Mechanical stability and clinical applicability assessment of novel orthodontic mini-implant design.

Ha Na Song1, Christine Hong, Robert Banh, Tania Ohebsion, Greg Asatrian, Ho-Yin Leung, Benjamin M Wu, Won Moon.   

Abstract

OBJECTIVE: To compare the stability and clinical applicability of a novel orthodontic mini-implant design (N2) with the most widely used commercially available (CA) design.
MATERIALS AND METHODS: Two groups of mini-implants were tested: a CA design (1.5-mm diameter, 6-mm length) and N2 (3-mm diameter, 2-mm length, tapered shape). Implants were inserted in bone blocks of cortical bone simulation with varying densities (20 pounds per cubic foot [pcf], 30 pcf, and 40 pcf). A torque test was used to measure maximum insertion torque (MIT) and maximum removal torque (MRT). Compression and tension force vectors were applied at angles of 10°, 20°, 30°, and 40° using customized load pins to determine primary stability.
RESULTS: Mean MIT and MRT were higher in the N2 than the CA design at all three cortical bone densities except MRT in 20 pcf bone (not statistically significant). The mean compression force required to displace the N2 at all distances and angulations was greater for the N2 than the CA design. At all displacement distances, the highest mean tension force required for N2 displacement was at 10° angulation, whereas at 30° and 40°, the mean tension force required to displace the CA design was greater.
CONCLUSIONS: The primary stability of the N2 is superior to that of the CA design and is promising for both orthodontic and orthopedic clinical applicability, especially under compression force. The short length of the N2 reduces risk of damage to anatomic structures and root proximity during placement and orthodontic treatment. The stability of the N2 may be compromised in areas of high bone density and highly angulated tension force.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23627420      PMCID: PMC8744521          DOI: 10.2319/111412-876.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  28 in total

1.  Mini-implants: Where are we?

Authors:  Birte Melsen
Journal:  J Clin Orthod       Date:  2005-09

2.  Mandibular molar intrusion with miniscrew anchorage.

Authors:  Seong Min Bae; Hee Moon Kyung
Journal:  J Clin Orthod       Date:  2006-02

3.  Bone density measurement in interdental areas with simulated placement of orthodontic miniscrew implants.

Authors:  Jin-Hugh Choi; Chang-Hun Park; Seung-Won Yi; Hoi-Jeong Lim; Hyeon-Shik Hwang
Journal:  Am J Orthod Dentofacial Orthop       Date:  2009-12       Impact factor: 2.650

4.  Skeletal anchorage system for open-bite correction.

Authors:  M Umemori; J Sugawara; H Mitani; H Nagasaka; H Kawamura
Journal:  Am J Orthod Dentofacial Orthop       Date:  1999-02       Impact factor: 2.650

5.  Miniscrew-assisted nonsurgical palatal expansion before orthognathic surgery for a patient with severe mandibular prognathism.

Authors:  Kee-Joon Lee; Young-Chel Park; Joo-Young Park; Woo-Sang Hwang
Journal:  Am J Orthod Dentofacial Orthop       Date:  2010-06       Impact factor: 2.650

6.  Patient cooperation in wearing orthodontic headgear.

Authors:  E J Clemmer; E W Hayes
Journal:  Am J Orthod       Date:  1979-05

7.  Timely relocation of mini-implants for uninterrupted full-arch distalization.

Authors:  Kyu-Rhim Chung; HyeRan Choo; Seong-Hun Kim; Peter Ngan
Journal:  Am J Orthod Dentofacial Orthop       Date:  2010-12       Impact factor: 2.650

8.  Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2.

Authors:  Christine Hong; Peter Truong; Ha Na Song; Benjamin M Wu; Won Moon
Journal:  Angle Orthod       Date:  2011-06-29       Impact factor: 2.079

9.  Comparing subjective and objective measures of headgear compliance.

Authors:  Annemieke Bos; Cornelis J Kleverlaan; Johan Hoogstraten; Birte Prahl-Andersen; Rein Kuitert
Journal:  Am J Orthod Dentofacial Orthop       Date:  2007-12       Impact factor: 2.650

10.  Density of the alveolar and basal bones of the maxilla and the mandible.

Authors:  Hyo-Sang Park; Youn-Ju Lee; Seong-Hwa Jeong; Tae-Geon Kwon
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-01       Impact factor: 2.650

View more
  4 in total

1.  Effects of monocortical and bicortical mini-implant anchorage on bone-borne palatal expansion using finite element analysis.

Authors:  Robert J Lee; Won Moon; Christine Hong
Journal:  Am J Orthod Dentofacial Orthop       Date:  2017-05       Impact factor: 2.650

Review 2.  Two-dimensional radiographs versus cone-beam computed tomography in planning mini-implant placement: A systematic review.

Authors:  Gabriela-Franco-da Rosa Caetano; Mariana-Quirino-Silveira Soares; Luciana-Butini Oliveira; José-Luiz-Cintra Junqueira; Monikelly-do Carmo-Chagas Nascimento
Journal:  J Clin Exp Dent       Date:  2022-08-01

3.  The efficacy of maxillary protraction protocols with the micro-implant-assisted rapid palatal expander (MARPE) and the novel N2 mini-implant-a finite element study.

Authors:  Won Moon; Kimberley W Wu; Matthew MacGinnis; Jay Sung; Howard Chu; George Youssef; Andre Machado
Journal:  Prog Orthod       Date:  2015-06-04       Impact factor: 2.750

4.  Comparison of mechanical and biological properties of zirconia and titanium alloy orthodontic micro-implants.

Authors:  Hae Won Choi; Young Seok Park; Shin Hye Chung; Min Ho Jung; Won Moon; Sang Hoon Rhee
Journal:  Korean J Orthod       Date:  2017-05-26       Impact factor: 1.372

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.