Effects of battery type and age on performance of rechargeable laryngoscopes.

Andrew D Milne1, Claire A Brousseau.   

Abstract

Optimal visualization of the glottis can be crucial to successful laryngoscopy. Limited information has been published on the light intensity delivered from laryngoscopes powered by rechargeable batteries. In this study the laryngoscope light intensity delivered from 10 nickel metal hydride (NiMH), 7 nickel cadmium (NiCAD), and 2 lithium (LI) batteries with 3-5 or more years of clinical usage were tested in comparison to 5 new NiMH batteries. Each battery was charged in a new laryngoscope handle and recharging unit for 24 h before testing. Light intensity (lux) from the bulb in the laryngoscope handle was recorded at 3-min intervals under continuous loading until battery depletion. The mean times ±1 standard deviation (SD) to minimum acceptable light output (2,000 lux from the handle) were new NiMH 70 ± 1 min, 3-year-old NiMH 96 ± 2 min, 5+ year-old NiCAD 45 ± 22 min, and 5+ year-old LI 117 ± 4 min. There were significant differences in the time to minimum light intensity among all groups (p = 0.00-0.04). All new and used batteries exceeded the minimum ISO standard of light intensity for more than 10 min. These data demonstrate that rechargeable laryngoscope batteries can safely be used for several years before requiring replacement.

Entities:  

Mesh:

Year:  2013        PMID: 23640370     DOI: 10.1007/s00540-013-1618-4

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  6 in total

1.  The effect of mechanical cleaning and thermal disinfection on light intensity provided by fibrelight Macintosh laryngoscopes.

Authors:  M J L Bucx; H M De Gast; J Veldhuis; L H Hassing; A Meulemans; A Kammeyer
Journal:  Anaesthesia       Date:  2003-05       Impact factor: 6.955

2.  Shelf life of rechargeable laryngoscope batteries without recharging.

Authors:  Andrew D Milne; Paul A Brousseau
Journal:  Can J Anaesth       Date:  2012-05-30       Impact factor: 5.063

3.  Changes in the light intensity of the fiberoptic laryngoscope blade by steam sterilization.

Authors:  Tomoki Nishiyama
Journal:  Anesth Analg       Date:  2007-04       Impact factor: 5.108

4.  Illumination of bulb-on-blade laryngoscopes in the out-of-hospital setting.

Authors:  Ka Wai Cheung; George Kovacs; J Adam Law; Paul Brousseau; William Hill
Journal:  Acad Emerg Med       Date:  2007-05       Impact factor: 3.451

5.  The effects of lead sulfate on new sealed lead acid batteries.

Authors:  M J Cleland; J P Maloney; B H Rowe
Journal:  J Emerg Med       Date:  2000-04       Impact factor: 1.484

6.  Premature failure of battery-powered syringe pumps.

Authors:  A Tatman; H Brunner; M Stokes
Journal:  Anaesthesia       Date:  1996-11       Impact factor: 6.955

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.