Patient-prosthesis mismatch and reduction in left ventricular mass after aortic valve replacement.

Kristian Kandler1, Christian H Møller, Christian Hassager, Peter S Olsen, Nikolaj Lilleør, Daniel A Steinbrüchel.   

Abstract

BACKGROUND: The presence of patient-prosthesis mismatch (PPM) after aortic valve replacement may influence patient survival. We examined the relationship between PPM and changes in left ventricular mass index at 3 months follow-up and also overall survival.
METHODS: From patients included in the Mosaic trial, we studied data from 266 patients who underwent aortic valve replacement with the Medtronic Mosaic porcine bioprosthesis and had an echocardiography performed 3 months postoperatively. Complete echocardiographic data, to calculate left ventricular mass index, was available in 78% of the patients. The primary outcome for this substudy was prevalence and severity of PPM. Secondary outcomes were reduction in left ventricular mass index at 3 months follow-up and medium-term survival. Patients without PPM were defined as having an indexed effective orifice area greater than 0.85 cm(2)/m(2), and those with moderate and severe PPM as having an indexed effective orifice area between 0.65 cm(2)/m(2) and 0.85 cm(2)/m(2) or below 0.65 cm(2)/m(2), respectively.
RESULTS: PPM was found in 217 (82%) patients. No difference in overall survival was found between patients with PPM and those without PPM. The change in left ventricular mass index was significantly different between groups (no PPM -31.4 ± 28.0 g/m(2), moderate PPM 1.1 ± 34.4 g/m(2), and severe PPM -5.9 ± 29.7 g/m(2), respectively (p = 0.01).
CONCLUSIONS: The presence of PPM did not influence medium-term survival. However, patients without PPM showed a marked reduction in left ventricular mass index as soon as 3 months postoperatively.
Copyright © 2013 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23684157     DOI: 10.1016/j.athoracsur.2013.03.089

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

1.  Prosthesis-patient mismatch after transcatheter aortic valve implantation: impact of 2D-transthoracic echocardiography versus 3D-transesophageal echocardiography.

Authors:  Cristina da Silva; Anders Sahlen; Reidar Winter; Magnus Bäck; Andreas Rück; Magnus Settergren; Aristomenis Manouras; Kambiz Shahgaldi
Journal:  Int J Cardiovasc Imaging       Date:  2014-08-08       Impact factor: 2.357

2.  Early and mid-term haemodynamic performance and clinical outcomes of St. Jude Medical Trifecta™ valve.

Authors:  Renata Raimundo; Soraia Moreira; Francisca Saraiva; Rui J Cerqueira; Pedro Teixeira; Elson Salgueiro; André Lourenço; Mário J Amorim; Jorge Almeida; Paulo Pinho; Adelino F Leite-Moreira
Journal:  J Thorac Dis       Date:  2018-02       Impact factor: 2.895

3.  Incidence, predictors, and outcome of prosthesis-patient mismatch after transcatheter aortic valve replacement: A meta-analysis.

Authors:  Shixin He; Zhenfei Fang
Journal:  Medicine (Baltimore)       Date:  2020-06-12       Impact factor: 1.817

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.