Essential opaque bubble layer elimination with novel LASIK flap settings in the FS200 Femtosecond Laser.

A John Kanellopoulos1, George Asimellis.   

Abstract

BACKGROUND: The purpose of this study is to evaluate the extent and incidence of opaque bubble layer (OBL) using laser-assisted in situ keratomileusis (LASIK) flaps created with the Alcon/WaveLight® FS200 femtosecond laser as a result of a recent change in flap programming parameters aiming to reduce further the incidence and extent of OBL.
METHODS: Intraoperative digital images of flaps from 36 consecutive patients (72 eyes) subjected to bilateral femtosecond-assisted LASIK were analyzed using a proprietary computerized technique. The incidence and extent of OBL was measured and reported as a percentage of the entire flap area. Flap creation was performed with a 1.7 mm wide canal, implemented as an updated design intended to reduce the extent of OBL (group A). The same OBL parameters were investigated and compared in an age-matched and procedure-matched patients in whom the previous standard setting of a 1.3 mm wide canal was implemented (group B).
RESULTS: In group A, the average extent of OBL was 3.69% of the flap area (range 0%-11.34%). In group B, the respective values were 6.06% (range 0%-20.24%). We found the difference to be statistically significant (one-tailed P = 0.00452).
CONCLUSION: This study suggests that there is a significant reduction in the incidence and extent of OBL when novel LASIK flap ventilation canal parameters of width and spot line separation are used.

Entities:  

Keywords:  Alcon/WaveLight FS200; bladeless laser-assisted in situ keratomileusis; femtosecond laser flap; opaque bubble layer; spot line separation

Year:  2013        PMID: 23620658      PMCID: PMC3633575          DOI: 10.2147/OPTH.S43723

Source DB:  PubMed          Journal:  Clin Ophthalmol        ISSN: 1177-5467


Introduction

Formation of opaque bubble layer (OBL) during creation of a laser-assisted in situ keratomileusis (LASIK) flap is a finding unique to use of femtosecond laser.1 OBL occurs along the lamellar dissection plane during the flap creation,2 and can be described simply as temporary stromal infiltration by compressed air generated by the intracorneal femtosecond laser action, that cannot escape.3 Although no serious complications have been reported as a result of its occurrence, OBL may temporarily obscure the pupil image used by most excimer laser trackers, in the subsequent excimer ablation. It may also interfere with reading of architectural landmarks on the iris used by some excimer laser trackers to compensate for cyclorotation, and may even obscure the patient’s fixation target. The purpose of this study was to compare quantitative differences in the presence and extent of OBL in flaps created using the FS200 femtosecond laser with a recently introduced wider venting canal design, and tighter line separation parameters, versus the predecessor design.

Materials and methods

This study was approved by the ethics committee of our institution, and was consistent with the tenets of the Declaration of Helsinki. Informed consent was obtained from each subject at the time of the LASIK intervention or the first clinical screening visit. The study was conducted in our clinical LASIK practice and recruited consecutive patients.

Patient inclusion criteria

The study group consisted of 36 consecutive patients (72 eyes) treated with bilateral primary myopic or hyperopic femtosecond-assisted LASIK between October 2012 and January 2013 at our center, forming the “wide canal” group (group A). Specifically, the canal width was 1.7 mm, with a canal spot separation of 4.0 μm and a canal line separation of 2.5 μm (details of the venting channel are shown in Figure 1). This wider canal with a denser line separation was introduced in order to create a potentially more efficient venting corridor. Of the 72 flaps in this group, the majority (n = 48) were programmed to a diameter of 8.00 mm, while 22 flaps were programmed to 8.50 mm and two flaps were programmed to 9.50 mm. This difference reflected our preference for flap parameters in standard myopic,4 wider optical zone myopic, and hyperopic LASIK,5 respectively.
Figure 1

Illustration of the gas venting chimney mechanism for the FS200 femtosecond laser. (A) Top view, intraoperative surgeon’s view. The canal and hinge are at the 12 o’clock position. (B) Artist’s rendering of the channel through the canal created during the procedure.

