| Literature DB >> 23523499 |
Kiyokazu Fukui1, Ayumi Kaneuji, Tanzo Sugimori, Toru Ichiseki, Tadami Matsumoto.
Abstract
We previously reported the results of wear comparison at a minimum of 5 years between highly cross-linked polyethylene (HXLPE) and conventional polyethylene (PE) against a zirconia femoral head. We now report the mean wear at 10 years for 52 patients (56 hips) of the original cohort of 61 patients (65 hips) who had undergone primary total hip arthroplasty at our hospital between November 1999 and August 2000. The mean steady-state linear wear rate of HXLPE was 0.045 mm/y, compared with 0.080 mm/y for conventional PE (P=.0003). The incidence of osteolysis was 25% in the conventional PE group compared with 0% in the HXLPE group. Our study demonstrated that the steady-state wear rate for HXLPE remains significantly lower than that for conventional PE against a zirconia femoral head at a mean of 10 years after implantation.Entities:
Keywords: conventional polyethylene; highly cross-linked polyethylene; steady-state wear; zirconia femoral head
Mesh:
Substances:
Year: 2013 PMID: 23523499 DOI: 10.1016/j.arth.2012.12.020
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757