Shear wave elastography using sound touch elastography and supersonic shear imaging for liver measurements: a comparative study.

Hongjin Xiang1, Wenwu Ling1, Lin Ma1, Lulu Yang1, Tang Lin1, Yan Luo1.   

Abstract

Background: Measurements of liver stiffness obtained with 2-dimensional shear wave elastography (2D-SWE) have been widely used to clinically assess liver fibrosis. However, differences between different 2D-SWE systems can lead to confusion when interpreting measurements. This study investigated the variability between a recently released sound touch elastography (STE) system and a supersonic shear imaging (SSI) system and assessed the degree of intersystem discrepancy using the different liver stiffness value (LSV) thresholds recommended by the Society of Radiologists in Ultrasound (SRU) for assessing liver fibrosis.
Methods: A total of 4,152 patients who had undergone STE and SSI on the same day were enrolled in this retrospective study. First, intrasystem agreement for STE and SSI was assessed. Then, intraclass correlation coefficients (ICCs) and the Bland-Altman method were used to assess intersystem variability for all cases, classified according to the thresholds recommended by the SRU. The effects of age, gender, and body-mass index (BMI) were evaluated using multivariate linear regression analysis and attributive intervals were computed for STE and SSI at each of the different thresholds.
Results: The ICCs for STE and SSI intrasystem agreement were 0.94 [95% confidence interval (CI): 0.937-0.943; P<0.001] and 0.984 (95% CI: 0.984-0.985; P<0.001), respectively. The 95% limit of agreement (LOA) for all cases ranged from -6.96 to 7.44 kPa. The 95% LOA increased as the threshold values rose, and intersystem variability was obvious, even at the smallest threshold (the 95% LOA at values ≤5 kPa was -0.85 to 2.08 kPa, while that at values >17 kPa was -20.81 to 14.71 kPa). The adjusted R2 for age, gender, and BMI was only 0.018 (all P value <0.05). Conclusions: There was clear variability between STE and SSI, in contrast with some previous studies with small sample sizes, and consistent with others. Intersystem variability increased with the elevation of the LSV thresholds recommended by the SRU. Gender and BMI had little effect on intersystem variability. Future research could compare STE and SSI in different liver diseases, assessing the feasibility of the SRU-recommended thresholds in proven pathologies and evaluating the test-retest repeatability. 2022 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  Sound touch elastography (STE); intersystem variability; rule of four; supersonic shear imaging (SSI)

Year:  2022        PMID: 35502377      PMCID: PMC9014163          DOI: 10.21037/qims-21-863

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  25 in total

1.  Reproducibility of 2 Liver 2-Dimensional Shear Wave Elastographic Techniques in the Fasting and Postprandial States.

Authors:  Paul Simkin; Anand Rattansingh; Kuan Liu; John M Hudson; Mostafa Atri; Hyun-Jung Jang; Tae Kyoung Kim; Korosh Khalili
Journal:  J Ultrasound Med       Date:  2018-12-11       Impact factor: 2.153

2.  EFSUMB Guidelines and Recommendations on the Clinical Use of Liver Ultrasound Elastography, Update 2017 (Long Version).

Authors:  Christoph F Dietrich; Jeffrey Bamber; Annalisa Berzigotti; Simona Bota; Vito Cantisani; Laurent Castera; David Cosgrove; Giovanna Ferraioli; Mireen Friedrich-Rust; Odd Helge Gilja; Ruediger Stephan Goertz; Thomas Karlas; Robert de Knegt; Victor de Ledinghen; Fabio Piscaglia; Bogdan Procopet; Adrian Saftoiu; Paul S Sidhu; Ioan Sporea; Maja Thiele
Journal:  Ultraschall Med       Date:  2017-04-13       Impact factor: 6.548

Review 3.  Liver Ultrasound Elastography: An Update to the World Federation for Ultrasound in Medicine and Biology Guidelines and Recommendations.

Authors:  Giovanna Ferraioli; Vincent Wai-Sun Wong; Laurent Castera; Annalisa Berzigotti; Ioan Sporea; Christoph F Dietrich; Byung Ihn Choi; Stephanie R Wilson; Masatoshi Kudo; Richard G Barr
Journal:  Ultrasound Med Biol       Date:  2018-09-09       Impact factor: 2.998

4.  The Asian Pacific Association for the Study of the Liver clinical practice guidelines for the diagnosis and management of metabolic associated fatty liver disease.

Authors:  Mohammed Eslam; Shiv K Sarin; Vincent Wai-Sun Wong; Jian-Gao Fan; Takumi Kawaguchi; Sang Hoon Ahn; Ming-Hua Zheng; Gamal Shiha; Yusuf Yilmaz; Rino Gani; Shahinul Alam; Yock Young Dan; Jia-Horng Kao; Saeed Hamid; Ian Homer Cua; Wah-Kheong Chan; Diana Payawal; Soek-Siam Tan; Tawesak Tanwandee; Leon A Adams; Manoj Kumar; Masao Omata; Jacob George
Journal:  Hepatol Int       Date:  2020-10-01       Impact factor: 6.047

5.  Liver Stiffness Values Are Lower in Pediatric Subjects than in Adults and Increase with Age: A Multifrequency MR Elastography Study.

Authors:  Emily Etchell; Lauriane Jugé; Alice Hatt; Ralph Sinkus; Lynne E Bilston
Journal:  Radiology       Date:  2016-10-17       Impact factor: 11.105

6.  Repeatability and Agreement of Shear Wave Speed Measurements in Phantoms and Human Livers Across 6 Ultrasound 2-Dimensional Shear Wave Elastography Systems.

Authors:  Leah A Gilligan; Andrew T Trout; Paula Bennett; Jonathan R Dillman
Journal:  Invest Radiol       Date:  2020-04       Impact factor: 6.016

7.  Shear wave-based sound touch elastography in liver fibrosis assessment for patients with autoimmune liver diseases.

Authors:  Lulu Yang; Wenwu Ling; Du He; Changli Lu; Lin Ma; Lin Tang; Yan Luo; Shigao Chen
Journal:  Quant Imaging Med Surg       Date:  2021-04

8.  Usefulness of New Shear Wave Elastography Technique for Noninvasive Assessment of Liver Fibrosis in Patients with Chronic Hepatitis B: A Prospective Multicenter Study.

Authors:  Jinfen Wang; Manli Wu; Runze Linghu; Jiandong Chang; Min Wu; Cheng Feng; Xinping Ren; Changzhu Liu; Jie Lin; Tao Song; Jionghui Gu; Yao Zhang; Ye Fang; Sumei Ma; Pingxiang Hu; Linsong Wu; Xue Han; Kaixuan Chen; Qiusheng Shi; Ruifang Zhang; Qi Zhou; Ruiqing Du; Yongyan Gao; Xiang Jing; Shunshi Yang; Chang Zhou; Jian Zheng; Ping Liang; Rong-Qin Zheng
Journal:  Ultraschall Med       Date:  2021-04-28       Impact factor: 6.548

9.  Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique.

Authors:  Hwaseong Ryu; Su Joa Ahn; Jeong Hee Yoon; Jeong Min Lee
Journal:  Ultrasonography       Date:  2018-11-23

10.  Intra- and inter-operator reproducibility of US point shear-wave elastography in various organs: evaluation in phantoms and healthy volunteers.

Authors:  Riwa Kishimoto; Katsuhiko Kikuchi; Atsuhisa Koyama; Jeff Kershaw; Tokuhiko Omatsu; Yasuhiko Tachibana; Mikio Suga; Takayuki Obata
Journal:  Eur Radiol       Date:  2019-05-14       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.