Alexander Nevelsky1, Nantakan Ieumwananonthachai, Orit Kaidar-Person, Raquel Bar-Deroma, Haitam Nasrallah, Rahamim Ben-Yosef, Abraham Kuten.
Abstract
The purpose of this study was to evaluate the feasibility of hippocampal-sparing whole-brain radiotherapy (HS WBRT) using the Elekta Infinity linear accelerator and Monaco treatment planning system (TPS). Ten treatment plans were created for HS-WBRT to a dose of 30 Gy (10 fractions). RTOG 0933 recommendations were applied for treatment planning. Intensity-modulated radiotherapy (IMRT) plans for the Elekta Infinity linear accelerator were created using Monaco 3.1 TPS-based on a nine-field arrangement and step-and-shoot delivery method. Plan evaluation was performed using D2% and D98% for the whole-brain PTV (defined as whole brain excluding hippocampus avoidance region), D100% and maximum dose to the hippocampus, and maximum dose to optic nerves and chiasm. Homogeneity index (HI) defined as (D2%-D98%)/Dmedian was used to quantify dose homogeneity in the PTV. The whole-brain PTV D2% mean value was 37.28 Gy (range 36.95-37.49Gy), and D98% mean value was 25.37 Gy (range 25.40-25.89 Gy). The hippocampus D100% mean value was 8.37 Gy (range 7.48-8.97 Gy) and the hippocampus maximum dose mean value was 14.35 Gy (range 13.48-15.40 Gy). The maximum dose to optic nerves and optic chiasm for all patients did not exceed 37.50 Gy. HI mean value was 0.36 (range 0.34-0.37). Mean number of segments was 105 (range 88-122) and mean number of monitor units was 1724 (range 1622-1914). Gamma evaluation showed that all plans passed 3%, 3 mm criteria with more than 99% of the measured points. These results indicate that Elekta equipment (Elekta Infinity linac and Monaco TPS) can be used for HS WBRT planning according to compliance criteria defined by the RTOG 0933 protocol.Entities:
Mesh:
Year: 2013 PMID: 23652251 PMCID: PMC5714429 DOI: 10.1120/jacmp.v14i3.4205
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
RTOG 0933 compliance criteria.
|
|
|
|---|---|
| Whole‐brain PTV |
|
|
| |
|
| |
| Hippocampus |
|
|
| |
| Maximum dose to optic chiasm and optic nerves | 37.5 Gy |
Beam arrangement used in our study.
|
|
|
|
|
|---|---|---|---|
| 1 | 10 | 45 | 0 |
| 2 | 60 | 45 | 0 |
| 3 | 130 | 45 | 0 |
| 4 | 170 | 45 | 0 |
| 5 | 220 | 45 | 0 |
| 6 | 270 | 45 | 0 |
| 7 | 320 | 45 | 0 |
| 8 | 290 | 0 | 90 |
| 9 | 330 | 0 | 90 |
Set of segmentation parameters which produced plans of acceptable quality with reasonable number of segments.
| Calculation grid | 0.2 cm |
| Monte Carlo variance per plan | 1% |
| Minimum segment area | 4 cm |
| Minimum segment change area | 4 cm |
| Minimum segment width | 0.5 cm |
| Minimum MU/segment | 6 MU |
Figure 1Dose distribution (axial, coronal, and sagittal views) at the level of the hippocampi for hippocampal avoidance during whole‐brain radiotherapy for one representative patient. Red contour represents the hippocampus. Blue isodose represents 12 Gy; light blue, 25 Gy; green, 30 Gy; orange, 35 Gy; red, 37.5 Gy, in ten fractions.
Figure 2A cumulative normalized dose‐volume histogram for whole‐brain radiotherapy with hippocampal sparing for one representative patient.
Detailed dosimetric results for ten patients.
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|
| 1 | 36.98 | 25.58 | 93.2 | 7.72 | 15.21 | 29.62 | 36.44 | 0.34 |
| 2 | 37.41 | 25.74 | 92.8 | 8.62 | 15.75 | 35.30 | 36.60 | 0.36 |
| 3 | 36.94 | 25.87 | 93.0 | 8.03 | 15.38 | 29.95 | 36.50 | 0.34 |
| 4 | 37.57 | 25.72 | 91.8 | 8.82 | 15.94 | 30.90 | 34.49 | 0.36 |
| 5 | 37.22 | 25.57 | 92.0 | 8.48 | 15.16 | 34.84 | 36.66 | 0.37 |
| 6 | 36.98 | 25.40 | 90.7 | 8.92 | 15.49 | 34.62 | 35.49 | 0.37 |
| 7 | 37.22 | 25.89 | 93.2 | 8.82 | 15.74 | 34.08 | 36.50 | 0.37 |
| 8 | 37.50 | 25.65 | 90.5 | 8.87 | 14.28 | 35.74 | 33.94 | 0.37 |
| 9 | 36.96 | 25.61 | 91.6 | 7.68 | 16.06 | 31.67 | 37.15 | 0.36 |
| 10 | 36.93 | 25.69 | 92.0 | 7.83 | 14.87 | 28.27 | 36.00 | 0.37 |
Detailed information on number of segments, number of MUs, and gamma values for ten patients.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| 1 | 104 | 1644 | 13:05 | 93.9 | 99.5 |
| 2 | 88 | 1640 | 11:48 | 93.4 | 99.2 |
| 3 | 122 | 1675 | 14:32 | 93.2 | 99.2 |
| 4 | 95 | 1640 | 12:18 | 93.3 | 99.4 |
| 5 | 119 | 1854 | 14:52 | 93.5 | 99.5 |
| 6 | 109 | 1914 | 14:11 | 93.5 | 99.4 |
| 7 | 106 | 1692 | 13:24 | 93.3 | 99.3 |
| 8 | 90 | 1622 | 11:56 | 93.5 | 99.3 |
| 9 | 109 | 1807 | 13:46 | 93.8 | 99.6 |
| 10 | 120 | 1750 | 14:42 | 93.4 | 99.2 |