Abstract
A high-performance liquid chromatography/positive ion electrospray tandem mass spectrometry method for the simultaneous quantification of efavirenz, emtricitabine and tenofovir was developed and validated with 100 microL human plasma. Following solid-phase extraction, the analytes were separated using a gradient mobile phase on a reverse-phase column and analyzed by MS/MS in the multiple reaction monitoring mode using the respective [M + H]+ ions, m/z 316 to 168 for efavirenz, m/z 248-130 for emtricitabine and m/z 288-176 for tenofovir, m/z 482-258 for rosuvastatin (IS), m/z 260-116 for propranolol (IS). The method exhibited a 100-fold linear dynamic range for all the three analytes in human plasma (20-2000, 2-200 and 20-2000 ng/mL for efavirenz, emtricitabine and tenofovir respectively). The lower limit of quantification was 2 ng/mL for emtricitabine and 20 ng/mL for both efavirenz and tenofovir with a relative standard deviation of less than 11%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. The total chromatographic run time of 4 min for each sample made it possible to analyze more than 250 human plasma samples per day. The method is precise and sensitive enough for its intended purpose. The method is also successfully applied to quantify efavirenz, emtricitabine and tenofovir concentrations in a rodent pharmacokinetic study.Entities:
Mesh:
Substances:
Year: 2009 PMID: 18937306 DOI: 10.1002/bmc.1125