Performance evaluation of ISFETs and other ISE sensors for whole blood ion assay.

J M Thompson1.   

Abstract

Performance evaluations of ion-selective field-effect transistors (ISFETs) have been carried out using both quality-control materials and whole arterial blood samples. Comparison of the results from these evaluations suggests that the whole blood evaluation may be more useful when assessing the value of particular sensors for clinical applications. The effect of outliers on the imprecision estimates is demonstrated for ISFETs and ISEs, both graphically and in the calculation of the estimates with and without the outliers present. Estimates of constant and proportional bias, against an alternative sensor, determined from the intercept and slope of the linear regression vary according to the regression method used. The bias estimates obtained for the K + ISFET against the Radiometer KNA1 using ordinary least squares regression are compared with the Deming/Mandel method and the three-group resistant line method of Tukey. The Thorn EMI ISFETs are demonstrated to have acceptable imprecision and only a small bias compared with direct ISE instruments for whole blood assay and can be considered suitable for incorporation into clinical instrumentation.

Mesh:

Substances:

Year:  1990        PMID: 2377000     DOI: 10.1007/bf02442678

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  7 in total

1.  Applications of statistics in clinical chemistry. A critical evaluation of regression lines.

Authors:  P J Wakkers; H B Hellendoorn; G J Op de Weegh; W Heerspink
Journal:  Clin Chim Acta       Date:  1975-10-15       Impact factor: 3.786

2.  A study of the accuracy and precision of clinical chemistry determinations in 170 Canadian laboratories.

Authors:  D B TONKS
Journal:  Clin Chem       Date:  1963-04       Impact factor: 8.327

3.  Comparison of potassium ISFETs with the Radiometer KNA1 and the Corning 902 ion selective electrode analysers for whole blood potassium ion estimation.

Authors:  J M Thompson; C Emmett; S C Smith; R Cramb; P Hutton
Journal:  Ann Clin Biochem       Date:  1989-05       Impact factor: 2.057

4.  Incorrect least-squares regression coefficients in method-comparison analysis.

Authors:  P J Cornbleet; N Gochman
Journal:  Clin Chem       Date:  1979-03       Impact factor: 8.327

5.  Online patient-monitoring system for the simultaneous analysis of blood K+, Ca2+, Na+ and pH using a quadruple-function ChemFET integrated-circuit sensor.

Authors:  A Sibbald; A K Covington; R F Carter
Journal:  Med Biol Eng Comput       Date:  1985-07       Impact factor: 2.602

6.  Medical significance of laboratory results.

Authors:  R N Barnett
Journal:  Am J Clin Pathol       Date:  1968-12       Impact factor: 2.493

Review 7.  Electrodes in clinical chemistry.

Authors:  A D Hirst; J F Stevens
Journal:  Ann Clin Biochem       Date:  1985-09       Impact factor: 2.057

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.