Temperature imaging of laser-induced thermotherapy (LITT) by MRI: evaluation of different sequences in phantom.

Babak Bazrafshan1, Frank Hübner, Parviz Farshid, Renate Hammerstingl, Jijo Paul, Vitali Vogel, Werner Mäntele, Thomas J Vogl.   

Abstract

The purpose of this study was to evaluate magnetic resonance (MR) temperature imaging of the laser-induced thermotherapy (LITT) comparing the proton resonance frequency (PRF) and T 1 thermometry methods. LITT was applied to a liver-mimicking acrylamide gel phantom. Temperature rise up to 70 °C was measured using a MR-compatible fiber-optic thermometer. MR imaging was performed by a 1.5-T scanner utilizing fast gradient echo sequences including a segmented echo planar imaging (seg-EPI) sequence for PRF and the following sequences for T 1 method: fast low-angle shot (FLASH), inversion recovery turbo flash (IRTF), saturation recovery turbo flash (SRTF), and true fast imaging (TRUFI). Temperature-induced change of the pixel values in circular regions of interest, selected on images under the temperature probe tip, was recorded. For each sequence, a calibration constant could be determined to be -0.0088 ± 0.0002 ppm °C(-1) (EPI), -1.15 ± 0.03 °C(-1) (FLASH), -1.49 ± 0.03 °C(-1) (IRTF), -1.21 ± 0.03 °C(-1) (SRTF), and -2.52 ± 0.12 °C(-1) (TRUFI). These constants were evaluated in further LITT experiments in phantom comparing the calculated temperatures with the fiber optic-measured ones; temperature precisions of 0.60 °C (EPI), 0.81 °C (FLASH), 1.85 °C (IRTF), 1.95 °C (SRTF), and 3.36 °C (TRUFI) were obtained. Furthermore, performing the Bland-Altman analysis, temperature accuracy was determined to be 0.23 °C (EPI), 0.31 °C (FLASH), 1.66 °C (IRTF), 1.19 °C (SRTF), and 3.20 °C (TRUFI). In conclusion, the seg-EPI sequence was found to be more convenient for MR temperature imaging of LITT due to its relatively high precision and accuracy. Among the T 1 method sequences, FLASH showed the highest accuracy and robustness.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23535892     DOI: 10.1007/s10103-013-1306-5

Source DB:  PubMed          Journal:  Lasers Med Sci        ISSN: 0268-8921            Impact factor:   3.161


  44 in total

1.  Temperature measurement using echo-shifted FLASH at low field for interventional MRI.

Authors:  Y C Chung; J L Duerk; A Shankaranarayanan; M Hampke; E M Merkle; J S Lewin
Journal:  J Magn Reson Imaging       Date:  1999-01       Impact factor: 4.813

Review 2.  Assessing extracranial tumors using diffusion-weighted whole-body MRI.

Authors:  Claudia Lenz; Markus Klarhöfer; Klaus Scheffler; Leopold Winter; Gregor Sommer
Journal:  Z Med Phys       Date:  2010-10-02       Impact factor: 4.820

Review 3.  MR thermometry.

Authors:  Viola Rieke; Kim Butts Pauly
Journal:  J Magn Reson Imaging       Date:  2008-02       Impact factor: 4.813

Review 4.  MR thermometry for monitoring tumor ablation.

Authors:  Baudouin Denis de Senneville; Charles Mougenot; Bruno Quesson; Iulius Dragonu; Nicolas Grenier; Chrit T W Moonen
Journal:  Eur Radiol       Date:  2007-05-22       Impact factor: 5.315

5.  Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences.

Authors:  A Kickhefel; Jörg Roland; Clifford Weiss; Fritz Schick
Journal:  Phys Med       Date:  2010-01-21       Impact factor: 2.685

6.  MRI thermometry in phantoms by use of the proton resonance frequency shift method: application to interstitial laser thermotherapy.

