Literature DB >> 23645529

Advanced techniques for chronic lead extraction: heading from the laser towards the evolution system.

Patrizio Mazzone1, Dimitris Tsiachris, Alessandra Marzi, Giuseppe Ciconte, Gabriele Paglino, Nicoleta Sora, Simone Gulletta, Pasquale Vergara, Paolo Della Bella.   

Abstract

AIM: The evolution mechanical dilator sheath has been reported to be an effective tool for chronic lead extraction (LE). We examined safety and efficacy of evolution system as compared with laser system. METHODS AND
RESULTS: From 2005 to 2009, all extractions requiring the use of a powered sheath were performed using the excimer laser system (n = 73). Since 2009, laser system was no longer available and the evolution system was introduced as the first-line method for powered extraction (n = 48). All procedures were performed by a single first operator. Success and complications were defined according to the current guidelines. Patients of the evolution group compared with those of the laser group had a greater number of extracted leads per patient (2.77 vs. 2.4, P = 0.049) and a longer implant duration (101.1 vs. 62.4 months, P = 0.001). Additional use of snare was required in 27.1% of the evolution group and 8.2% of the laser group (P = 0.005). Complete procedural success was achieved in 91.7% of the evolution group and 97.3% of the laser group (P = 0.16). There was also no difference between evolution and laser groups in clinical success (97.9 vs. 98.6%, P = 0.76), as well as regarding major (4.2 vs. 2.7%, P = 0.66) or minor complications (4.2 vs. 5.5%, P = 0.76).
CONCLUSION: Use of the recently introduced evolution system for LE exhibit acceptably high levels of safety, as well as of procedural and clinical success, although additional use of snare was required more frequently in the evolution compared with the laser group.

Entities:  

Keywords:  Lead extraction; Powered sheath; Snare

Mesh:

Year:  2013        PMID: 23645529     DOI: 10.1093/europace/eut126

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  10 in total

1.  Re-evaluation of transvenous lead extraction with modified standard technique: a prospective study in 229 patients.

Authors:  Xian-Ming Chu; Xue-Bin Li; Ping Zhang; Long Wang; Ding Li; Jiang-Bo Duan; Bing Li; Ji-Hong Guo
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2013-10-20

2.  Transvenous lead extraction outcomes using a novel hand-powered bidirectional rotational sheath as a first-line extraction tool in a low-volume centre.

Authors:  Jus Ksela; Jan Prevolnik; Mark Racman
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-04-08

3.  Cardiac implantable electronic device lead extraction using the lead-locking device system: keeping it simple, safe, and inexpensive with mechanical tools and local anesthesia.

Authors:  Antonis S Manolis; Georgios Georgiopoulos; Sofia Metaxa; Spyridon Koulouris; Dimitris Tsiachris
Journal:  Anatol J Cardiol       Date:  2017-08-11       Impact factor: 1.596

4.  Editorial: What makes transvenous extraction more difficult?

Authors:  Hideo Okamura
Journal:  J Cardiol Cases       Date:  2015-12-04

5.  Use of the new rotating dilator sheath TightRail™ for lead extraction: A bicentric experience.

Authors:  Patrizio Mazzone; Francesco Melillo; Andrea Radinovic; Alessandra Marzi; Gabriele Paglino; Paolo Della Bella; Giosuè Mascioli
Journal:  J Arrhythm       Date:  2020-02-14

6.  The COVID-19 challenge to cardiac electrophysiologists: optimizing resources at a referral center.

Authors:  Patrizio Mazzone; Giovanni Peretto; Andrea Radinovic; Luca Rosario Limite; Alessandra Marzi; Simone Sala; Manuela Cireddu; Pasquale Vegara; Francesca Baratto; Gabriele Paglino; Giuseppe D'Angelo; Lorenzo Cianfanelli; Savino Altizio; Felicia Lipartiti; Antonio Frontera; Caterina Bisceglia; Simone Gulletta; Paolo Della Bella
Journal:  J Interv Card Electrophysiol       Date:  2020-05-18       Impact factor: 1.900

7.  Comparison between laser sheaths, femoral approach and rotating mechanical sheaths for lead extraction.

Authors:  F A Bracke; N Rademakers; N Verberkmoes; M Van 't Veer; B M van Gelder
Journal:  Neth Heart J       Date:  2021-12-21       Impact factor: 2.854

8.  Transvenous extraction of pacemaker and implantable cardioverter defibrillator leads using Evolution® mechanical dilator sheath: a single center confirmatory experience.

Authors:  Uğur Kocabaş; Hamza Duygu; Nihan Kahya Eren; Zehra İlke Akyıldız; Serhan Özyıldırım; Selcen Yakar Tülüce; Tuncay Kırış; Cem Nazlı
Journal:  Springerplus       Date:  2016-03-22

9.  Risk Factors Predicting Complications of Transvenous Lead Extraction.

Authors:  Wojciech Jacheć; Anna Polewczyk; Maciej Polewczyk; Andrzej Tomasik; Marianna Janion; Andrzej Kutarski
Journal:  Biomed Res Int       Date:  2018-12-18       Impact factor: 3.411

10.  Efficacy and safety of transvenous lead extraction using a liberal combined superior and femoral approach.

Authors:  Sing-Chien Yap; Rohit E Bhagwandien; Dominic A M J Theuns; Yunus Emre Yasar; John de Heide; Mark G Hoogendijk; Charles Kik; Tamas Szili-Torok
Journal:  J Interv Card Electrophysiol       Date:  2020-10-07       Impact factor: 1.900