Literature DB >> 23523513

Influence of different frequencies and insertion depths on the diagnostic accuracy of liver elastography by acoustic radiation force impulse imaging (ARFI).

Andrej Potthoff1, Dina Attia, Sven Pischke, Janina Kirschner, Ingmar Mederacke, Heiner Wedemeyer, Michael P Manns, Michael J Gebel, Kinan Rifai.   

Abstract

BACKGROUND: Acoustic Radiation Force Impulse Imaging (ARFI) is an innovative elastography for staging of liver fibrosis. We evaluated the diagnostic accuracy of different probes to perform ARFI at different insertion depths.
METHODS: In a prospective study, 89 chronic HCV infected patients underwent ARFI elastography using both available probes (c-ARFI: C4-1-MHz; l-ARFI: L9-4 MHz) in comparison to Fibroscan(®). Variability of ARFI elastography at different insertion depths was systematically evaluated in 39 patients (44%). According to Fibroscan(®) elastography, 32 patients (36%) presented with liver cirrhosis, 23 patients (26%) had significant fibrosis and 34 patients (38%) had no significant fibrosis.
RESULTS: Mean propagation velocity with c-ARFI was 1.70±0.67m/s and 1.91±0.87m/s with l-ARFI. Results of both probes were correlated to each other (p<0.001; r=0.70) and to Fibroscan(®) (p<0.001, r=0.82 and 0.84, respectively). In patients with significant fibrosis or with cirrhosis, mean values by l-ARFI were significantly higher than by c-ARFI (p<0.001). For detection of liver cirrhosis, AUROC was 0.97 for c-ARFI (cut-off level 1.72m/s) and 0.90 for l-ARFI (cut-off 2.04m/s). Correlation coefficients of c-ARFI with Fibroscan(®) were highest at an insertion depth of 5-6cm (r=0.882 and 0.864, respectively, p<0.001) and at 3-4cm for l-ARFI (r=0.850 and 0.838, respectively, p<0.001).
CONCLUSIONS: ARFI elastography with the linear and with the convex probes showed comparable validity and accuracy in the estimation of liver stiffness. The linear probe gave higher ARFI values. The most accurate insertion depth was 5-6cm for c-ARFI and 3-4cm for l-ARFI indicating that measurements should not be performed close to the liver capsule.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23523513     DOI: 10.1016/j.ejrad.2013.02.011

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  16 in total

Review 1.  [Ultrasound elastography].

Authors:  R S Goertz
Journal:  Radiologe       Date:  2015-11       Impact factor: 0.635

2.  Diagnostic Performance of MR Elastography and Vibration-controlled Transient Elastography in the Detection of Hepatic Fibrosis in Patients with Severe to Morbid Obesity.

Authors:  Jun Chen; Meng Yin; Jayant A Talwalkar; Jennifer Oudry; Kevin J Glaser; Thomas C Smyrk; Véronique Miette; Laurent Sandrin; Richard L Ehman
Journal:  Radiology       Date:  2016-11-18       Impact factor: 11.105

3.  Guidelines for Finite-Element Modeling of Acoustic Radiation Force-Induced Shear Wave Propagation in Tissue-Mimicking Media.

Authors:  Mark L Palmeri; Bo Qiang; Shigao Chen; Matthew W Urban
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2016-12-21       Impact factor: 2.725

4.  Shear wave elastography using sound touch elastography and supersonic shear imaging for liver measurements: a comparative study.

Authors:  Hongjin Xiang; Wenwu Ling; Lin Ma; Lulu Yang; Tang Lin; Yan Luo
Journal:  Quant Imaging Med Surg       Date:  2022-05

5.  Liver reserve function assessment by acoustic radiation force impulse imaging.

Authors:  Xiao-Lan Sun; Li-Wei Liang; Hui Cao; Qiong Men; Ke-Zhu Hou; Zhen Chen; Ya-E Zhao
Journal:  World J Gastroenterol       Date:  2015-08-28       Impact factor: 5.742

6.  The Role of Spleen Stiffness in Determining the Severity and Bleeding Risk of Esophageal Varices in Cirrhotic Patients.

Authors:  Hwi Young Kim; Eun Hyo Jin; Won Kim; Jae Young Lee; Hyunsik Woo; Sohee Oh; Ji-Yeon Seo; Hong Sang Oh; Kwang Hyun Chung; Yong Jin Jung; Donghee Kim; Byeong Gwan Kim; Kook Lae Lee
Journal:  Medicine (Baltimore)       Date:  2015-06       Impact factor: 1.889

7.  Shear wave elastography in chronic kidney disease: a pilot experience in native kidneys.

Authors:  Anthony E Samir; Andrew S Allegretti; Qingli Zhu; Manish Dhyani; Arash Anvari; Dorothy A Sullivan; Caitlin A Trottier; Sarah Dougherty; Winfred W Williams; Jodie L Babitt; Julia Wenger; Ravi I Thadhani; Herbert Y Lin
Journal:  BMC Nephrol       Date:  2015-07-31       Impact factor: 2.388

8.  Evaluation of Transient Elastography, Acoustic Radiation Force Impulse Imaging (ARFI), and Enhanced Liver Function (ELF) Score for Detection of Fibrosis in Morbidly Obese Patients.

Authors:  Thomas Karlas; Arne Dietrich; Veronica Peter; Christian Wittekind; Ralf Lichtinghagen; Nikita Garnov; Nicolas Linder; Alexander Schaudinn; Harald Busse; Christiane Prettin; Volker Keim; Michael Tröltzsch; Tatjana Schütz; Johannes Wiegand
Journal:  PLoS One       Date:  2015-11-03       Impact factor: 3.240

9.  Influence of age, sex, body mass index, alcohol, and smoking on shear wave velocity (p-SWE) of the pancreas.

Authors:  Sabina Stumpf; Heike Jaeger; Tilmann Graeter; Suemeyra Oeztuerk; Julian Schmidberger; Mark Martin Haenle; Wolfgang Kratzer
Journal:  Abdom Radiol (NY)       Date:  2016-07

Review 10.  Evaluation of kidney allograft status using novel ultrasonic technologies.

Authors:  Cheng Yang; Mushuang Hu; Tongyu Zhu; Wanyuan He
Journal:  Asian J Urol       Date:  2015-07-06