A real-time regional adaptive exposure method for saving dose-area product in x-ray fluoroscopy.

Steve Burion1, Michael A Speidel, Tobias Funk.   

Abstract

PURPOSE: Reduction of radiation dose in x-ray imaging has been recognized as a high priority in the medical community. Here the authors show that a regional adaptive exposure method can reduce dose-area product (DAP) in x-ray fluoroscopy. The authors' method is particularly geared toward providing dose savings for the pediatric population.
METHODS: The scanning beam digital x-ray system uses a large-area x-ray source with 8000 focal spots in combination with a small photon-counting detector. An imaging frame is obtained by acquiring and reconstructing up to 8000 detector images, each viewing only a small portion of the patient. Regional adaptive exposure was implemented by varying the exposure of the detector images depending on the local opacity of the object. A family of phantoms ranging in size from infant to obese adult was imaged in anteroposterior view with and without adaptive exposure. The DAP delivered to each phantom was measured in each case, and noise performance was compared by generating noise arrays to represent regional noise in the images. These noise arrays were generated by dividing the image into regions of about 6 mm(2), calculating the relative noise in each region, and placing the relative noise value of each region in a one-dimensional array (noise array) sorted from highest to lowest. Dose-area product savings were calculated as the difference between the ratio of DAP with adaptive exposure to DAP without adaptive exposure. The authors modified this value by a correction factor that matches the noise arrays where relative noise is the highest to report a final dose-area product savings.
RESULTS: The average dose-area product saving across the phantom family was (42 ± 8)% with the highest dose-area product saving in the child-sized phantom (50%) and the lowest in the phantom mimicking an obese adult (23%).
CONCLUSIONS: Phantom measurements indicate that a regional adaptive exposure method can produce large DAP savings without compromising the noise performance in the image regions with highest noise.

Entities:  

Mesh:

Year:  2013        PMID: 23635281      PMCID: PMC3651207          DOI: 10.1118/1.4801908

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  X-ray beam equalization: feasibility and performance of an automated prototype system in a phantom and swine.

Authors:  S Molloi; A Van Drie; F Wang
Journal:  Radiology       Date:  2001-12       Impact factor: 11.105

2.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

3.  Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Timothy D Betts; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

4.  Guidelines for patient radiation dose management.

Authors:  Michael S Stecker; Stephen Balter; Richard B Towbin; Donald L Miller; Eliseo Vañó; Gabriel Bartal; J Fritz Angle; Christine P Chao; Alan M Cohen; Robert G Dixon; Kathleen Gross; George G Hartnell; Beth Schueler; John D Statler; Thierry de Baère; John F Cardella
Journal:  J Vasc Interv Radiol       Date:  2009-07       Impact factor: 3.464

5.  Filter wheel equalization in chest radiography: demonstration with a prototype system.

Authors:  J M Boone; J Duryea; E H Moore
Journal:  Radiology       Date:  1995-09       Impact factor: 11.105

6.  Pediatric interventional radiography equipment: safety considerations.

Authors:  Keith J Strauss
Journal:  Pediatr Radiol       Date:  2006-09

Review 7.  Management of pediatric radiation dose using Philips fluoroscopy systems DoseWise: perfect image, perfect sense.

Authors:  Dick Stueve
Journal:  Pediatr Radiol       Date:  2006-09

8.  Solid cancer incidence in atomic bomb survivors exposed in utero or as young children.

Authors:  Dale L Preston; Harry Cullings; Akihiko Suyama; Sachiyo Funamoto; Nobuo Nishi; Midori Soda; Kiyohiko Mabuchi; Kazunori Kodama; Fumiyoshi Kasagi; Roy E Shore
Journal:  J Natl Cancer Inst       Date:  2008-03-11       Impact factor: 13.506

9.  Radiation doses in interventional radiology procedures: the RAD-IR study: part I: overall measures of dose.

Authors:  Donald L Miller; Stephen Balter; Patricia E Cole; Hollington T Lu; Beth A Schueler; Michael Geisinger; Alejandro Berenstein; Robin Albert; Jeffrey D Georgia; Patrick T Noonan; John F Cardella; James St George; Eric J Russell; Tim W Malisch; Robert L Vogelzang; George L Miller; Jon Anderson
Journal:  J Vasc Interv Radiol       Date:  2003-06       Impact factor: 3.464

10.  X-ray dose reduction through adaptive exposure in fluoroscopic imaging.

Authors:  Steve Burion; Tobias Funk
Journal:  J Vis Exp       Date:  2011-09-11       Impact factor: 1.355

View more
  3 in total

1.  Dynamic electronic collimation method for 3-D catheter tracking on a scanning-beam digital x-ray system.

Authors:  David A P Dunkerley; Jordan M Slagowski; Tobias Funk; Michael A Speidel
Journal:  J Med Imaging (Bellingham)       Date:  2017-04-18

2.  Monte Carlo simulation of inverse geometry x-ray fluoroscopy using a modified MC-GPU framework.

Authors:  David A P Dunkerley; Michael T Tomkowiak; Jordan M Slagowski; Bradley P McCabe; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-02-21

3.  Method for dose-reduced 3D catheter tracking on a scanning-beam digital x-ray system using dynamic electronic collimation.

Authors:  David A P Dunkerley; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.