Literature DB >> 27375314

Method for dose-reduced 3D catheter tracking on a scanning-beam digital x-ray system using dynamic electronic collimation.

David A P Dunkerley1, Tobias Funk2, Michael A Speidel3.   

Abstract

Scanning-beam digital x-ray (SBDX) is an inverse geometry x-ray fluoroscopy system capable of tomosynthesis-based 3D catheter tracking. This work proposes a method of dose-reduced 3D tracking using dynamic electronic collimation (DEC) of the SBDX scanning x-ray tube. Positions in the 2D focal spot array are selectively activated to create a region-of-interest (ROI) x-ray field around the tracked catheter. The ROI position is updated for each frame based on a motion vector calculated from the two most recent 3D tracking results. The technique was evaluated with SBDX data acquired as a catheter tip inside a chest phantom was pulled along a 3D trajectory. DEC scans were retrospectively generated from the detector images stored for each focal spot position. DEC imaging of a catheter tip in a volume measuring 11.4 cm across at isocenter required 340 active focal spots per frame, versus 4473 spots in full-FOV mode. The dose-area-product (DAP) and peak skin dose (PSD) for DEC versus full field-of-view (FOV) scanning were calculated using an SBDX Monte Carlo simulation code. DAP was reduced to 7.4% to 8.4% of the full-FOV value, consistent with the relative number of active focal spots (7.6%). For image sequences with a moving catheter, PSD was 33.6% to 34.8% of the full-FOV value. The root-mean-squared-deviation between DEC-based 3D tracking coordinates and full-FOV 3D tracking coordinates was less than 0.1 mm. The 3D distance between the tracked tip and the sheath centerline averaged 0.75 mm. Dynamic electronic collimation can reduce dose with minimal change in tracking performance.

Entities:  

Keywords:  X-ray fluoroscopy; catheter tracking; collimation; inverse geometry; scanning-beam digital x-ray

Year:  2016        PMID: 27375314      PMCID: PMC4925103          DOI: 10.1117/12.2216892

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  16 in total

Review 1.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

Review 2.  Radiofrequency ablation of cardiac arrhythmias: past, present and future.

Authors:  J P Joseph; K Rajappan
Journal:  QJM       Date:  2011-11-12

3.  Feasibility of low-dose single-view 3D fiducial tracking concurrent with external beam delivery.

Authors:  Michael A Speidel; Brian P Wilfley; Annie Hsu; Dimitre Hristov
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

Review 4.  Structural heart intervention. Beyond transcatheter valve therapy.

Authors:  Igor F Palacios; Dabit Arzamendi
Journal:  Rev Esp Cardiol (Engl Ed)       Date:  2012-03-20

5.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

6.  Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Timothy D Betts; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

7.  Predicting error in rigid-body point-based registration.

Authors:  J M Fitzpatrick; J B West; C R Maurer
Journal:  IEEE Trans Med Imaging       Date:  1998-10       Impact factor: 10.048

8.  Monte Carlo simulation of inverse geometry x-ray fluoroscopy using a modified MC-GPU framework.

Authors:  David A P Dunkerley; Michael T Tomkowiak; Jordan M Slagowski; Bradley P McCabe; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-02-21

9.  Computed tomography-fluoroscopy image integration-guided catheter ablation of atrial fibrillation.

Authors:  Jasbir Sra; Girish Narayan; David Krum; Angela Malloy; Ryan Cooley; Atul Bhatia; Anwer Dhala; Zalmen Blanck; Vikram Nangia; Masood Akhtar
Journal:  J Cardiovasc Electrophysiol       Date:  2007-01-30

10.  Detector, collimator and real-time reconstructor for a new scanning-beam digital x-ray (SBDX) prototype.

Authors:  Michael A Speidel; Michael T Tomkowiak; Amish N Raval; David A P Dunkerley; Jordan M Slagowski; Paul Kahn; Jamie Ku; Tobias Funk
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015
View more
  1 in total

1.  Improving the cardiac cath-lab interventional imaging eco-system.

Authors:  Benjamin R Ciske; Michael A Speidel; Amish N Raval
Journal:  Transl Pediatr       Date:  2018-01
  1 in total