Moria M2 single use microkeratome head in 100 consecutive LASIK procedures.

A John Kanellopoulos1, Lawrence H Pe, Lynda Kleiman.   

Abstract

PURPOSE: To evaluate the safety and efficacy of the Moria M2 single use 130 microkeratome head in consecutive LASIK procedures for correction of myopia and myopic astigmatism.
METHODS: One hundred eyes of 55 patients underwent LASIK in which the flaps were created with the Moria M2 microkeratome using the single use 130 head and excimer laser ablation was done with the Allegretto Wave-light laser. Flap parameters measured were: thickness, diameter, hinge length, and overall quality. Preoperative uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity (BSCVA), refraction, wavefront aberrations, and low contrast sensitivity were compared to postoperative values at 6-month follow-up.
RESULTS: Mean flap thickness was 145 +/- 17.5 microm, mean flap diameter was 8.5 +/- 0.40 mm, and mean hinge cord length was 4.05 +/- 0.35 mm. At 6-month follow-up, UCVA improved from 20/200 (+/-0.24) to 20/18.5 (+/-0.12) and BSCVA improved from 20/20.5 (+/-0.18) to 20/17.5 (+/-0.11).
CONCLUSIONS: The Moria M2 single use 130 microkeratome head appears to be safe and effective in performing LASIK procedures.

Entities:  

Mesh:

Year:  2005        PMID: 16209445     DOI: 10.3928/1081-597X-20050901-09

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  5 in total

1.  Essential opaque bubble layer elimination with novel LASIK flap settings in the FS200 Femtosecond Laser.

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-04-19

2.  Visumax femtolasik versus Moria M2 microkeratome in mild to moderate myopia: efficacy, safety, predictability, aberrometric changes and flap thickness predictability.

Authors:  Magda A Torky; Yousif A Al Zafiri; Abeer M Khattab; Rania K Farag; Eman A Awad
Journal:  BMC Ophthalmol       Date:  2017-07-17       Impact factor: 2.209

3.  Corneal flap thickness with the Moria M2 single-use head 90 microkeratome in 72 consecutive LASIK procedures.

Authors:  Yunus Karabela; Orkun Muftuoglu; Faruk Kaya
Journal:  Clin Ophthalmol       Date:  2017-03-03

4.  Three-dimensional LASIK flap thickness variability: topographic central, paracentral and peripheral assessment, in flaps created by a mechanical microkeratome (M2) and two different femtosecond lasers (FS60 and FS200).

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-04-03

5.  Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique.

Authors:  A John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-02-15