| Literature DB >> 8943596 |
A Tatman1, H Brunner, M Stokes.
Abstract
Failure of battery-powered equipment during the interhospital transfer of patients is potentially life threatening. The time to failure of 60 fully-charged identical syringe pumps (20 from each batch purchased in 1992, 1994 and 1995) was measured. Older pumps were associated with less predictable charge capacity, with 40% of the 1992 pumps failing within 60 min. The premature failure of these pumps is most likely due to poor battery care. It is unsafe to assume that a fully-charged, battery-powered pump will continue to function throughout a long transfer. The routine carriage of spare pumps or a backup power supply is recommended.Entities:
Mesh:
Year: 1996 PMID: 8943596 DOI: 10.1111/j.1365-2044.1996.tb15000.x