Optimization of dual-energy xenon-computed tomography for quantitative assessment of regional pulmonary ventilation.

Matthew K Fuld1, Ahmed F Halaweish, John D Newell, Bernhard Krauss, Eric A Hoffman.   

Abstract

OBJECTIVE: Dual-energy x-ray computed tomography (DECT) offers visualization of the airways and quantitation of regional pulmonary ventilation using a single breath of inhaled xenon gas. In this study, we sought to optimize scanning protocols for DECT xenon gas ventilation imaging of the airways and lung parenchyma and to characterize the quantitative nature of the developed protocols through a series of test-object and animal studies.
MATERIALS AND METHODS: The Institutional Animal Care and Use Committee approved all animal studies reported here. A range of xenon/oxygen gas mixtures (0%, 20%, 25%, 33%, 50%, 66%, 100%; balance oxygen) were scanned in syringes and balloon test-objects to optimize the delivered gas mixture for assessment of regional ventilation while allowing for the development of improved 3-material decomposition calibration parameters. In addition, to alleviate gravitational effects on xenon gas distribution, we replaced a portion of the oxygen in the xenon/oxygen gas mixture with helium and compared gas distributions in a rapid-prototyped human central-airway test-object. Additional syringe tests were performed to determine if the introduction of helium had any effect on xenon quantitation. Xenon gas mixtures were delivered to anesthetized swine to assess airway and lung parenchymal opacification while evaluating various DECT scan acquisition settings.
RESULTS: Attenuation curves for xenon were obtained from the syringe test-objects and were used to develop improved 3-material decomposition parameters (Hounsfield unit enhancement per percentage xenon: within the chest phantom, 2.25 at 80 kVp, 1.7 at 100 kVp, and 0.76 at 140 kVp with tin filtration; in open air, 2.5 at 80 kVp, 1.95 at 100 kVp, and 0.81 at 140 kVp with tin filtration). The addition of helium improved the distribution of xenon gas to the gravitationally nondependent portion of the airway tree test-object, while not affecting the quantitation of xenon in the 3-material decomposition DECT. The mixture 40% Xe/40% He/20% O2 provided good signal-to-noise ratio (SNR), greater than the Rose criterion (SNR > 5), while avoiding gravitational effects of similar concentrations of xenon in a 60% O2 mixture. Compared with 100/140 Sn kVp, 80/140 Sn kVp (Sn = tin filtered) provided improved SNR in a swine with an equivalent thoracic transverse density to a human subject with a body mass index of 33 kg/m. Airways were brighter in the 80/140 Sn kVp scan (80/140 Sn, 31.6%; 100/140 Sn, 25.1%) with considerably lower noise (80/140 Sn, coefficient of variation of 0.140; 100/140 Sn, coefficient of variation of 0.216).
CONCLUSION: To provide a truly quantitative measure of regional lung function with xenon-DECT, the basic protocols and parameter calibrations need to be better understood and quantified. It is critically important to understand the fundamentals of new techniques to allow for proper implementation and interpretation of their results before widespread usage. With the use of an in-house derived xenon calibration curve for 3-material decomposition rather than the scanner supplied calibration and a xenon/helium/oxygen mixture, we demonstrate highly accurate quantitation of xenon gas volumes and avoid gravitational effects on gas distribution. This study provides a foundation for other researchers to use and test these methods with the goal of clinical translation.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23571834      PMCID: PMC3873149          DOI: 10.1097/RLI.0b013e31828ad647

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  41 in total

1.  Collateral ventilation in a canine model with bronchial obstruction: assessment with xenon-enhanced dual-energy CT.

Authors:  Eun Jin Chae; Joon Beom Seo; Namkug Kim; Koun-Sik Song; Ji Hoon Shin; Tae-Hyung Kim; Yedaun Lee
Journal:  Radiology       Date:  2010-06       Impact factor: 11.105

2.  Characteristics of airflow in a CT-based ovine lung: a numerical study.

Authors:  Senthil Kabilan; Ching-Long Lin; Eric A Hoffman
Journal:  J Appl Physiol (1985)       Date:  2006-11-16

3.  Recirculation of inhaled xenon does not alter lung CT density.

Authors:  Jeffrey B Hoag; Matthew Fuld; Robert H Brown; Brett A Simon
Journal:  Acad Radiol       Date:  2007-01       Impact factor: 3.173

4.  Quantitative imaging of element composition and mass fraction using dual-energy CT: three-material decomposition.

Authors:  Xin Liu; Lifeng Yu; Andrew N Primak; Cynthia H McCollough
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

5.  Chronic obstructive pulmonary disease: quantitative and visual ventilation pattern analysis at xenon ventilation CT performed by using a dual-energy technique.

Authors:  Eun-Ah Park; Jin Mo Goo; Sang Joon Park; Hyun Ju Lee; Chang Hyun Lee; Chang Min Park; Chul-Gyu Yoo; Jong Hyo Kim
Journal:  Radiology       Date:  2010-07-22       Impact factor: 11.105

6.  Dual-source dual-energy CT with additional tin filtration: Dose and image quality evaluation in phantoms and in vivo.

