A formalism for independent checking of Gamma Knife dose calculations.

J S Tsai1, M J Engler, M J Rivard, A Mahajan, J A Borden, Z Zheng.   

Abstract

For stereotactic radiosurgery using the Leksell Gamma Knife system, it is important to perform a pre-treatment verification of the maximum dose calculated with the Leksell GammaPlan (DLGP) stereotactic radiosurgery system. This verification can be incorporated as part of a routine quality assurance (QA) procedure to minimize the chance of a hazardous overdose. To implement this procedure, a formalism has been developed to calculate the dose DCAL(X,Y,Z,dav,t) using the following parameters: average target depth (dav), coordinates (X,Y,Z) of the maximum dose location or any other dose point(s) to be verified, 3-dimensional (3-dim) beam profiles or off-centerratios (OCR) of the four helmets, helmet size i, output factor Oi, plug factor Pi, each shot j coordinates (x,y,z)i,j, and shot treatment time (ti,j). The average depth of the target dav was obtained either from MRI/CT images or ruler measurements of the Gamma Knife Bubble Head Frame. DCAL and DLGP were then compared to evaluate the accuracy of this independent calculation. The proposed calculation for an independent check of DLGP has been demonstrated to be accurate and reliable, and thus serves as a QA tool for Gamma Knife stereotactic radiosurgery.

Entities:  

Mesh:

Year:  2001        PMID: 11585215     DOI: 10.1118/1.1398038

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  The variable ellipsoid modeling technique as a verification method for the treatment planning system of gamma knife radiosurgery.

Authors:  Beong Ik Hur; Byung Kwan Choi; Soon Ki Sung; Won Ho Cho; Seung Heon Cha; Chang Hwa Choi
Journal:  J Korean Neurosurg Soc       Date:  2010-02-28

2.  Validation of Gamma Knife Perfexion Dose Profile Distribution by a Modified Variable Ellipsoid Modeling Technique.

Authors:  Beong Ik Hur; Seong Jin Jin; Gyeong Rip Kim; Jong Hyeok Kwak; Young Ha Kim; Sang Weon Lee; Soon Ki Sung
Journal:  J Korean Neurosurg Soc       Date:  2021-01-01

3.  Clinical application of gamma knife dose verification method in multiple brain tumors : modified variable ellipsoid modeling technique.

Authors:  Beong Ik Hur; Jae Min Lee; Won Ho Cho; Dong Wan Kang; Choong Rak Kim; Byung Kwan Choi
Journal:  J Korean Neurosurg Soc       Date:  2013-02-28

4.  Two-year experience with the commercial Gamma Knife Check software.

Authors:  Andy Yuanguang Xu; Jagdish Bhatnagar; Greg Bednarz; Josef Novotny; John Flickinger; L Dade Lunsford; M Saiful Huq
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

5.  Development and validation of a comprehensive patient-specific quality assurance program for a novel stereotactic radiation delivery system for breast lesions.

Authors:  Stewart J Becker; Ying Niu; Yildirim Mutaf; Shifeng Chen; Yannick Poirier; Elizabeth M Nichols; ByongYong Yi
Journal:  J Appl Clin Med Phys       Date:  2019-12       Impact factor: 2.102