Literature DB >> 33732836

Radiographic False Evidence of a Tibial Baseplate Fracture After Total Knee Arthroplasty.

David S Constantinescu1, Jeremy A Ross2, Nirav K Patel2, Benjamin M Strong3, Laura A Giambra2, Gregory J Golladay2.   

Abstract

Fracture of the tibial baseplate is a rare but dramatic cause of typically late fatigue failure in the setting of loosening after total knee arthroplasty. A 58-year-old female presented 4 months after total knee arthroplasty for evaluation of contralateral knee pain. Plain radiographs of the left knee incidentally suggested the possibility of tibial baseplate fracture despite minimal, expected postoperative symptoms. Subsequent computed tomography imaging demonstrated no confirmatory evidence of component failure or fracture. Malalignment and fatigue fracture are proposed etiologies of baseplate fractures. The presented case illustrates the importance of computed tomography imaging and clinical correlation when a diagnosis of baseplate fracture is suspected to avoid an unnecessary revision surgery.
© 2021 The Authors.

Entities:  

Keywords:  Complications; Implant failure; Knee arthroplasty

Year:  2021        PMID: 33732836      PMCID: PMC7943966          DOI: 10.1016/j.artd.2021.02.004

Source DB:  PubMed          Journal:  Arthroplast Today        ISSN: 2352-3441


  7 in total

1.  Retrieval study of tibial baseplate fracture after total knee arthroplasty.

Authors:  Urjit Chatterji; Mark J Ashworth; Antony L Smith; Nigel Brewster; Peter L Lewis
Journal:  J Arthroplasty       Date:  2005-01       Impact factor: 4.757

2.  Fractured Whiteside Ortholoc II knee components.

Authors:  E Swarts; S J Miller; C V Keogh; G Lim; R J Beaver
Journal:  J Arthroplasty       Date:  2001-10       Impact factor: 4.757

3.  Fatigue fracture of tibial arthroplasty implant masked by contralateral knee arthritis.

Authors:  Barry J O'Neill; May Cleary; John P McElwain
Journal:  Int J Surg Case Rep       Date:  2013-03-14

4.  Tibial baseplate fracture associated with polyethylene wear and osteolysis after total knee arthroplasty.

Authors:  Deniz Cankaya; Serdar Yilmaz; Alper Deveci; Bulent Ozkurt
Journal:  Int J Surg Case Rep       Date:  2013-12-21

5.  Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty.

Authors:  Kazuaki Mineta; Masahiko Okada; Soshi Matsumoto; Daisuke Hamada; Tomohiro Goto; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2017-05-16

6.  Tibial tray fracture in a modern prosthesis with retrieval analysis.

Authors:  John J Callaghan; David E DeMik; Nicholas A Bedard; Andrew N Odland; William M Kane; Steven M Kurtz
Journal:  Arthroplast Today       Date:  2018-02-10

7.  Catastrophic tibial baseplate failure of a modern cementless total knee arthroplasty implant.

Authors:  William F Scully; Matthew E Deren; Michael R Bloomfield
Journal:  Arthroplast Today       Date:  2019-10-16
  7 in total
  1 in total

1.  Catastrophic Failure of the Tibial Component After Total Knee Arthroplasty: Fracture and Dissociation Between the Baseplate and Stem.

Authors:  Christopher A Worgul; Brandon Lentine; Quinten G Dicken; David M Freccero
Journal:  Arthroplast Today       Date:  2022-03-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.