A second group comprising 36 patients (72 eyes) previously treated between March 2012 and October 2012 with bilateral primary myopic or hyperopic femtosecond-assisted LASIK at our center using the “narrow canal” design, was selected to match group A in regard to flap diameter distribution (group B). In addition to the 1.3 mm wide canal, the canal line separation in group B patients was sparser at 4.0 μm. In all procedures, which were performed by the same surgeon (AJK), the LASIK flap was created using the WaveLight® FS200 femtosecond laser (Alcon Surgical, Fort Worth, TX, USA), and subsequent excimer ablation was performed using the WaveLight EX500 excimer laser.6 The femtosecond laser flap settings common to both groups were: stromal bed cut spot separation, 8 μm; line separation, 8 μm; side cut bed separation, 5 μm; line separation, 3 μm; side cut angle, 70°; bed cut pulse energy, 0.80 μJ; side cut pulse energy, 0.80 μJ; canal spot separation, 4.0 μm; and hinge angle, 45°.

Imaging and measurement

Intraoperative images were collected from the applanated corneas using the Alcon/WaveLight Refractive Suite WaveNet documentation software. These images are created by default during the refractive procedure and stored in the system software, and are available for documentation. Digital analysis of these images was used to determine if OBL was present, as described in detail elsewhere.10 Linear regression analysis was performed to seek possible correlations between intended versus achieved flap dimensions. Descriptive statistics (average, minimum, maximum, standard deviation, and range), comparative statistics, and linear regression were performed using Microsoft Excel 2010 (Microsoft Corporation, Redmond, WA, USA) and OriginLab version 9.0 Build B45 (OriginLab Corporation, Northampton, MA, USA). Analysis of variance between the two groups was performed using the OriginLab statistics tool.

Results

The mean subject age in group A at the time of surgery was 28.7 ± 6.6 (18–41) years, and the normality test P value was 0.078. The mean subject age in group B was 29.6 ± 7.8 (17–44) years, and the normality test P value was 0.054. OBL in group A (wide canal) was digitally measured to have a mean area of 3.69% ± 3.78%, where the percentage indicates the fraction of the OBL area with respect to the total flap area. The maximum OBL percentage was 11.3%, and the minimum was 0%. Of the 72 flaps examined, 27 had no OBL present (that is, OBL area 0%). The comparative OBL incidence metrics for both groups are presented in Table 1, and histogram data and box plots for the incidence of OBL in both groups are shown in Table 2 and Figure 2, respectively. The one-tailed t-test was performed because the results were expanding only in the positive direction, and yielded a value of P = 0.00452 between the groups.
Table 1

Incidence of opaque bubble layer, expressed as a percentage of total flap area, with comparative results for the wide and narrow canal groups

Wide canal groupNarrow canal group
Eyes (n)7272
Mean OBL area3.69%6.06%
Standard deviation3.78%6.58%
Minimum0.00%0.00%
Maximum11.34%20.34%
Table 2

Comparative OBL histogram data, expressed as % fraction of total flap area for the two groups in the study

OBL area (% of total flap area)number of cases
group A wide canal groupgroup B wide canal group
0%–2%3532
2%–4%25
4%–6%64
6%–8%175
8%–10%84
10%–12%45
12%–14%03
14%–16%07
16%–18%07
>18%2
Figure 2

Box plots for OBL, expressed as fraction of the total flap area for the two groups, indicating the 99% point with the × sign, and the mean point with the □ sign.

Note: Vertical axis, range of extent of OBL as a fraction of total flap area.

