Airpuff tonometry versus applanation tonometry.

O Lagerlöf.   

Abstract

In order to study the reliability of airpuff tonometry, 230 eyes of 115 patients had their IOP measured with 2 airpuff tonometers, NCT and Pulsair, and one applanation tonometer, Perkins handheld Goldmann Type, between Dec 87 and March 89. Each patient had three readings per eye with each airpuff instrument and two readings per eye with the Perkins instrument. Mean age for the group was 68 +/- 11 (SD), range 36-87 years. The mean values were for Perkins: 21.21 +/- 6.27 R.E., 21.28 +/- 5.67 L.E.; for NCT: 21.11 +/- 6.95 R.E., 21.27 +/- 5.41 L.E.; for Pulsair: 23.05 +/- 6.00 R.E., 23.76 +/- 5.39 L.E. The ranges were for Perkins: 9-45 R.E., 9-38 L.E.; for NCT: 8-59 R.E., 11-38 L.E.; for Pulsair: 9-52 R.E., 11-38 L.E. For the Pulsair a value of 18 was found to be a cut-off point below which there were no false negatives, and for the NCT this value was found to be 17. A linear relation between airpuff and applanation tonometry was found, but also a large spread, both for false positives and false negatives.

Entities:  

Mesh:

Year:  1990        PMID: 2356713     DOI: 10.1111/j.1755-3768.1990.tb01909.x

Source DB:  PubMed          Journal:  Acta Ophthalmol (Copenh)        ISSN: 0001-639X


  4 in total

1.  Clinical evaluation of two non-contact tonometers.

Authors:  S A Koopmans; A C Kooijman; G van Rij; J Eisses; J de Groot-Woltjer
Journal:  Doc Ophthalmol       Date:  1991       Impact factor: 2.379

2.  Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer.

Authors:  Qasim K Farhood
Journal:  Clin Ophthalmol       Date:  2012-12-27

3.  Intraocular Pressure Based on Dynamic Bidirectional Applanation and Air-puff Tonometry: A Comparative Study.

Authors:  Ayse E Bahadir Kilavuzoglu; Cemile B Cosar; Ali Rc Celebi; Ugur E Al Parmak
Journal:  J Curr Glaucoma Pract       Date:  2019 May-Aug

4.  Comparative Assessment of The Goldmann Applanation and Noncontact Tonometers in Intraocular Pressure Measurements in a Sample of Glaucoma Patients in the Cape Coast Metropolis, Ghana.

Authors:  Samuel Kyei; Cynthia Pakyennu Gboglu; Michael Agyemang Kwarteng; Frank Assiamah
Journal:  Niger Med J       Date:  2020-12-19