SPECT system optimization against a discrete parameter space.

L J Meng1, N Li.   

Abstract

In this paper, we present an analytical approach for optimizing the design of a static SPECT system or optimizing the sampling strategy with a variable/adaptive SPECT imaging hardware against an arbitrarily given set of system parameters. This approach has three key aspects. First, it is designed to operate over a discretized system parameter space. Second, we have introduced an artificial concept of virtual detector as the basic building block of an imaging system. With a SPECT system described as a collection of the virtual detectors, one can convert the task of system optimization into a process of finding the optimum imaging time distribution (ITD) across all virtual detectors. Thirdly, the optimization problem (finding the optimum ITD) could be solved with a block-iterative approach or other nonlinear optimization algorithms. In essence, the resultant optimum ITD could provide a quantitative measure of the relative importance (or effectiveness) of the virtual detectors and help to identify the system configuration or sampling strategy that leads to an optimum imaging performance. Although we are using SPECT imaging as a platform to demonstrate the system optimization strategy, this development also provides a useful framework for system optimization problems in other modalities, such as positron emission tomography and x-ray computed tomography (Moore et al (2009 IEEE Nucl. Sci. Symp. Conf. Rec. pp 4154-7), Freed et al (2008 Med. Phys. 35 1912-25)).

Entities:  

Mesh:

Year:  2013        PMID: 23587609      PMCID: PMC3703762          DOI: 10.1088/0031-9155/58/9/3037

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  33 in total

1.  Reconstruction of two- and three-dimensional images from synthetic-collimator data.

Authors:  D W Wilson; H H Barrett; E W Clarkson
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

2.  Resolution and noise properties of MAP reconstruction for fully 3-D PET.

Authors:  J Qi; R M Leahy
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

3.  Theoretical study of lesion detectability of MAP reconstruction using computer observers.

Authors:  J Qi; R H Huesman
Journal:  IEEE Trans Med Imaging       Date:  2001-08       Impact factor: 10.048

4.  A modified uniform Cramer-Rao bound for multiple pinhole aperture design.

Authors:  L J Meng; N H Clinthorne
Journal:  IEEE Trans Med Imaging       Date:  2004-07       Impact factor: 10.048

Review 5.  In vivo molecular and genomic imaging: new challenges for imaging physics.

Authors:  Simon R Cherry
Journal:  Phys Med Biol       Date:  2004-02-07       Impact factor: 3.609

Review 6.  Small animal SPECT and its place in the matrix of molecular imaging technologies.

Authors:  Steven R Meikle; Peter Kench; Michael Kassiou; Richard B Banati
Journal:  Phys Med Biol       Date:  2005-10-24       Impact factor: 3.609

7.  Design and simulation of a high-resolution stationary SPECT system for small animals.

Authors:  Freek J Beekman; Brendan Vastenhouw
Journal:  Phys Med Biol       Date:  2004-10-07       Impact factor: 3.609

8.  Theoretical study of penalized-likelihood image reconstruction for region of interest quantification.

Authors:  Jinyi Qi; Ronald H Huesman
Journal:  IEEE Trans Med Imaging       Date:  2006-05       Impact factor: 10.048

9.  Fast LROC analysis of Bayesian reconstructed emission tomographic images using model observers.

Authors:  Parmeshwar Khurd; Gene Gindi
Journal:  Phys Med Biol       Date:  2005-03-22       Impact factor: 3.609

10.  The use of low glycemic and high satiety index food dishes in Mexico: a low cost approach to prevent and control obesity and diabetes.

Authors:  A Jiménez-Cruz; V Manuel Loustaunau-López; M Bacardi-Gascón
Journal:  Nutr Hosp       Date:  2006 May-Jun       Impact factor: 1.057

View more
  1 in total

1.  Simulation study of the second-generation MR-compatible SPECT system based on the inverted compound-eye gamma camera design.

Authors:  Xiaochun Lai; Ling-Jian Meng
Journal:  Phys Med Biol       Date:  2018-02-12       Impact factor: 3.609

  1 in total