Authors:  Andrew N Primak; Juan Carlos Ramirez Giraldo; Christian D Eusemann; Bernhard Schmidt; Birgit Kantor; Joel G Fletcher; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2010-11       Impact factor: 3.959

7.  Xenon ventilation imaging using dual-energy computed tomography in asthmatics: initial experience.

Authors:  Eun Jin Chae; Joon Beom Seo; Jeongjin Lee; Namkug Kim; Hyun Woo Goo; Hyun Joo Lee; Choong Wook Lee; Seung Won Ra; Yeon-Mok Oh; You Sook Cho
Journal:  Invest Radiol       Date:  2010-06       Impact factor: 6.016

8.  Effect of low-xenon and krypton supplementation on signal/noise of regional CT-based ventilation measurements.

Authors:  Deokiee Chon; Kenneth C Beck; Brett A Simon; Hidenori Shikata; Osama I Saba; Eric A Hoffman
Journal:  J Appl Physiol (1985)       Date:  2006-11-22

9.  Xenon ventilation CT using dual-source and dual-energy technique in children with bronchiolitis obliterans: correlation of xenon and CT density values with pulmonary function test results.

Authors:  Hyun Woo Goo; Dong Hyun Yang; Soo-Jong Hong; Jinho Yu; Byoung-Ju Kim; Joon Beom Seo; Eun Jin Chae; Jeongjin Lee; Bernhard Krauss
Journal:  Pediatr Radiol       Date:  2010-04-22

10.  Xenon ventilation CT with a dual-energy technique of dual-source CT: initial experience.

Authors:  Eun Jin Chae; Joon Beom Seo; Hyun Woo Goo; Namkug Kim; Koun-Sik Song; Sang Do Lee; Soo-Jong Hong; Bernhard Krauss
Journal:  Radiology       Date:  2008-08       Impact factor: 11.105

View more
  9 in total

1.  Ventilation/Perfusion Relationships and Gas Exchange: Measurement Approaches.

Authors:  Susan R Hopkins
Journal:  Compr Physiol       Date:  2020-07-08       Impact factor: 9.090

Review 2.  Pulmonary CT and MRI phenotypes that help explain chronic pulmonary obstruction disease pathophysiology and outcomes.

Authors:  Eric A Hoffman; David A Lynch; R Graham Barr; Edwin J R van Beek; Grace Parraga
Journal:  J Magn Reson Imaging       Date:  2015-07-22       Impact factor: 4.813

Review 3.  Assessment of Heterogeneity in Lung Structure and Function During Mechanical Ventilation: A Review of Methodologies.

Authors:  Jacob Herrmann; Michaela Kollisch-Singule; Joshua Satalin; Gary F Nieman; David W Kaczka
Journal:  J Eng Sci Med Diagn Ther       Date:  2022-05-11

Review 4.  Lung aeration, ventilation, and perfusion imaging.

Authors:  Lorenzo Ball; Gaetano Scaramuzzo; Jacob Herrmann; Maurizio Cereda
Journal:  Curr Opin Crit Care       Date:  2022-06-01       Impact factor: 3.359

Review 5.  Origins of and lessons from quantitative functional X-ray computed tomography of the lung.

Authors:  Eric A Hoffman
Journal:  Br J Radiol       Date:  2022-03-01       Impact factor: 3.629

Review 6.  Imaging the Injured Lung: Mechanisms of Action and Clinical Use.

Authors:  Maurizio Cereda; Yi Xin; Alberto Goffi; Jacob Herrmann; David W Kaczka; Brian P Kavanagh; Gaetano Perchiazzi; Takeshi Yoshida; Rahim R Rizi
Journal:  Anesthesiology       Date:  2019-09       Impact factor: 7.892

7.  Ventilation heterogeneity imaged by multibreath wash-ins of hyperpolarized 3 He and 129 Xe in healthy rabbits.

Authors:  Hooman Hamedani; Stephen Kadlecek; Kai Ruppert; Yi Xin; Ian Duncan; Rahim R Rizi
Journal:  J Physiol       Date:  2021-08-10       Impact factor: 6.228

8.  Regional Gas Transport During Conventional and Oscillatory Ventilation Assessed by Xenon-Enhanced Computed Tomography.

Authors:  Jacob Herrmann; Sarah E Gerard; Joseph M Reinhardt; Eric A Hoffman; David W Kaczka
Journal:  Ann Biomed Eng       Date:  2021-05-04       Impact factor: 4.219

Review 9.  Functional lung imaging in thoracic tumor radiotherapy: Application and progress.

Authors:  Pi-Xiao Zhou; Shu-Xu Zhang
Journal:  Front Oncol       Date:  2022-09-23       Impact factor: 5.738

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.