| Literature DB >> 23516268 |
Simon Chun Ho Yu1, Thomas Wai Hong Leung, Kwok Tung Lee, Lawrence Ka Sing Wong.
Abstract
BACKGROUND: Symptomatic brain hemorrhage was a significant cause of periprocedural stroke or death following stenting in the Stenting and Aggressive Medical Management for Preventing Recurrent Stroke in Intracranial Stenosis trial, which called into question the safety of Wingspan stenting for intracranial atherosclerosis. This study analyzed the role of a learning curve in the safety and outcome of Wingspan stenting from the experience of 95 consecutive patients at a single center.Entities:
Keywords: Angioplasty; Atherosclerosis; Intervention; Stent; Stroke
Mesh:
Substances:
Year: 2013 PMID: 23516268 PMCID: PMC3963535 DOI: 10.1136/neurintsurg-2012-010593
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Technical problems encountered during procedure and causes of procedure failure
| Patient sequence | Stenosis location | Stenosis degree | Technical problem encountered during the procedure | Cause of procedure failure |
|---|---|---|---|---|
| 4 | ICA and MCA | 50% | Difficult stent tracking and delivery, fracture core of stent delivery system | |
| 5 | ICA | 60% | Nose cone bending caused stent sheath kinking and stretching during stent deployment | |
| 7 | ICA | 60% | Stretching of stent sheath during stent deployment | |
| 8 | BA | 75% | Arterial rupture after angioplasty | |
| 15 | ICA and MCA | 50% | TIA during manipulation of balloon catheter across stenosis | TIA |
| 27 | ICA | 90% | Nose cone trapped between guidewire and stent tip after stent deployment | |
| 42 | ICA | 75% | Difficult stent tracking and delivery, fracture core of stent delivery system, TIA | TIA |
| 44 | MCA | 83% | TIA | |
| 47 | ICA and MCA | 96% | Failure passage of balloon across stenosis | Failure passage of balloon |
| 48 | MCA | 90% | Kinking of balloon catheter after guidewire removal | |
| 56 | MCA | 70% | Transient blindness and severe headache during stent system advancement | |
| 59 | MCA | 87% | Transient blindness during stent system advancement | TIA |
| 66 | ICA | 75% | Perforation of M2 perforators by guidewire during angioplasty |
BA, basilar artery; ICA, internal carotid artery; MCA, middle cerebral artery; TIA, transient ischemic attack.
Stroke or death after 30 days in patients with Wingspan successfully placed
| Incidence (%) of events occurring during the period of assessment | |||||
|---|---|---|---|---|---|
| Event | 30 days to 1 year | 2nd year | 3rd year | 4th year | 5th year |
| No of patients evaluated | 75 | 63 | 49 | 37 | 17 |
| Major hemorrhagic stroke in related territory (mRS>3) | 0 | 0 | 0 | 0 | 0 |
| Major ischemic stroke in related territory (mRS>3) | 0 | 0 | 0 | 0 | 0 |
| Minor hemorrhagic stroke in related territory (mRS≤3) | 0 | 0 | 0 | 0 | 0 |
| Minor ischemic stroke in related territory (mRS≤3) | 0 | 0 | 0 | 1 | 1 |
| Minor ischemic stroke in unrelated territory (mRS≤3) | 3 | 0 | 0 | 0 | 0 |
| Stroke-related death | 0 | 0 | 0 | 0 | 0 |
| Unrelated death | 1 | 1 | 2 | 1 | 1 |
mRS, modified Rankin Scale.
Figure 1Occurrence of study endpoints in chronological sequence.
Leaning curve analysis using data splitting method
| Variables | 1st quarter | 2nd quarter | 3rd quarter | 4th quarter | Trend difference across quarters |
|---|---|---|---|---|---|
| Patient sequence (patient no) | 1–24 (24) | 25–48 (24) | 49–72 (24) | 73–95 (23) | |
| Age (mean±SD) | 67.46±10.484 | 62.75±11.433 | 62.13±10.999 | 66.57±9.336 | No difference across quartersp>0.085 |
| 95% CI | 63.03 to 71.89 | 57.92 to 67.58 | 57.48 to 66.77 | 62.53 to 70.6 | |
| Degree of stenosis (mean±SD) | 69.96±12.539 | 77.82±8.368 | 76.39±9.942 | 77.91±8.286 | Difference between 1st and 2nd quarter, p=0.033. No difference across other quarters |
| 95% CI | 64.53 to 75.38 | 74.29 to 81.35 | 72.19 to 80.59 | 74.24 to 81.58 | |
| Location | |||||
| MCA | 10 | 16 | 15 | 10 | Difference between 1st and 2nd quarter in frequency of location at VBA, p=0.023. No difference across other quarters |
| ICA | 6 | 7 | 5 | 8 | |
| VBA | 8 | 1 | 4 | 5 | |
| Event to stenting <30 days (rate) | 10 (41.7%) | 15 (62.5%) | 14 (58.3%) | 18 (78.3%) | Difference between 1st and 4th quarter, p=0.011. Difference between 4th quarter and the rest, p=0.05 |
| Procedural problem (rate) | 4 (16.7%) | 5 (20.8%) | 2 (8.3%) | 0 | Difference between 4th quarter and the rest, p=0.046 |
| Technical failure (rate) | 2 (8.3%) | 2 (8.3%) | 2 (8.3%) | 0 | No difference across quarters,p>0.33 |
| Guidewire- or angioplasty-related hemorrhagic complication <30 days (rate) | 1 (4.2%) | 1 (4.2%) | 1 (4.2%) | 0 | No difference across quarters,p=1 |