Fatigue fracture of tibial arthroplasty implant masked by contralateral knee arthritis.

Barry J O'Neill1, May Cleary, John P McElwain.   

Abstract

INTRODUCTION: The wear of polyethylene components is a well-recognised long-term complication of total knee arthroplasty. PRESENTATION OF CASE: We present a case of fatigue fracture of a geomedic tibial tray 17 years after implantation. The symptoms were masked by primary osteoarthritis of the contralateral knee. DISCUSSION: The mode of failure in this case differs from that most commonly reported in the literature in that it occurred late as the result of prolonged use, not due to implant failure, physiological factors, or surgical technique. The case also demonstrates the incredible 'masking' effect of arthritic knee pain.
CONCLUSION: This case highlights the importance to clinicians of keeping a high index of suspicion for prosthesis complications in patients with unilateral joint replacement with contralateral arthropathy.
Copyright © 2013 Surgical Associates Ltd. Published by Elsevier Ltd. All rights reserved.

Entities:  

Year:  2013        PMID: 23562901      PMCID: PMC3731697          DOI: 10.1016/j.ijscr.2013.02.019

Source DB:  PubMed          Journal:  Int J Surg Case Rep        ISSN: 2210-2612


  20 in total

1.  Influence of pre-operative bone mineral content of the proximal tibia on revision rate after uncemented knee arthroplasty.

Authors:  M Therbo; M M Petersen; J E Varmarken; C A Olsen; B Lund
Journal:  J Bone Joint Surg Br       Date:  2003-09

2.  Bilateral fatigue fracture of the femoral components in a cruciate-retaining cementless total knee prosthesis.

Authors:  Shu Saito; Yasuaki Tokuhashi; Takao Ishii; Sei Mori; Kunihiro Hosaka; Keinosuke Ryu; Gen Suzuki
Journal:  Orthopedics       Date:  2011-10-05       Impact factor: 1.390

3.  Segmental stem fracture of a cemented femoral prosthesis.

Authors:  Kristiina Jarvi; Robert M Kerry
Journal:  J Arthroplasty       Date:  2007-01-22       Impact factor: 4.757

4.  Fracture of the porous-coated metal tray of a biologically fixed knee prosthesis. Report of a case.

Authors:  B F Morrey; E Y Chao
Journal:  Clin Orthop Relat Res       Date:  1988-03       Impact factor: 4.176

5.  Retrieval analysis of porous-coated components for total knee arthroplasty. A report of two cases.

Authors:  C S Ranawat; N A Johanson; C M Rimnac; T M Wright; R E Schwartz
Journal:  Clin Orthop Relat Res       Date:  1986-08       Impact factor: 4.176

6.  Fracture of the metallic tibial tray following total knee replacement. Report of two cases.

Authors:  R D Scott; F C Ewald; P S Walker
Journal:  J Bone Joint Surg Am       Date:  1984-06       Impact factor: 5.284

7.  Fractures of modern high nitrogen stainless steel cemented stems: cause, mechanism, and avoidance in 14 cases.

Authors:  Piers J Yates; Nasir A Quraishi; Allan Kop; Donald W Howie; Clare Marx; Eric Swarts
Journal:  J Arthroplasty       Date:  2007-10-24       Impact factor: 4.757

8.  Knee flexion influences periprosthetic BMD measurement in the tibia. Suggestions for a reproducible clinical scan protocol.

Authors:  Maiken Stilling; Kjeld Søballe; Kristian Larsen; Niels Trolle Andersen; Ole Rahbek
Journal:  Acta Orthop       Date:  2010-08       Impact factor: 3.717

Review 9.  The influence of obesity on the complication rate and outcome of total knee arthroplasty: a meta-analysis and systematic literature review.

Authors:  Gino M M J Kerkhoffs; Elvire Servien; Warren Dunn; Diane Dahm; Jos A M Bramer; Daniel Haverkamp
Journal:  J Bone Joint Surg Am       Date:  2012-10-17       Impact factor: 5.284

10.  Bone mineral density in the proximal femur and contralateral knee after total knee arthroplasty.

Authors:  Tarja A Soininvaara; Hannu J A Miettinen; Jukka S Jurvelin; Esko M Alhava; Heikki P J Kröger
Journal:  J Clin Densitom       Date:  2004       Impact factor: 2.963

View more
  5 in total

1.  Early Tibial Component Fractures in a Cementless, 3D-Printed, Titanium Implant.

Authors:  Alan D Lam; Gavan P Duffy
Journal:  Arthroplast Today       Date:  2022-10-12

2.  Tibial baseplate fracture associated with polyethylene wear and osteolysis after total knee arthroplasty.

Authors:  Deniz Cankaya; Serdar Yilmaz; Alper Deveci; Bulent Ozkurt
Journal:  Int J Surg Case Rep       Date:  2013-12-21

3.  Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty.

Authors:  Kazuaki Mineta; Masahiko Okada; Soshi Matsumoto; Daisuke Hamada; Tomohiro Goto; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2017-05-16

4.  Radiographic False Evidence of a Tibial Baseplate Fracture After Total Knee Arthroplasty.

Authors:  David S Constantinescu; Jeremy A Ross; Nirav K Patel; Benjamin M Strong; Laura A Giambra; Gregory J Golladay
Journal:  Arthroplast Today       Date:  2021-03-08

5.  Tibial tray fracture in a modern prosthesis with retrieval analysis.

Authors:  John J Callaghan; David E DeMik; Nicholas A Bedard; Andrew N Odland; William M Kane; Steven M Kurtz
Journal:  Arthroplast Today       Date:  2018-02-10
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.