| Literature DB >> 27066369 |
Uğur Kocabaş1, Hamza Duygu2, Nihan Kahya Eren1, Zehra İlke Akyıldız1, Serhan Özyıldırım3, Selcen Yakar Tülüce1, Tuncay Kırış1, Cem Nazlı1.
Abstract
OBJECTIVE: In recent years there has been an increase in clinical situations requiring lead extraction procedures of implanted cardiac devices. In our clinic, extraction procedures are performed with Evolution® mechanical lead extraction system. In this manuscript we aimed to evaluate our lead extraction procedures.Entities:
Keywords: Defibrillator; Extraction; Lead; Pacemaker
Year: 2016 PMID: 27066369 PMCID: PMC4803708 DOI: 10.1186/s40064-016-1987-x
Source DB: PubMed Journal: Springerplus ISSN: 2193-1801
Fig. 1Combined use of two mechanical dilator sheaths for separation of extremely adhered leads
Fig. 2The extracted lead
Baseline characteristics of the study population
| Age, mean ± SD | 61.5 ± 18.5 median 67 (23–85) |
| Gender, n (%) | 30 Male (73.2 %) 11 female (26.8 %) |
| Indication of initial device implantation, n (%) | |
| Sick sinus syndrome | 16 (39 %) |
| Complete AV block | 11 (26.8 %) |
| Ventricular arrhythmias, heart failure | 14 (34.2 %) |
| Implanted device, n (%) | |
| Single chamber PM,VVI/VDD | 7 (17 %), 6 (14.8 %) |
| Dual chamber PM | 14 (34 %) |
| Single chamber ICD | 6 (14.8 %) |
| Dual chamber ICD | 7 (17 %) |
| Biventricular ICD | 1 (2.4 %) |
| Prevalence of diseases, n (%) | |
| Heart failure | 12 (29.2 %) |
| Coronary artery disease | 13 (31.7 %) |
| Valvular disease | 2 (4.8 %) |
| Hypertension | 19 (46.3 %) |
| Diabetes | 7 (17 %) |
| Congenital heart disease | 3 (7.2 %) |
| Chronic obstructive lung disease | 2 (4.8 %) |
| Chronic renal failure | 2 (4.8 %) |
| Hypertrophic cardiomyopathy | 1 (2.4 %) |
| Arrythmogenic right ventricular disease | 1 (2.4 %) |
| Cirrohis of liver | 1 (2.4 %) |
| Brugada syndrome | 1 (2.4 %) |
Procedure and implanted device characteristics
| Implanted device, n (%) | Pacemaker 27 (65.8 %) ICD 14 (34.2 %) |
| Mean duration after initial implantation | 88.4 ± 62.5 months median 84 (6–240 months) |
| Cardiac leads per patient, total n (%) | 67 |
| One lead | 17 (41.4 %) |
| Two leads | 23(56.2 %) |
| Four leads | 1 (2.4 %) |
| Lead type, n (%) | 67 |
| Atrial | 22 (32.8 %) |
| Ventricular | 30 (44.2 %) |
| Defibrillation coil | 14 (21.5 %) |
| Coronary sinus lead | 1 (1.5 %) |
| Fixation mechanism, n (%) | |
| Active fixation | 11 (16.4 %) |
| Passive fixation | 56 (83.6 %) |
| Indication for removal, n (%) | |
| PM pocket dekubitis, infection | 29 (70.8 %) |
| Lead dysfunction | 11 (26.8 %) |
| Subclavian vein thrombosis | 1 (2.4 %) |
Procedure and procedure related complications
| Clinical success, n (%) | 40 (97.5 %) |
| Procedural success, n (%) | 39 (95.1 %) |
| Success rate without snare use, n (%) | 35 (85.3 %) |
| Major complications | 2 (4.8 %) |
| Mortality after procedure | 1 (2.4 %) |
| Right ventricular rupture caused by temporary transvenous pacemaker | 1 (2.4 %) |
| Cardiovascular surgery due to failure of procedure | 1 (2.4 %) |
| Minor complications | 5 (12.2 %) |
| Vascular repair operation by vascular surgeons | 2 (4.8 %) |
| Pacemaker pocket hematoma requiring surgery | 1 (2.4 %) |
| Pericardial effusion | 1 (2.4 %) |
| Pleural effusion not requiring drainage | 1 (2.4 %) |
| Snare need for lead or its components | 5 (12.2 %) |
| Device implantation indication after procedure | 31 (75.6 %) |
| Patients with device implantation after procedure | 25 (61.1 %) |
Main characteristics and results from different centers using mechanical extraction devices and our center’s results
| Trial | Patient number | Age of patients (years) | Mean implantation duration (months) | Clinical success rate (%) | Procedural success rate (%) | Major complication rate (%) | Minor Complication rate (%) |
|---|---|---|---|---|---|---|---|
| Husseina et al. ( | 29 | 65 ± 19 | 65 (12–409) | 100 | 86 | 0 | 0 |
| Otoa et al. ( | 23 | 58 ± 14 | 74 (25–180) | 100 | 82 | 0 | 4.3 |
| Otoa et al. ( | 66 | 55 ± 11 | 85 (22–240) | 98.5 | 87.9 | 1.5 | 3 |
| Mazzonea et al. ( | 48 | 65 ± 14 | 101 ± 66.4 | 97.9 | 91.7 | 4.2 | 2.7 |
| Starcka et al. ( | 122 | 60.4 | 69.6 (1–384) | 97 | 88.9 | 2 | 3 |
| Kocabasa et al. ( | 20 | 61 ± 19 | 97 (8–204) | 95 | 95 | 5 | 25 |
| Aytemirb et al. ( | 23 | 59 ± 13 | 72 (8–216) | 100 | 95.7 | 0 | 0 |
| This studya | 41 | 61 ± 18 | 88 (6–240) | 97.5 | 95.1 | 4.8 | 12.2 |
aEvolution mechanical lead extraction system
bTightRail mechanical lead extraction system