| Literature DB >> 35108776 |
Haein Lee1, Yoon Joo Cho1, Eun Ju Ha1, Jayoung Moon1, You Na Kim1, Minji Kim1, Kyung-Min Lee1, Sung Hyun An1.
Abstract
PURPOSE: The aim of this study was to assess the feasibility and efficacy of an ultrasound needle guidance system (NGS) based on standard needle magnetization in a phantom study of thyroid nodule (TN)-targeting punctures.Entities:
Keywords: Fine-needle aspiration; Needle guidance; Thyroid nodule; Ultrasound
Year: 2021 PMID: 35108776 PMCID: PMC9262671 DOI: 10.14366/usg.21211
Source DB: PubMed Journal: Ultrasonography ISSN: 2288-5919
Fig. 1.Phantom model for needle puncture.
Schematic illustration (A), ultrasonography (B), and photograph (C) of the phantom show two targets (7 mm and 4 mm) mimicking thyroid nodules.
Fig. 2.Preparation of the needle guidance system.
A. A portable ultrasonographic apparatus with a standard needle magnetization system and a needle are used. B. After a brief insertion of the needle into the magnetizer (1-2 seconds), the performance indicator in the bottom left corner of the display shows full green bars, indicating good functionality of the system.
Fig. 3.Ultrasonography-guided procedure with a needle guidance system.
The electromagnetic needle guidance system shows spatial information of the needle (dashed lines), as well as a target box superimposed on the currently acquired ultrasonographic images.
Comparison of the technical success rate and the number of punctures between the conventional in-plane technique and the technique assisted by the needle guidance system
| Group 1 | Group 2 | Group 3 | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Conventional | With NGS | P-value | Conventional | With NGS | P-value | Conventional | With NGS | P-value | ||
| Technical success rate (%) | 60.0 (24/40) | 80.0 (32/40) | 0.046 | 95.0 (38/40) | 100 (40/40) | 0.157 | 100 (60/60) | 98.3 (59/60) | 0.317 | |
| Right-handed needling | 60.0 (12/20) | 85.0 (17/20) | 0.059 | 90.0 (18/20) | 100 (20/20) | 0.157 | 100 (30/30) | 100 (30/30) | >0.99 | |
| 7 mm | 50.0 (5/10) | 80.0 (8/10) | 0.180 | 90.0 (9/10) | 100 (10/10) | 0.317 | 100 (15/15) | 100 (15/15) | >0.99 | |
| 4 mm | 70.0 (7/10) | 90.0 (9/10) | 0.157 | 90.0 (9/10) | 100 (10/10) | 0.317 | 100 (15/15) | 100 (15/15) | >0.99 | |
| Left-handed needling | 60.0 (12/20) | 75.0 (15/20) | 0.317 | 100 (20/20) | 100 (20/20) | >0.99 | 100 (30/30) | 96.7 (29/30) | 0.317 | |
| 7 mm | 60.0 (6/10) | 80.0 (8/10) | 0.157 | 100 (10/10) | 100 (10/10) | >0.99 | 100 (15/15) | 93.3 (14/15) | 0.317 | |
| 4 mm | 60.0 (6/10) | 70.0 (7/10) | 0.705 | 100 (10/10) | 100 (10/10) | >0.99 | 100 (15/15) | 100 (15/15) | >0.99 | |
| No. of punctures[ | 2.2 (87/40) | 1.2 (48/40) | 0.005 | 1.1 (42/40) | 1.0 (40/40) | 0.157 | 1.0 (60/60) | 1.0 (61/60) | 0.317 | |
| Right-handed needling | 2.0 (40/20) | 1.15 (23/20) | 0.018 | 1.1 (22/20) | 1.0 (20/20) | 0.157 | 1.0 (30/30) | 1.0 (30/30) | >0.99 | |
| 7 mm | 2.2 (22/10) | 1.2 (12/10) | 0.068 | 1.1 (11/10) | 1.0 (10/10) | 0.317 | 1.0 (15/15) | 1.0 (15/15) | >0.99 | |
| 4 mm | 1.8 (18/10) | 1.1 (11/10) | 0.102 | 1.1 (11/10) | 1.0 (10/10) | 0.317 | 1.0 (15/15) | 1.0 (15/15) | >0.99 | |
| Left-handed needling | 2.4 (47/20) | 1.3 (25/20) | 0.103 | 1.0 (20/20) | 1.0 (20/20) | >0.99 | 1.0 (30/30) | 1.0 (31/30) | 0.317 | |
| 7 mm | 2.5 (25/10) | 1.2 (12/10) | 0.180 | 1.0 (10/10) | 1.0 (10/10) | >0.99 | 1.0 (15/15) | 1.1 (16/15) | 0.317 | |
| 4 mm | 2.2 (22/10) | 1.3 (13/10) | 0.380 | 1.0 (10/10) | 1.0 (10/10) | >0.99 | 1.0 (15/15) | 1.0 (15/15) | >0.99 | |
Group 1 indicates radiology residents with <50 cases of experience with thyroid fine-needle aspiration, while group 2 indicates radiology residents with ≥50 cases and <100 cases of experience with thyroid fine-needle aspiration. Group 3 indicates experienced radiology residents and a staff radiologist with ≥100 cases of experience with thyroid fine-needle aspiration.
NGS, needle guidance system.
Numbers are mean values.
Comparison of the total procedure time between the conventional in-plane technique and the technique using the needle guidance system
| Group 1 | Group 2 | Group 3 | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Conventional | With NGS | P-value | Conventional | With NGS | P-value | Conventional | With NGS | P-value | ||
| Procedure time (s) | 46.3±59.2 | 24.4±23.1 | 0.041 | 27.9±21.3 | 17.6±7.9 | 0.010 | 6.3±2.4 | 9.0±3.6 | <0.001 | |
| Right-handed needling | 53.4±59.7 | 34.7±31.0 | 0.027 | 25.6±19.9 | 18.7±9.0 | 0.109 | 5.6±2.0 | 8.7±3.9 | 0.001 | |
| 7 mm | 56.6±48.1 | 31.1±38.3 | 0.126 | 29.0±23.3 | 20.0±7.0 | 0.373 | 4.4±0.5 | 6.0±1.7 | 0.038 | |
| 4 mm | 50.1±72.0 | 22.2±19.1 | 0.123 | 22.1±16.3 | 17.3±10.8 | 0.138 | 6.8±2.3 | 11.4±3.6 | 0.005 | |
| Left-handed needling | 39.2±59.3 | 22.2±14.2 | 0.641 | 30.3±22.9 | 16.5±6.8 | 0.070 | 7.0±2.6 | 9.2±3.4 | 0.018 | |
| 7 mm | 45.0±67.5 | 19.9±14.9 | 0.359 | 34.8±21.5 | 17.3±6.5 | 0.074 | 6.6±1.8 | 9.4±3.4 | 0.131 | |
| 4 mm | 33.4±52.8 | 24.5±13.8 | 0.799 | 25.7±24.4 | 15.6±7.3 | 0.677 | 7.4±3.4 | 9.0±3.7 | 0.078 | |
Values are presented as mean±standard deviation. Group 1 indicates radiology residents with <50 cases of experience with thyroid fine-needle aspiration, while group 2 indicates radiology residents with ≥50 cases of experience with thyroid fine-needle aspiration. Group 3 indicates experienced radiology residents and a staff radiologist with ≥100 cases of experience with thyroid fine-needle aspiration.
NGS, needle guidance system.