Abbreviation: OBL, opaque bubble layer

Discussion

Creation of a LASIK flap with a femtosecond laser is considered advantageous to microceratome7,8 for a more centered, higher controlled-geometry, both in depth9 as well as diameter.10 In an earlier effort to validate the precision and accuracy of flap creation, we had introduced a quantitative digital analysis technique for accurate measurement of flap diameter and extent of OBL for flaps created using the Alcon/WaveLight FS200 femtosecond laser during LASIK and prior to lifting of the flap.10 A major finding of this study was that OBL was rare and consistently of the “delayed” form, and that there was a “signature” of accumulation near the sides of the canal and towards the limbus (Figure 3). Our hypothesis to explain why the extent of OBL was so small is related to the proprietary technique used by the FS200 laser to address this issue, ie, flap creation starts with a venting dissection corridor which acts as a venting “chimney” along a canal designed by the surgeon. The canal, as seen by the intraoperative surgeon’s view, begins from the limbal area towards the flap hinge, and in the cross-sectional view, begins from the top of the conjunctiva towards the lamellar cut bed (Figures 1 and 4). Other lasers, such as the IntraLase FS60, create a vertical gas pocket near the hinge, acting as an escape route for gas that has built up within the cornea.9 This pocket may not solve the problem, because it just presses the gas into deep looser corneal tissue, whereas the venting canals helps bring the gas out of the eye. The other disadvantages of the pocket is possible gas breakthough towards the anterior chamber, hindering subsequent eximer laser ablation.
Figure 3

Two representative flaps from group A (canal width 1.7 mm, denser canal line separation, 2.5 μm) and group B (canal width 1.3 mm, sparser canal line separation, 4.0 μm).

Figure 4

Digital analysis of extent of OBL.

Notes: In this example, the OBL area is 2688 pixels, while the entire flap is 29,225 pixels. The extent of the flap was measured to be 9.20% of the entire flap area.

The novel settings were implemented to test our theory that OBL can be further reduced if the chimney volume provided by this FS200 design is increased. Until recently, our standard canal width was 1.3 mm. We have since introduced the 1.7 mm canal width, ie, the wide canal. In addition, the denser line spacing within the canal also helps to create a more efficient chimney. Despite the significance of OBL in femtosecond-laser LASIK flap creation, there are very few reports in the peer-reviewed literature regarding the extent and characteristics of OBL.2,11 Our analysis is based on immediate digital image processing of the flap creation (Figure 4), and we believe it is a novel way to document and measure the presence of OBL in a quantitative manner. Our study shows that the incidence of OBL is small, and it appears that the novel parameters used in flap canal creation are beneficial in terms of outcome, by improving the channeling of gas buildup outside the cornea at the limbus during lamellar cornea dissection. Although not specifically evaluated in this study, the wider canal was not associated with any astigmatic induction. The reason why OBL always appears in the hinge side of the flap is that OBL is caused by a buildup of trapped gas which is unable to vent. As the lamellar dissection of the laser progresses, the lamellar pocket space created becomes larger, thereby reducing this phenomenon. The type of OBL observed in our study was the so-called “soft” or “delayed’ OBL, as opposed to “hard” or “advancing” OBL ahead of the laser breakdown raster line. Thus, even this small and further reduced appearance of delayed OBL does not interfere with laser focusing during flap creation. With the implementation of the novel flap parameter settings, OBL was clinically nonexistent, with only with only a minute presence. Specifically, 50% of the flaps had less than 2.7% of total flap trace OBL, and only one of five of the flaps would be characterized as having actual OBL, albeit minimal, and affecting up to approximately 16% of the flap area. The reader has to consider that these incidence percentages cannot be compared with previous published studies or reports with OBL except ours. Our benchmark objective digital analysis is able to document and register clinically minute presence of OBL. It is therefore an objective and extremely stern criterion for the existence of OBL and its exact relative flap surface extent. Since the maximum OBL measured in the novel settings group is under 12%, one can claim clinical assessment is eliminated. Larger independent studies may be needed to validate these findings further. In the analysis reported here, our small sample yielded statistically significant data. The two sample groups were age-matched and procedure-matched, making this small study quite representative in our opinion.

Conclusion

Our study suggests that the wider canal settings along with customized denser spot application implemented in LASIK flap creation via the FS200 femtosecond laser significantly reduces the risk of OBL.
  11 in total

1.  Comparison of the IntraLase femtosecond laser and mechanical keratomes for laser in situ keratomileusis.

Authors:  Guy M Kezirian; Karl G Stonecipher
Journal:  J Cataract Refract Surg       Date:  2004-04       Impact factor: 3.351

2.  Incidence, possible risk factors, and potential effects of an opaque bubble layer created by a femtosecond laser.

Authors:  Igor Kaiserman; Hillel S Maresky; Irit Bahar; David S Rootman
Journal:  J Cataract Refract Surg       Date:  2008-03       Impact factor: 3.351

Review 3.  Femtosecond lasers in ophthalmology.

