use of digital subtraction acquisitions, avoiding lateral angulation, using higher magnification levels when possible, and being diligent about the use of shielding during fluoroscopy cases.
Copyright © 2013. Published by Mosby, Inc.

Entities:  

Mesh:

Year:  2013        PMID: 23810300     DOI: 10.1016/j.jvs.2013.02.032

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  13 in total

Review 1.  Physician and Patient Radiation Exposure During Endovascular Procedures.

Authors:  Andrew M Goldsweig; J Dawn Abbott; Herbert D Aronow
Journal:  Curr Treat Options Cardiovasc Med       Date:  2017-02

2.  Measures of patient radiation exposure during endoscopic retrograde cholangiography: beyond fluoroscopy time.

Authors:  Toufic Kachaamy; Edwyn Harrison; Rahul Pannala; William Pavlicek; Michael D Crowell; Douglas O Faigel
Journal:  World J Gastroenterol       Date:  2015-02-14       Impact factor: 5.742

3.  Typical exposure parameters, organ doses and effective doses for endovascular aortic aneurysm repair: Comparison of Monte Carlo simulations and direct measurements with an anthropomorphic phantom.

Authors:  Monika Foerth; Michael C Seidenbusch; Mojtaba Sadeghi-Azandaryani; Ursula Lechel; Karla Maria Treitl; Marcus Treitl
Journal:  Eur Radiol       Date:  2015-04-09       Impact factor: 5.315

4.  Investigation of reference levels and radiation dose associated with abdominal EVAR (endovascular aneurysm repair) procedures across several European Centres.

Authors:  E Tuthill; L O'Hora; M O'Donohoe; S Panci; P Gilligan; D Campion; R Trenti; E Fox; D Catania; L Rainford
Journal:  Eur Radiol       Date:  2017-05-18       Impact factor: 5.315

5.  A survey on knowledge, attitude, and practices of workplace radiation safety amongst anaesthesiology personnel in northern Indian tertiary care institutes.

Authors:  Rudrashish Haldar; Rafat Shamim; Himel Mondal; Ashish Kumar Kannaujia; Prabhakar Mishra; Anil Agarwal
Journal:  Indian J Anaesth       Date:  2022-05-17

6.  Three-dimensional fusion computed tomography decreases radiation exposure, procedure time, and contrast use during fenestrated endovascular aortic repair.

Authors:  Michael M McNally; Salvatore T Scali; Robert J Feezor; Daniel Neal; Thomas S Huber; Adam W Beck
Journal:  J Vasc Surg       Date:  2014-08-28       Impact factor: 4.268

7.  Radiation Awareness for Endovascular Abdominal Aortic Aneurysm Repair in the Hybrid Operating Room. An Instant Patient Risk Chart for Daily Practice.

Authors:  Quirina M de Ruiter; Crystel M Gijsberts; Constantijn E Hazenberg; Frans L Moll; Joost A van Herwaarden
Journal:  J Endovasc Ther       Date:  2017-04-10       Impact factor: 3.487

8.  Impact of Allura Clarity Technology on Radiation Dose Exposure During Left Atrial Appendage Closure.

Authors:  Emilia Studzińska; Maria Anna Staniszewska
Journal:  Pol J Radiol       Date:  2017-10-20

9.  A Magnetorheological Fluids-Based Robot-Assisted Catheter/Guidewire Surgery System for Endovascular Catheterization.

Authors:  Linshuai Zhang; Shuoxin Gu; Shuxiang Guo; Takashi Tamiya
Journal:  Micromachines (Basel)       Date:  2021-05-30       Impact factor: 2.891

Review 10.  Radiation protection in the cardiac catheterization laboratory.

Authors:  Sylvia Marie R Biso; Mladen I Vidovich
Journal:  J Thorac Dis       Date:  2020-04       Impact factor: 3.005

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.