| Literature DB >> 23793151 |
Anna J Warren1, Wes Armour, Danny Axford, Mark Basham, Thomas Connolley, David R Hall, Sam Horrell, Katherine E McAuley, Vitaliy Mykhaylyk, Armin Wagner, Gwyndaf Evans.
Abstract
The focus in macromolecular crystallography is moving towards even more challenging target proteins that often crystallize on much smaller scales and are frequently mounted in opaque or highly refractive materials. It is therefore essential that X-ray beamline technology develops in parallel to accommodate such difficult samples. In this paper, the use of X-ray microradiography and microtomography is reported as a tool for crystal visualization, location and characterization on the macromolecular crystallography beamlines at the Diamond Light Source. The technique is particularly useful for microcrystals and for crystals mounted in opaque materials such as lipid cubic phase. X-ray diffraction raster scanning can be used in combination with radiography to allow informed decision-making at the beamline prior to diffraction data collection. It is demonstrated that the X-ray dose required for a full tomography measurement is similar to that for a diffraction grid-scan, but for sample location and shape estimation alone just a few radiographic projections may be required.Entities:
Keywords: lipid cubic phase; membrane proteins; microradiography; microtomography
Mesh:
Substances:
Year: 2013 PMID: 23793151 PMCID: PMC3689528 DOI: 10.1107/S0907444913011359
Source DB: PubMed Journal: Acta Crystallogr D Biol Crystallogr ISSN: 0907-4449
Figure 1The experimental setup for X-ray tomography and radiography, showing (a) a schematic representation demonstrating the basic principles of X-ray microtomography and radiography, (b) the setup on beamline I04 at Diamond viewed from above and (c) the same setup viewed from the side, showing the installation of the equipment below the detector support.
Figure 2(a) View of membrane protein crystals of the human A2A adenosine G-protein coupled receptor in lipid cubic phase mounted on a nylon loop using a visible microscope; (b) the same orientation of the sample viewed as a radiograph. It is unclear in (a) where the crystals are located, while after X-ray imaging (b) it is evident that two crystals are located within the loop.
Figure 3View of the membrane protein crystal bacteriorhodopsin in lipid cubic phase mounted on a nylon loop (a) with the grid-scan set up for data collection, (b) the results of the good Bragg candidates from the grid-scan calculated using DISTL (Zhang et al., 2006 ▶) overlaid on the sample image, (c) the same orientation of the sample viewed as a radiograph and (d) the grid-scan results overlaid on the radiograph. The grid-scan and the radiograph both demonstrate how they can be used for determination of the sample location and both results map well, as can be seen in (d). The radiograph in (c) provides more information about the crystal size and shape which can then be used to allow better decision-making at the beamline for the diffraction experiment.
Data-collection statistics before and after the tomography experiments
| Before tomography | After tomography | |
|---|---|---|
| Wavelength (Å) | 1.380 | 1.380 |
| Space group |
|
|
| Unit-cell parameters (Å, °) |
|
|
| Resolution (Å) | 42.43–2.21 (2.28–2.21) | 42.53–2.21 (2.28–2.21) |
|
| 0.032 (0.155) | 0.045 (0.335) |
| 〈 | 23.3 (5.6) | 18.0 (2.7) |
| Completeness (%) | 100.0 (100.0) | 100.0 (100.0) |
| Multiplicity | 21.7 (21.8) | 21.7 (21.7) |
| No. of reflections | 315180 (26786) | 317994 (27142) |
| No. of unique reflections | 14544 (1230) | 14654 (1249) |
Figure 4A plot of the scaling results from SCALA (Evans, 2006 ▶) for CuNiR plotted as image number versus the B factor. Images 0–180 are from before the imaging experiments and images 181–360 are the diffraction images after the experiments.
Figure 5Visual hull reconstructions of the same slice through the A2A crystal along the pin axis, with a varying number of projections evenly spaced over the sequence of 180 × 1° frames. (a) Two projections (90° apart), (b) four projections (45° apart), (c) nine projections (20° apart), (d) 36 projections (5° apart) and (e) 180 projections (all frames). The same slice through a tomographic reconstruction is also shown in (f), where the crystal is visible in black surrounded by LCP in the loop in dark grey. It can be clearly seen that even with a visual hull reconstruction using four projections (b) a good approximation to the correct shape can be made when compared with the tomographic slice (f).