A CT-based study investigating the relationship between pedicle screw placement and stimulation threshold of compound muscle action potentials measured by intraoperative neurophysiological monitoring.

Gerit Kulik1, Etienne Pralong, John McManus, Damien Debatisse, Constantin Schizas.   

Abstract

PURPOSE: Neurophysiological monitoring aims to improve the safety of pedicle screw placement, but few quantitative studies assess specificity and sensitivity. In this study, screw placement within the pedicle is measured (post-op CT scan, horizontal and vertical distance from the screw edge to the surface of the pedicle) and correlated with intraoperative neurophysiological stimulation thresholds.
METHODS: A single surgeon placed 68 thoracic and 136 lumbar screws in 30 consecutive patients during instrumented fusion under EMG control. The female to male ratio was 1.6 and the average age was 61.3 years (SD 17.7). Radiological measurements, blinded to stimulation threshold, were done on reformatted CT reconstructions using OsiriX software. A standard deviation of the screw position of 2.8 mm was determined from pilot measurements, and a 1 mm of screw-pedicle edge distance was considered as a difference of interest (standardised difference of 0.35) leading to a power of the study of 75 % (significance level 0.05).
RESULTS: Correct placement and stimulation thresholds above 10 mA were found in 71 % of screws. Twenty-two percent of screws caused cortical breach, 80 % of these had stimulation thresholds above 10 mA (sensitivity 20 %, specificity 90 %). True prediction of correct position of the screw was more frequent for lumbar than for thoracic screws.
CONCLUSION: A screw stimulation threshold of >10 mA does not indicate correct pedicle screw placement. A hypothesised gradual decrease of screw stimulation thresholds was not observed as screw placement approaches the nerve root. Aside from a robust threshold of 2 mA indicating direct contact with nervous tissue, a secondary threshold appears to depend on patients' pathology and surgical conditions.

Entities:  

Mesh:

Year:  2013        PMID: 23686531      PMCID: PMC3777061          DOI: 10.1007/s00586-013-2754-0

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  26 in total

1.  Posterior lumbar interbody fusion: a retrospective study of complications after facet joint excision and pedicle screw fixation in 148 cases.

Authors:  K Okuyama; E Abe; T Suzuki; Y Tamura; M Chiba; K Sato
Journal:  Acta Orthop Scand       Date:  1999-08

2.  Pedicle screws with high electrical resistance: a potential source of error with stimulus-evoked EMG.

Authors:  D Greg Anderson; Lawrence R Wierzbowski; Daniel M Schwartz; Alan S Hilibrand; Alexander R Vaccaro; Todd J Albert
Journal:  Spine (Phila Pa 1976)       Date:  2002-07-15       Impact factor: 3.468

3.  Electrical stimulation for intraoperative evaluation of thoracic pedicle screw placement.

Authors:  Yong-bing Shi; Michael Binette; William H Martin; James M Pearson; Robert A Hart
Journal:  Spine (Phila Pa 1976)       Date:  2003-03-15       Impact factor: 3.468

4.  Probing for thoracic pedicle screw tract violation(s): is it valid?

Authors:  Ronald A Lehman; Benjamin K Potter; Timothy R Kuklo; Audrey S Chang; David W Polly; Scott B Shawen; Joseph R Orchowski
Journal:  J Spinal Disord Tech       Date:  2004-08

5.  Reduction in radiation (fluoroscopy) while maintaining safe placement of pedicle screws during lumbar spine fusion.

Authors:  Christopher D Chaput; Keri George; Amer F Samdani; John I Williams; John Gaughan; Randal R Betz
Journal:  Spine (Phila Pa 1976)       Date:  2012-10-01       Impact factor: 3.468

Review 6.  Roentgenographic measurement of pedicle screw penetration.

Authors:  T S Whitecloud; T C Skalley; S D Cook; E L Morgan
Journal:  Clin Orthop Relat Res       Date:  1989-08       Impact factor: 4.176

7.  Continuous intraoperative electromyographic recording during spinal surgery.

Authors:  R M Beatty; P McGuire; J M Moroney; F P Holladay
Journal:  J Neurosurg       Date:  1995-03       Impact factor: 5.115

8.  Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members.

Authors:  S I Esses; B L Sachs; V Dreyzin
Journal:  Spine (Phila Pa 1976)       Date:  1993-11       Impact factor: 3.468

9.  Comparison of transcranial electric motor and somatosensory evoked potential monitoring during cervical spine surgery.

Authors:  Alan S Hilibrand; Daniel M Schwartz; Venkat Sethuraman; Alexander R Vaccaro; Todd J Albert
Journal:  J Bone Joint Surg Am       Date:  2004-06       Impact factor: 5.284

10.  Can triggered electromyograph thresholds predict safe thoracic pedicle screw placement?

Authors:  Barry L Raynor; Lawrence G Lenke; Yongjung Kim; Darrell S Hanson; Tracy J Wilson-Holden; Keith H Bridwell; Anne M Padberg
Journal:  Spine (Phila Pa 1976)       Date:  2002-09-15       Impact factor: 3.468

View more
  4 in total

1.  Intraoperative neurophysiological monitoring for minimally invasive 1- and 2-level transforaminal lumbar interbody fusion: does it improve patient outcome?

Authors:  Juanita Garces; J Franklin Berry; Edison P Valle-Giler; Wale A R Sulaiman
Journal:  Ochsner J       Date:  2014

2.  Prevention of spinal cord injury using brain-evoked muscle-action potential (Br(E)-MsEP) monitoring in cervical spinal screw fixation.

Authors:  Kazuyoshi Kobayashi; Shiro Imagama; Zenya Ito; Kei Ando; Tetsuro Hida; Naoki Ishiguro
Journal:  Eur Spine J       Date:  2016-12-31       Impact factor: 3.134

3.  Combining pedicle screw stimulation with spinal navigation, a protocol to maximize the safety of neural elements and minimize radiation exposure in thoracolumbar spine instrumentation.

Authors:  Sebouh Z Kassis; Loay K Abukwedar; Abdul Karim Msaddi; Catalin N Majer; Walid Othman
Journal:  Eur Spine J       Date:  2015-04-29       Impact factor: 3.134

4.  Intraoperative electromyographic monitoring to optimize safe lumbar pedicle screw placement - a retrospective analysis.

Authors:  Arun-Kumar Kaliya-Perumal; Jiun-Ran Charng; Chi-Chien Niu; Tsung-Ting Tsai; Po-Liang Lai; Lih-Huei Chen; Wen-Jer Chen
Journal:  BMC Musculoskelet Disord       Date:  2017-05-30       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.