Comparison of the new rebound tonometer with Goldmann applanation tonometer in a clinical setting.

Kyoung Nam Kim1, Jin Wook Jeoung, Ki Ho Park, Min Kyu Yang, Dong Myung Kim.   

Abstract

PURPOSE: To evaluate the clinical usefulness of a new rebound tonometer, Icare(®) PRO (Icare PRO), by comparison with Goldmann applanation tonometry (GAT) in a study on patients with glaucoma.
METHODS: One hundred and seventy-two eyes of 86 subjects were enrolled in this study. All of the subjects were examined with an autorefractometer, Icare PRO, slit-lamp biomicroscope, GAT, ultrasound A-scan and pachymeter. Three intraocular pressure (IOP) measurements were obtained by Icare PRO and GAT. The intraobserver reliabilities were established by calculating the intraclass correlation coefficients. The Bland-Altman plot was used to compare the Icare PRO and GAT.
RESULTS: There was a good correlation between the IOP measurement by GAT and that by Icare PRO (r = 0.6995, p < 0.001). The intraclass correlation coefficients of Icare PRO and GAT were 0.778 and 0.955, respectively. The IOP differences between Icare PRO and GAT (mean: 1.92 mmHg; SD: 3.29 mmHg; 95% limit of agreement: -4.52 to 8.37 mmHg) did not vary over the wide range of central corneal thickness (p = 0.498), age (p = 0.248), axial length (p = 0.277) or spherical equivalent (p = 0.075).
CONCLUSIONS: Although IOP with Icare PRO was higher than that with GAT, especially at lower GAT IOP value, Icare PRO was found to be a reliable method and showed a good correlation with GAT. The IOP difference between Icare PRO and GAT was not affected by the central corneal thickness, age, axial length or spherical equivalent. Icare PRO can be expected not only to be a good screening tool but also to be a good substitute for GAT.
© 2013 The Authors Acta Ophthalmologica © 2013 Acta Ophthalmologica Scandinavica Foundation. Published by Blackwell Publishing Ltd.

Entities:  

Keywords:  Goldmann applanation tonometer; Icare rebound tonometer; glaucoma; intraocular pressure

Mesh:

Year:  2013        PMID: 23521889     DOI: 10.1111/aos.12109

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  18 in total

1.  Early and frequent development of ocular hypertension in children with nephrotic syndrome.

Authors:  Emi Kawaguchi; Kenji Ishikura; Riku Hamada; Yoshinobu Nagaoka; Yoshihiko Morikawa; Tomoyuki Sakai; Yuko Hamasaki; Hiroshi Hataya; Eiichiro Noda; Masaru Miura; Takashi Ando; Masataka Honda
Journal:  Pediatr Nephrol       Date:  2014-05-24       Impact factor: 3.714

2.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

3.  Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Authors:  Senay Asik Nacaroglu; Emine Seker Un; Mehmet Giray Ersoz; Yelda Tasci
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

Review 4.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

5.  Comparison of Diaton transpalpebral tonometer with applanation tonometry in keratoconus.

Authors:  Robert Pl Wisse; Natalie Peeters; Saskia M Imhof; Allegonda van der Lelij
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

6.  One Year of Glaucoma Research in Review: 2012 to 2013.

Authors:  Charles Kim; Anna M Demetriades; Nathan M Radcliffe
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2014 Jan-Feb

7.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

Review 8.  Systematic Review of Current Devices for 24-h Intraocular Pressure Monitoring.

Authors:  Sabita M Ittoop; Jeffrey R SooHoo; Leonard K Seibold; Kaweh Mansouri; Malik Y Kahook
Journal:  Adv Ther       Date:  2016-08-16       Impact factor: 3.845

9.  Comparison of self-measured diurnal intraocular pressure profiles using rebound tonometry between primary angle closure glaucoma and primary open angle glaucoma patients.

Authors:  Shaoying Tan; Nafees Baig; Linda Hansapinyo; Vishal Jhanji; Shihui Wei; Clement C Tham
Journal:  PLoS One       Date:  2017-03-23       Impact factor: 3.240

10.  Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.

Authors:  Blanche D Rodrigues; Fabiano Montiani-Ferreira; Mariza Bortolini; André T Somma; András M Komáromy; Peterson Triches Dornbusch
Journal:  Vet Ophthalmol       Date:  2020-10-17       Impact factor: 1.644

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.