Illumination of bulb-on-blade laryngoscopes in the out-of-hospital setting.

Ka Wai Cheung1, George Kovacs, J Adam Law, Paul Brousseau, William Hill.   

Abstract

OBJECTIVES: To determine what percentage of out-of-hospital laryngoscopes meet a predetermined minimal illumination criterion and what factors may be altered to improve illumination.
METHODS: This was an observational study of the illumination of laryngoscopes currently in use by the Emergency Health Services of Nova Scotia. Each laryngoscope was measured at baseline. This illumination was compared with the illumination after replacement with new batteries, replacement with a new bulb, replacement with new batteries and a new bulb together, and attachment of a disposable blade. The percentage of laryngoscopes that met a previously defined minimal brightness criterion was determined.
RESULTS: Fifty-one laryngoscopes were measured. These laryngoscopes had a mean (+/-SD) illumination of 624 (+/-297) lux at baseline. Laryngoscope illumination increased after replacement with new batteries by 168 lux (95% confidence interval [CI] = 121 to 216), replacement with a new bulb by 679 lux (95% CI = 524 to 834), replacement with new batteries and a new bulb by 937 lux (95% CI = 770 to 1,104), and attachment of a disposable blade by 2,401 lux (95% CI = 2,075 to 2,740). Fourteen percent of laryngoscopes (7/51) at baseline met the minimal illumination criterion.
CONCLUSIONS: Only a small percentage of out-of-hospital laryngoscopes met the minimal illumination criterion. There was a statistically significant increase in illumination after replacement with new batteries, replacement with a new bulb, replacement with new batteries and a new bulb, or attachment of a disposable blade. Optimal changing of lightbulbs and batteries in the out-of-hospital setting will have to be more clearly defined.

Entities:  

Mesh:

Year:  2007        PMID: 17392520     DOI: 10.1197/j.aem.2007.01.004

Source DB:  PubMed          Journal:  Acad Emerg Med        ISSN: 1069-6563            Impact factor:   3.451


  1 in total

1.  Effects of battery type and age on performance of rechargeable laryngoscopes.

Authors:  Andrew D Milne; Claire A Brousseau
Journal:  J Anesth       Date:  2013-05-03       Impact factor: 2.078

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.