Authors:  J Olsrud; R Wirestam; S Brockstedt; A M Nilsson; K G Tranberg; F Ståhlberg; B R Persson
Journal:  Phys Med Biol       Date:  1998-09       Impact factor: 3.609

7.  Assessment of radiofrequency self-heating around a metallic wire with MR T1-based thermometry.

Authors:  V Detti; D Grenier; E Perrin; O Beuf
Journal:  Magn Reson Med       Date:  2011-02-28       Impact factor: 4.668

8.  Safety and efficacy of microwave ablation of hepatic tumors: a prospective review of a 5-year experience.

Authors:  Robert C G Martin; Charles R Scoggins; Kelly M McMasters
Journal:  Ann Surg Oncol       Date:  2009-08-26       Impact factor: 5.344

9.  Microwave ablation of lung malignancies: effectiveness, CT findings, and safety in 50 patients.

Authors:  Farrah J Wolf; David J Grand; Jason T Machan; Thomas A Dipetrillo; William W Mayo-Smith; Damian E Dupuy
Journal:  Radiology       Date:  2008-03-27       Impact factor: 11.105

10.  Ultrasound-guided percutaneous thermal ablation of hepatocellular carcinoma using microwave and radiofrequency ablation.

Authors:  H-X Xu; X-Y Xie; M-D Lu; J-W Chen; X-Y Yin; Z-F Xu; G-J Liu
Journal:  Clin Radiol       Date:  2004-01       Impact factor: 2.350

View more
  9 in total

1.  Effect of size, concentration, and type of spherical gold nanoparticles on heat evolution following laser irradiation using tissue-simulating phantoms.

Authors:  Hossam Zakaria; Wessameldin S Abdelaziz; Tareq Youssef
Journal:  Lasers Med Sci       Date:  2016-02-09       Impact factor: 3.161

2.  Optothermal profile of an ablation catheter with integrated microcoil for MR-thermometry during Nd:YAG laser interstitial thermal therapies of the liver—an in-vitro experimental and theoretical study.

Authors:  Evdokia M Kardoulaki; Richard R A Syms; Ian R Young; Kaushal Choonee; Marc Rea; Wladyslaw M W Gedroyc
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

3.  Validation of a mathematical model for laser-induced thermotherapy in liver tissue.

Authors:  F Hübner; C Leithäuser; B Bazrafshan; N Siedow; T J Vogl
Journal:  Lasers Med Sci       Date:  2017-06-22       Impact factor: 3.161

4.  Magnetic resonance temperature imaging of laser-induced thermotherapy using proton resonance frequency shift: evaluation of different sequences in phantom and porcine brain at 7 T.

Authors:  Guojun Xu; Zhiyong Zhao; Kedi Xu; Junming Zhu; Anna W Roe; Bin Xu; Xiaotong Zhang; Jianqi Li; Dongrong Xu
Journal:  Jpn J Radiol       Date:  2022-04-17       Impact factor: 2.701

5.  Improved MR thermometry for laser interstitial thermotherapy.

Authors:  Henrik Odéen; Dennis L Parker
Journal:  Lasers Surg Med       Date:  2019-01-15       Impact factor: 4.025

Review 6.  Advances in Magnetic Noble Metal/Iron-Based Oxide Hybrid Nanoparticles as Biomedical Devices.

Authors:  Laura M Sanchez; Vera A Alvarez
Journal:  Bioengineering (Basel)       Date:  2019-08-28

7.  Optimization of sapphire capillary needles for interstitial and percutaneous laser medicine.

Authors:  Irina N Dolganova; Irina A Shikunova; Gleb M Katyba; Arsen K Zotov; Elena E Mukhina; Marina A Shchedrina; Valery V Tuchin; Kirill I Zaytsev; Vladimir N Kurlov
Journal:  J Biomed Opt       Date:  2019-12       Impact factor: 3.170

8.  Quantitative MR thermometry based on phase-drift correction PRF shift method at 0.35 T.

Authors:  Yuping Chen; Mengke Ge; Rizwan Ali; Hejun Jiang; Xiaoyan Huang; Bensheng Qiu
Journal:  Biomed Eng Online       Date:  2018-04-10       Impact factor: 2.819

9.  Fiber Bragg Grating Sensors for Performance Evaluation of Fast Magnetic Resonance Thermometry on Synthetic Phantom.

Authors:  Martina De Landro; Jacopo Ianniello; Maxime Yon; Alexey Wolf; Bruno Quesson; Emiliano Schena; Paola Saccomandi
Journal:  Sensors (Basel)       Date:  2020-11-12       Impact factor: 3.576

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.