| Literature DB >> 23560174 |
Beong Ik Hur1, Jae Min Lee, Won Ho Cho, Dong Wan Kang, Choong Rak Kim, Byung Kwan Choi.
Abstract
OBJECTIVE: The Leksell Gamma Knife® (LGK) is based on a single-fraction high dose treatment strategy. Therefore, independent verification of the Leksell GammaPlan® (LGP) is important for ensuring patient safety and minimizing the risk of treatment errors. Although several verification techniques have been previously developed and reported, no method has ever been tested statistically on multiple LGK target treatments. The purpose of this study was to perform and to evaluate the accuracy of a verification method (modified variable ellipsoid modeling technique, MVEMT) for multiple target treatments.Entities:
Keywords: Brain tumor; Gamma Knife; Radiation dose; Radiosurgery
Year: 2013 PMID: 23560174 PMCID: PMC3611052 DOI: 10.3340/jkns.2013.53.2.102
Source DB: PubMed Journal: J Korean Neurosurg Soc ISSN: 1225-8245
Fig. 1Bland-Altman Plots for each individual case. The middle line represents the mean difference between the LGP and MVEMT. The two extreme lines are the +1.96 and -1.96 standard deviations of the difference between the two methods. MVEMT : modified variable ellipsoid modeling technique, LGP : Leksell GammaPlan.
Fig. 2A : Bland-Altman Plot for all pooled data (n=500). The mean difference between doses calculated by two methods, LGP and MVEMT (95% CI) was -0.057. B : Bland-Altman Plot for data excluding case 6 (n=450). The mean difference between doses calculated by two methods (95% CI) was -0.140. The middle line is the mean difference between LGP and MVEMT. The two extreme lines are the +1.96 and -1.96 standard deviations of the difference between two methods. MVEMT : modified variable ellipsoid modeling technique, LGP : Leksell GammaPlan.
Paired t-test for difference of Leksell GammaPlan and MVEMT
MVEMT : modified variable ellipsoid modeling technique, SD : standard deviation
Maximal dose and its location
*p=0.323. MVEMT : modified variable ellipsoid modeling technique, LGP : Leksell GammaPlan, SD : standard deviation
Comparison of target volume (50%)
*p=0.626. MVEMT : modified variable ellipsoid modeling technique, LGP : Leksell GammaPlan, SD : standard deviation