Authors:  H Kaz Soong; João Baptista Malta
Journal:  Am J Ophthalmol       Date:  2008-10-18       Impact factor: 5.258

4.  In vivo structural characteristics of the femtosecond LASIK-induced opaque bubble layers with ultrahigh-resolution SD-OCT.

Authors:  Volkan Hurmeric; Sonia H Yoo; Jane Fishler; Victoria S Chang; Jianhua Wang; William W Culbertson
Journal:  Ophthalmic Surg Lasers Imaging       Date:  2010 Nov-Dec

Review 5.  Femtosecond lasers for LASIK flap creation: a report by the American Academy of Ophthalmology.

Authors:  Ayad A Farjo; Alan Sugar; Steven C Schallhorn; Parag A Majmudar; David J Tanzer; William B Trattler; John B Cason; Kendall E Donaldson; George D Kymionis
Journal:  Ophthalmology       Date:  2012-11-20       Impact factor: 12.079

6.  Refractive and keratometric stability in high myopic LASIK with high-frequency femtosecond and excimer lasers.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  J Refract Surg       Date:  2013-10-04       Impact factor: 3.573

7.  Moria M2 single use microkeratome head in 100 consecutive LASIK procedures.

Authors:  A John Kanellopoulos; Lawrence H Pe; Lynda Kleiman
Journal:  J Refract Surg       Date:  2005 Sep-Oct       Impact factor: 3.573

8.  Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite.

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-02-21

9.  Three-dimensional LASIK flap thickness variability: topographic central, paracentral and peripheral assessment, in flaps created by a mechanical microkeratome (M2) and two different femtosecond lasers (FS60 and FS200).

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-04-03

10.  Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique.

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-02-15
View more
  6 in total

1.  [Scanning electron microscopic investigations of cutting edge quality in lamellar keratotomy using the Wavelight femtosecond laser (FS-200) : What influence do spot distance and an additional tunnel have?]

Authors:  T Hammer; T Höche; J Heichel
Journal:  Ophthalmologe       Date:  2018-01       Impact factor: 1.059

2.  Opaque bubble layer incidence in Femtosecond laser-assisted LASIK: comparison among different flap design parameters.

Authors:  Leonardo Mastropasqua; Roberta Calienno; Manuela Lanzini; Niccolò Salgari; Sergio De Vecchi; Rodolfo Mastropasqua; Mario Nubile
Journal:  Int Ophthalmol       Date:  2016-08-12       Impact factor: 2.031

3.  Flap-making patterns and corneal characteristics influence opaque bubble layer occurrence in femtosecond laser-assisted laser in situ keratomileusis.

Authors:  Xi He; Shi-Ming Li; Changbin Zhai; Li Zhang; Yue Wang; Xiumei Song; Yi Wang
Journal:  BMC Ophthalmol       Date:  2022-07-11       Impact factor: 2.086

4.  Clear-cornea cataract surgery: pupil size and shape changes, along with anterior chamber volume and depth changes. A Scheimpflug imaging study.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2014-10-24

5.  Incidence and Outcomes of Anterior Chamber Gas Bubble during Femtosecond Flap Creation for Laser-Assisted In Situ Keratomileusis.

Authors:  Sloan W Rush; Philip Cofoid; Ryan B Rush
Journal:  J Ophthalmol       Date:  2015-04-20       Impact factor: 1.909

Review 6.  The evolution of corneal and refractive surgery with the femtosecond laser.

Authors:  Antonis Aristeidou; Elise V Taniguchi; Michael Tsatsos; Rodrigo Muller; Colm McAlinden; Roberto Pineda; Eleftherios I Paschalis
Journal:  Eye Vis (Lond)       Date:  2015-07-14
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.