Literature DB >> 18280411

Fractures of modern high nitrogen stainless steel cemented stems: cause, mechanism, and avoidance in 14 cases.

Piers J Yates1, Nasir A Quraishi, Allan Kop, Donald W Howie, Clare Marx, Eric Swarts.   

Abstract

We present 14 cases of fracture of modern, high-nitrogen, stainless steel stems. Our clinical and radiological data suggest that heavy patients with small stems and poor proximal support are at risk for fracturing their implants. "Champagne-glass" canals can lead to the use of smaller stems often placed in varus, which can lead to cantilever bending and fatigue failure in the distal half of the stem. Metallurgical assessment of the retrieved high-nitrogen, stainless steel stems reveals microstructural inconsistencies that may contribute to their failure. Based on our findings, careful consideration and attention to technique is required when using stainless steel stems in patients with high body mass index or high weight. Technique is particularly important in femurs with champagne-glass canals.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18280411     DOI: 10.1016/j.arth.2006.12.051

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  5 in total

1.  A comparison of the omega and posterior approaches on patient reported function and radiological outcomes following total hip replacement.

Authors:  James R Berstock; Ashley W Blom; Michael R Whitehouse
Journal:  J Orthop       Date:  2017-06-24

2.  Initial evaluation on subsidence of cemented collarless polished tapered stem applied to the patients with narrow femoral medullar canal.

Authors:  Katsuyuki Dairaku; Masaji Ishii; Shinji Kobayashi; Hiroyuki Kawaji; Kan Sasaki; Yuya Takakubo; Michiaki Takagi
Journal:  Open Orthop J       Date:  2010-03-16

3.  Fatigue fracture of tibial arthroplasty implant masked by contralateral knee arthritis.

Authors:  Barry J O'Neill; May Cleary; John P McElwain
Journal:  Int J Surg Case Rep       Date:  2013-03-14

4.  Long-term results of cementless hip arthroplasty with ceramic-on-ceramic articulation.

Authors:  Marek Synder; Marek Drobniewski; Marcin Sibiński
Journal:  Int Orthop       Date:  2012-08-19       Impact factor: 3.075

5.  Femoral stem fractures after primary and revision hip replacements: A single-center experience.

Authors:  Alper Köksal; Ali Öner; Osman Çimen; Osman Emre Aycan; Hakan Akgün; Furkan Yapıcı; Yalkın Çamurcu
Journal:  Jt Dis Relat Surg       Date:  2020
  5 in total