Multinight recording and analysis of continuous positive airway pressure airflow in the home for titration and management of sleep disordered breathing.

Cynthia Y Callahan1, Robert G Norman, Zachary Taxin, Anne M Mooney, David M Rapoport, Indu Ayappa.   

Abstract

OBJECTIVES: The authors examined magnitude/variability of residual sleep disordered breathing (SDB) at pressures around the therapeutic continuous positive airway pressure (CPAP), and described a multinight approach to CPAP titration/retitration consisting of recording airflow and summarizing SDB over multiple nights at multiple pressures and choosing an optimal pressure from these summarized data.
DESIGN: Prospective, single-center nonblinded study. PATIENTS: Ten female/18 male patients with obstructive sleep apnea-hypopnea syndrome (OSAHS) (respiratory disturbance index [RDI] 67/h), 17 newly-initiated, 11 chronic CPAP users.
INTERVENTIONS: A custom CPAP device (Fisher & Paykel Healthcare) recording airflow and pre-programmed to vary CPAP between 2-3 cm H2O below and 1-2 cm H2O above prescription pressure as determined by a full laboratory titration.
RESULTS: Airflow and pressure continuously recorded for multiple nights (15.9 ± 5.1 nights) at four to seven different pressures in each patient. SDB events manually scored from the airflow as apnea (airflow reduction > 90%), hypopnea (airflow reduction > 30% lasting 10 to 120 sec with inspira-tory flow limitation [IFL]) and runs of sustained IFL > 2 min identified. RDI = (apnea + hypopnea)/total sleep time calculated for each night and an obstruction index, including sustained IFL, also was calculated. PressureMultinight was obtained for each patient from multiple nights of data using two mathematical techniques. Night-to-night variability of SDB indices was low in some patients and significant in others. PressureMultinight could be determined in 17 of 28 patients and was similar to the in-laboratory pressure.
CONCLUSIONS: This study showed that recording multiple nights of CPAP airflow in the home and analyzing these data for residual SDB provided useful information, including the possibility of determining a therapeutic prescription for fixed CPAP in most patients and identification of others with significant physiologic variability of SDB.

Entities:  

Keywords:  CPAP Titration; General; Obstructive Sleep Apnea; Severity Metric; Sleep Disordered Breathing; Treatment Algorithm

Mesh:

Year:  2013        PMID: 23543909      PMCID: PMC3595181          DOI: 10.5665/sleep.2540

Source DB:  PubMed          Journal:  Sleep        ISSN: 0161-8105            Impact factor:   5.849


  37 in total

1.  Is (re)titration of nasal continuous positive airway pressure for obstructive sleep apnoea necessary?

Authors:  S Choi; R Mullins; J H Crosby; R J Davies; J R Stradling
Journal:  Sleep Med       Date:  2001-09       Impact factor: 3.492

2.  Assessment of inspiratory flow limitation invasively and noninvasively during sleep.

Authors:  S A Clark; C R Wilson; M Satoh; D Pegelow; J A Dempsey
Journal:  Am J Respir Crit Care Med       Date:  1998-09       Impact factor: 21.405

3.  Continuous positive airway pressure titration for obstructive sleep apnoea: automatic versus manual titration.

Authors:  Nigel McArdle; Bhajan Singh; Michelle Murphy; Kevin R Gain; Christine Maguire; Sarah Mutch; David R Hillman
Journal:  Thorax       Date:  2010-07       Impact factor: 9.139

4.  Continuous positive airway pressure therapy for treating sleepiness in a diverse population with obstructive sleep apnea: results of a meta-analysis.

Authors:  Sanjay R Patel; David P White; Atul Malhotra; Michael L Stanchina; Najib T Ayas
Journal:  Arch Intern Med       Date:  2003-03-10

5.  Positive airway pressure initiation: a randomized controlled trial to assess the impact of therapy mode and titration process on efficacy, adherence, and outcomes.

Authors:  Clete A Kushida; Richard B Berry; Alexander Blau; Tami Crabtree; Ingo Fietze; Meir H Kryger; Samuel T Kuna; G Vernon Pegram; Thomas Penzel
Journal:  Sleep       Date:  2011-08-01       Impact factor: 5.849

6.  The need for pressure changes in CPAP therapy 2-3 months after initial treatment: a prospective trial in 905 patients with sleep-disordered breathing.

Authors:  Nikolaus C Netzer; János Juhász; Markus Hofmann; Kathrin Hohl; Kingman P Strohl; Thomas E A H Küpper
Journal:  Sleep Breath       Date:  2010-03-04       Impact factor: 2.816

7.  Non-Invasive detection of respiratory effort-related arousals (REras) by a nasal cannula/pressure transducer system.

Authors:  I Ayappa; R G Norman; A C Krieger; A Rosen; R L O'malley; D M Rapoport
Journal:  Sleep       Date:  2000-09-15       Impact factor: 5.849

8.  Relative accuracy of algorithm-based prescription of nasal CPAP in OSA.

Authors:  J R Stradling; M Hardinge; J Paxton; D M Smith
Journal:  Respir Med       Date:  2004-02       Impact factor: 3.415

9.  Flow limitation as a noninvasive assessment of residual upper-airway resistance during continuous positive airway pressure therapy of obstructive sleep apnea.

Authors:  R Condos; R G Norman; I Krishnasamy; N Peduzzi; R M Goldring; D M Rapoport
Journal:  Am J Respir Crit Care Med       Date:  1994-08       Impact factor: 21.405

Review 10.  Pressure modification for improving usage of continuous positive airway pressure machines in adults with obstructive sleep apnoea.

Authors:  Ian Smith; Toby J Lasserson
Journal:  Cochrane Database Syst Rev       Date:  2009-10-07
View more
  4 in total

1.  Is there an optimal nasal pressure for treating obstructive sleep apnea-and if so, what is it?

Authors:  Ramon Farré; Daniel Navajas; Josep M Montserrat
Journal:  Sleep       Date:  2013-04-01       Impact factor: 5.849

2.  Short-term effects of a vibrotactile neck-based treatment device for positional obstructive sleep apnea: preliminary data on tolerability and efficacy.

Authors:  Simone Scarlata; Isaura Rossi Bartoli; Simona Santangelo; Gilda Giannunzio; Claudio Pedone; Raffaele Antonelli Incalzi
Journal:  J Thorac Dis       Date:  2016-07       Impact factor: 2.895

3.  Auto-titrating CPAP for the treatment of obstructive sleep apnea in children.

Authors:  Ilya Khaytin; Ignacio E Tapia; Melissa S Xanthopoulos; Christopher Cielo; Ji Young Kim; Julianne Smith; Edward C Matthews; Suzanne E Beck
Journal:  J Clin Sleep Med       Date:  2020-06-15       Impact factor: 4.062

4.  Alternative Procedure to Individual Nasal Pressure Titration for Sleep Apnea.

Authors:  Ramon Farré; David Gozal; Josep M Montserrat
Journal:  J Clin Med       Date:  2021-04-01       Impact factor: 4.241

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.