OBJECTIVES/HYPOTHESIS: This pilot study details the use of a software tool that uses continuous impedance measurement during electrode insertion, with the eventual potential to assess and optimize electrode position and reduce insertional trauma. STUDY
DESIGN: Software development and experimental study with human cadaveric cochleae and two live surgeries.
METHODS: A prototype program to measure intracochlear electrode impedance and display it graphically in real time has been developed. The software was evaluated in human cadaveric temporal bones while simultaneously making real-time fluoroscopic recordings and in two live surgeries during intracochlear electrode insertion.
RESULTS: Impedance changes were observed with various scalar positions, and values were consistent with those obtained using clinically available software. Using Contour Advance electrodes, impedance values increased after stylet removal, particularly when using the monopolar mode.
CONCLUSIONS: Impedance values seem systematically affected by electrode position, with higher values being associated with proximity to the cochlear wall. The new software is capable of acquiring impedance measurements during electrode insertion, and these data may be useful to guide surgeons to achieve optimal and atraumatic electrode insertion, to guide robotic electrode insertion, and to provide insights about electrode position in the cochlea.
Copyright © 2013 The American Laryngological, Rhinological, and Otological Society, Inc.

Entities:  

Mesh:

Year:  2013        PMID: 23529884      PMCID: PMC3616339          DOI: 10.1002/lary.23714

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  8 in total

1.  Inroads toward robot-assisted cochlear implant surgery using steerable electrode arrays.

Authors:  Jian Zhang; Wei Wei; Jienan Ding; J Thomas Roland; Spiros Manolidis; Nabil Simaan
Journal:  Otol Neurotol       Date:  2010-10       Impact factor: 2.311

2.  A new software tool to optimize frequency table selection for cochlear implants.

Authors:  Daniel Jethanamest; Chin-Tuan Tan; Matthew B Fitzgerald; Mario A Svirsky
Journal:  Otol Neurotol       Date:  2010-10       Impact factor: 2.311

3.  Electrode impedance in adults and children using the Nucleus 24 cochlear implant system.

Authors:  P A Busby; K L Plant; L A Whitford
Journal:  Cochlear Implants Int       Date:  2002-09

4.  Neural response thresholds in the Nucleus Contour cochlear implant before and after stylet removal.

Authors:  Robinson Koji Tsuji; Maria Valéria Schmidt Goffi-Gomez; Cristina Ornelas Peralta; Mariana Cardoso Guedes; Ana Tereza Magalhães; Rubens Brito Neto; Ricardo Ferreira Bento
Journal:  Acta Otolaryngol       Date:  2009-11       Impact factor: 1.494

5.  A computer simulation of the generation and distribution of cochlear potentials.

Authors:  D Strelioff
Journal:  J Acoust Soc Am       Date:  1973-09       Impact factor: 1.840

6.  Potential distributions and neural excitation patterns in a rotationally symmetric model of the electrically stimulated cochlea.

Authors:  J H Frijns; S L de Snoo; R Schoonhoven
Journal:  Hear Res       Date:  1995-07       Impact factor: 3.208

7.  Role of electrode placement as a contributor to variability in cochlear implant outcomes.

Authors:  Charles C Finley; Timothy A Holden; Laura K Holden; Bruce R Whiting; Richard A Chole; Gail J Neely; Timothy E Hullar; Margaret W Skinner
Journal:  Otol Neurotol       Date:  2008-10       Impact factor: 2.311

8.  Design of a Tool Integrating Force Sensing With Automated Insertion in Cochlear Implantation.

Authors:  Daniel Schurzig; Robert F Labadie; Andreas Hussong; Thomas S Rau; Robert J Webster
Journal:  IEEE ASME Trans Mechatron       Date:  2011-02-22       Impact factor: 5.303

  8 in total
  5 in total

1.  Durability of Hearing Preservation after Cochlear Implantation with Conventional-Length Electrodes and Scala Tympani Insertion.

Authors:  Alex D Sweeney; Jacob B Hunter; Matthew L Carlson; Alejandro Rivas; Marc L Bennett; Rene H Gifford; Jack H Noble; David S Haynes; Robert F Labadie; George B Wanna
Journal:  Otolaryngol Head Neck Surg       Date:  2016-02-23       Impact factor: 3.497

2.  Scanning electrochemical microscopy as a novel proximity sensor for atraumatic cochlear implant insertion.

Authors:  H Watanabe; J Velmurugan; M V Mirkin; M A Svirsky; A K Lalwani; R R Llinas
Journal:  IEEE Trans Biomed Eng       Date:  2014-06       Impact factor: 4.538

3.  Real-Time Localization of Cochlear-Implant Electrode Arrays Using Bipolar Impedance Sensing.

Authors:  Trevor L Bruns; Katherine E Riojas; Robert F Labadie; Robert J Webster Iii
Journal:  IEEE Trans Biomed Eng       Date:  2022-01-20       Impact factor: 4.538

4.  Insertion Guidance Based on Impedance Measurements of a Cochlear Electrode Array.

Authors:  Enver Salkim; Majid Zamani; Dai Jiang; Shakeel R Saeed; Andreas Demosthenous
Journal:  Front Comput Neurosci       Date:  2022-06-23       Impact factor: 3.387

5.  Impedance Measures During in vitro Cochlear Implantation Predict Array Positioning.

Authors:  Christopher Kenneth Giardina; Elliot Samuel Krause; Kanthaiah Koka; Douglas Carl Fitzpatrick
Journal:  IEEE Trans Biomed Eng       Date:  2018-02       Impact factor: 4.538

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.