Literature DB >> 23683378

Comparison of a novel real-time SonixGPS needle-tracking ultrasound technique with traditional ultrasound for vascular access in a phantom gel model.

Daniel S Kopac1, Jerry Chen, Raymond Tang, Andrew Sawka, Himat Vaghadia.   

Abstract

OBJECTIVE: Ultrasound-guided percutaneous vascular access for endovascular procedures is well established in surgical practice. Despite this, rates of complications from venous and arterial access procedures remain a significant cause of morbidity. We hypothesized that the use of a new technique of vascular access using an ultrasound with a novel needle-guidance positioning system (GPS) would lead to improved success rates of vascular puncture for both in-plane and out-of-plane techniques compared with traditional ultrasound.
METHODS: A prospective, randomized crossover study of medical students from all years of medical school was conducted using a phantom gel model. Each medical student performed three ultrasound-guided punctures with each of the four modalities (in-plane no GPS, in-plane with GPS, out-of-plane no GPS, out-of-plane with GPS) for a total of 12 attempts. The success or failure was judged by the ability to aspirate a simulated blood solution from the model. The time to successful puncture was also recorded. A poststudy validated NASA Task Load Index workload questionnaire was conducted to assess the student's perceptions of the two different techniques.
RESULTS: A total of 30 students completed the study. There was no significant difference seen in the mean times of vascular access for each of the modalities. Higher success rates for vascular access using the GPS for both the in-plane (94% vs 91%) and the out-of-plane (86% vs 70%) views were observed; however, this was not statistically significant. The students perceived the mental demand (median 12.0 vs 14.00; P = .035) and effort to be lower (mean 11.25 vs 14.00; P = .044) as well as the performance to be higher (mean 15.50 vs 14.00; P = .041) for the GPS vs the traditional ultrasound-guided technique. Students also perceived their ability to access vessels increased with the aid of the GPS (7.00 vs 6.50; P = .007). The majority of students expressed a preference for GPS (26/30, 87%) as opposed to the traditional counterpart.
CONCLUSIONS: Use of the novel SonixGPS needle-tracking ultrasound system (UltraSonix, Richmond, BC, Canada) was not associated with a higher success rate of vascular puncture compared with the traditional ultrasound-guided technique. Assessment of mental task load significantly favored the use of the ultrasound GPS over the traditional ultrasound technique.
Copyright © 2013 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Mesh:

Year:  2013        PMID: 23683378     DOI: 10.1016/j.jvs.2013.03.007

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  10 in total

1.  Comparison of in-plane and out-of-plane needle insertion with vs. without needle guidance.

Authors:  Verena Constanze Meiser; Helene Kreysa; Orlando Guntinas-Lichius; Gerd Fabian Volk
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-10-23       Impact factor: 2.503

Review 2.  [Seeing more : Technical innovations in regional anesthesia].

Authors:  T Wiesmann; T Steinfeldt; T Volk; U Schwemmer; P Kessler; H Wulf
Journal:  Anaesthesist       Date:  2014-11       Impact factor: 1.041

3.  Ultra-Low Radiation Dose CT Fluoroscopy for Percutaneous Interventions: A Porcine Feasibility Study.

Authors:  Martin G Wagner; J Louis Hinshaw; Yinsheng Li; Timothy P Szczykutowicz; Paul Laeseke; Charles A Mistretta; Fred T Lee
Journal:  Radiology       Date:  2019-01-15       Impact factor: 11.105

4.  Systematic review of measurement tools to assess surgeons' intraoperative cognitive workload.

Authors:  R D Dias; M C Ngo-Howard; M T Boskovski; M A Zenati; S J Yule
Journal:  Br J Surg       Date:  2018-02-21       Impact factor: 6.939

5.  A new biplane ultrasound probe for real-time visualization and cannulation of the internal jugular vein.

Authors:  Jeremy Kaplowitz; Paul Bigeleisen
Journal:  Case Rep Anesthesiol       Date:  2014-03-13

6.  Needle tip tracking for ultrasound-guided peripheral nerve block procedures-An observer blinded, randomised, controlled, crossover study on a phantom model.

Authors:  Trine Kåsine; Luis Romundstad; Leiv Arne Rosseland; Kyrre Ullensvang; Morten Wang Fagerland; Per Kristian Hol; Paul Kessler; Axel Rudolf Sauter
Journal:  Acta Anaesthesiol Scand       Date:  2019-04-30       Impact factor: 2.105

7.  Development and Characterization of an Ex Vivo Testing Platform for Evaluating Automated Central Vascular Access Device Performance.

Authors:  Emily N Boice; David Berard; Sofia I Hernandez Torres; Guy Avital; Eric J Snider
Journal:  J Pers Med       Date:  2022-08-05

8.  Computer-assisted needle navigation for pediatric internal jugular central venous cannulation: A feasibility study.

Authors:  Christopher Gallo; Pezhman Foroughi; Elisabeth Meagher; Ranjith Vellody; Bhupender Yadav; Anthony Ho; Alican Demir; Dorothee Heisenberg; Kevin Cleary; Karun Sharma
Journal:  J Vasc Access       Date:  2020-04-27       Impact factor: 2.283

9.  Technical feasibility and efficacy of a standard needle magnetization system for ultrasound needle guidance in thyroid nodule-targeting punctures: a phantom study.

Authors:  Haein Lee; Yoon Joo Cho; Eun Ju Ha; Jayoung Moon; You Na Kim; Minji Kim; Kyung-Min Lee; Sung Hyun An
Journal:  Ultrasonography       Date:  2021-12-10

10.  Ultrasound-guided lumbar puncture with a needle-guidance system: A prospective and controlled study to evaluate the learnability and feasibility of a newly developed approach.

Authors:  Tilo Backhaus; Moritz von Cranach; Jochen Brich
Journal:  PLoS One       Date:  2018-04-09       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.