Literature DB >> 23650901

Patch pump versus conventional pump: postprandial glycemic excursions and the influence of wear time.

Yoeri M Luijf1, Sabine Arnolds, Angelo Avogaro, Carsten Benesch, Daniela Bruttomesso, Anne Farret, Lutz Heinemann, Jerome Place, Eric Renard, Rachele Scotton, J Hans DeVries.   

Abstract

BACKGROUND AND AIMS: The aim of this study was to compare blood glucose and plasma insulin profiles after bolus insulin infusion by a patch pump (PP) versus a conventional pump (CP), directly after placement and after Day 3 of use. PATIENTS AND METHODS: Twenty patients with type 1 diabetes came in for two blocks of visits: one block of two visits while wearing the OmniPod® (Insulet Corp., Bedford, MA) insulin pump (PP) and one block of two visits while wearing the Medtronic Diabetes (Northridge, CA) Paradigm® pump (CP). Patients administered an identical mealtime insulin bolus of at least 6 IU.
RESULTS: For PP, maximum glucose levels were 28.7% lower on Day 3 (P=0.020), when maximum insulin levels were 30.3% higher (P=0.002). For CP, maximum glucose levels were 26.5% lower on Day 3 (P=0.015), when maximum insulin levels were 46.4% higher (P=0.003). Glucose levels (mean [interquartile range]) were significantly lower on Day 3 for PP (168.2 [145.8] mg/dL vs. 139.4 [77.8] mg/dL; P=0.013), but not significantly so for CP (159.0 [66.1] mg/dL vs. 139.5 [57.9] mg/dL; P=0.084). Mean insulin levels were significantly higher on Day 3 for CP (195 [120] pmol/L vs. 230 [90] pmol/L; P=0.01), but not significantly so for PP (178 [106] pmol/L vs. 194 [120] pmol/L; P=0.099). There were no significant differences between the two catheter lengths.
CONCLUSIONS: Postprandial glycemic excursions were lower on Day 3 of catheter wear time, but there were no differences between PPs and CPs. These findings support the proposal that catheter wear time plays an important role in insulin absorption.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23650901      PMCID: PMC3709595          DOI: 10.1089/dia.2013.0016

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  10 in total

1.  AP@home: a novel European approach to bring the artificial pancreas home.

Authors:  Lutz Heinemann; Carsten Benesch; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

2.  Accelerating and improving the consistency of rapid-acting analog insulin absorption and action for both subcutaneous injection and continuous subcutaneous infusion using recombinant human hyaluronidase.

Authors:  Douglas B Muchmore; Daniel E Vaughn
Journal:  J Diabetes Sci Technol       Date:  2012-07-01

Review 3.  Advances in pump technology: insulin patch pumps, combined pumps and glucose sensors, and implanted pumps.

Authors:  P Schaepelynck; P Darmon; L Molines; M F Jannot-Lamotte; C Treglia; D Raccah
Journal:  Diabetes Metab       Date:  2011-12       Impact factor: 6.041

4.  Insulin patch pumps: their development and future in closed-loop systems.

Authors:  Henry Anhalt; Nancy J V Bohannon
Journal:  Diabetes Technol Ther       Date:  2010-06       Impact factor: 6.118

5.  Pilot study for assessment of optimal frequency for changing catheters in insulin pump therapy-trouble starts on day 3.

Authors:  Volkmar Schmid; Cloth Hohberg; Marcus Borchert; Thomas Forst; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2010-07-01

6.  Effect of age of infusion site and type of rapid-acting analog on pharmacodynamic parameters of insulin boluses in youth with type 1 diabetes receiving insulin pump therapy.

Authors:  Karena L Swan; James D Dziura; Garry M Steil; Gayane R Voskanyan; Kristin A Sikes; Amy T Steffen; Melody L Martin; William V Tamborlane; Stuart A Weinzimer
Journal:  Diabetes Care       Date:  2008-11-18       Impact factor: 17.152

7.  Effect of insulin catheter wear-time on subcutaneous adipose tissue blood flow and insulin absorption in humans.

Authors:  Trine Schnedler Clausen; Peter Kaastrup; Bente Stallknecht
Journal:  Diabetes Technol Ther       Date:  2009-09       Impact factor: 6.118

Review 8.  Insulin infusion set: the Achilles heel of continuous subcutaneous insulin infusion.

Authors:  Lutz Heinemann; Lars Krinelke
Journal:  J Diabetes Sci Technol       Date:  2012-07-01

9.  The OmniPod Insulin Management System: the latest innovation in insulin pump therapy.

Authors:  Howard C Zisser
Journal:  Diabetes Ther       Date:  2010-09-13       Impact factor: 2.945

10.  Use of the site of subcutaneous insulin administration for the measurement of glucose in patients with type 1 diabetes.

Authors:  Stefan Lindpointner; Stefan Korsatko; Gerd Köhler; Hans Köhler; Roland Schaller; Ruthy Kaidar; Ofer Yodfat; Lukas Schaupp; Martin Ellmerer; Thomas R Pieber; Werner Regittnig
Journal:  Diabetes Care       Date:  2009-12-29       Impact factor: 17.152

  10 in total
  6 in total

Review 1.  AP@home: The Artificial Pancreas Is Now at Home.

Authors:  Lutz Heinemann; Carsten Benesch; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2016-06-28

2.  Patch Pumps: Are They All the Same?

Authors:  Lutz Heinemann; Delia Waldenmaier; Bernd Kulzer; Ralph Ziegler; Barry Ginsberg; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-08-22

3.  Intradermal insulin infusion achieves faster insulin action than subcutaneous infusion for 3-day wear.

Authors:  Christopher James Rini; Elaine McVey; Diane Sutter; Stephen Keith; Heinz-Joerg Kurth; Leszek Nosek; Christoph Kapitza; Kerstin Rebrin; Laurence Hirsch; Ronald J Pettis
Journal:  Drug Deliv Transl Res       Date:  2015-08       Impact factor: 4.617

Review 4.  Continuous subcutaneous insulin infusion in diabetes: patient populations, safety, efficacy, and pharmacoeconomics.

Authors:  Paolo Pozzilli; Tadej Battelino; Thomas Danne; Roman Hovorka; Przemyslawa Jarosz-Chobot; Eric Renard
Journal:  Diabetes Metab Res Rev       Date:  2015-06-22       Impact factor: 4.876

Review 5.  Recommendations for Using Real-Time Continuous Glucose Monitoring (rtCGM) Data for Insulin Adjustments in Type 1 Diabetes.

Authors:  Jeremy Pettus; Steven V Edelman
Journal:  J Diabetes Sci Technol       Date:  2016-08-20

Review 6.  Use of fast-acting insulin aspart in insulin pump therapy in clinical practice.

Authors:  Mark Evans; Antonio Ceriello; Thomas Danne; Christophe De Block; J Hans DeVries; Marcus Lind; Chantal Mathieu; Kirsten Nørgaard; Eric Renard; Emma G Wilmot
Journal:  Diabetes Obes Metab       Date:  2019-06-19       Impact factor: 6.577