Literature DB >> 23507457

European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing.

F Hannemann1, A Hartmann, J Schmitt, J Lützner, A Seidler, P Campbell, C P Delaunay, H Drexler, H B Ettema, E García-Cimbrelo, H Huberti, K Knahr, J Kunze, D J Langton, W Lauer, I Learmonth, C H Lohmann, M Morlock, M A Wimmer, L Zagra, K P Günther.   

Abstract

INTRODUCTION: There is an ongoing debate about the optimal use of metal-on-metal (MoM) bearings in total hip replacement, since there are uncertainties about local and systemic adverse effects due to wear and corrosion of these bearings. Despite various national recommendations, efforts to achieve international harmonization of specific evidence-based recommendations for best practice are still lacking. HYPOTHESIS: An international consensus study group should be able to develop recommendations on the use and monitoring of MoM bearings, preferably at the European level, through a multidisciplinary approach, by integrating the perspectives of various stakeholders.
MATERIALS AND METHODS: Twenty-one experts representing three stakeholder groups and eight countries participated in this European consensus study, which consisted of a consensus meeting, subsequent structured discussion, and consensus voting.
RESULTS: The current statement defines first of all benefits, local and systemic risks, as well as uncertain issues related to MoM bearings. Safety assessment after implantation of MoM comprises all patients. A closer follow-up is recommended for large head MoM (≥36mm) and resurfacing. In these implants basic follow-up should consist of x-rays and metal ion measurement of cobalt in whole blood, performed with GF-AAS or ICP-MS. Clinical and/or radiographic abnormality as well as elevated ion levels needs additional imaging (ultrasound, CT-scan and/or MARS-MRI). Cobalt values less than 2 μg/L are probably devoid of clinical concern, the threshold value for clinical concern is expected to be within the range of 2-7 μg/L. DISCUSSION: This is the first multinational, interdisciplinary, and multiprofessional approach for developing a recommendation for the use and monitoring of MoM bearings in total hip replacement. The current recommendations are in partial agreement with previous statements regarding the extent of follow-up and imaging techniques. They however differ from previous communications regarding measurement of metal ions and especially the investigated medium, technique, and eventual threshold levels. LEVEL OF EVIDENCE: Level V, expert opinion/agreement conference.
Copyright © 2013 Elsevier Masson SAS. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23507457     DOI: 10.1016/j.otsr.2013.01.005

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  41 in total

1.  Large-diameter metal-on-metal total hip arthroplasty: dislocation infrequent but survivorship poor.

Authors:  Adolph V Lombardi; Keith R Berend; Michael J Morris; Joanne B Adams; Michael A Sneller
Journal:  Clin Orthop Relat Res       Date:  2014-11-04       Impact factor: 4.176

2.  Cementless Total Hip Arthroplasty With Metasul Bearings Provides Good Results in Active Young Patients: A Concise Followup.

Authors:  Christian P Delaunay; Sophie Putman; Benjamin Puliéro; Matthieu Bégin; Henri Migaud; François Bonnomet
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

3.  In-situ Generated Tribomaterial in Metal/Metal Contacts: current understanding and future implications for implants.

Authors:  N Espallargas; A Fischer; A Igual Muñoz; S Mischler; M A Wimmer
Journal:  Biotribology (Oxf)       Date:  2017-05-12

4.  Pleomorphic clinical spectrum of metallosis in total hip arthroplasty.

Authors:  Francesco Pisanu; Carlo Doria; Matteo Andreozzi; Marco Bartoli; Laura Saderi; Giovanni Sotgiu; Paolo Tranquilli Leali
Journal:  Int Orthop       Date:  2018-09-29       Impact factor: 3.075

Review 5.  Revision total hip arthroplasty for metal-on-metal failure.

Authors:  Justin S Chang; Fares S Haddad
Journal:  J Clin Orthop Trauma       Date:  2019-10-04

6.  Long-term monitoring for nanomedicine implants and drugs.

Authors:  Michaela Kendall; Iseult Lynch
Journal:  Nat Nanotechnol       Date:  2016-03       Impact factor: 39.213

7.  [Complications of metal-on-metal tribological pairing].

Authors:  M Stiehler; F Zobel; F Hannemann; J Schmitt; J Lützner; S Kirschner; K-P Günther; A Hartmann
Journal:  Orthopade       Date:  2014-01       Impact factor: 1.087

8.  Outcomes after revision of metal on metal hip resurfacing to total arthroplasty using the direct anterior approach.

Authors:  Victoire Bouveau; Thomas-Xavier Haen; Joel Poupon; Christophe Nich
Journal:  Int Orthop       Date:  2018-03-08       Impact factor: 3.075

9.  Evaluation of impingement behaviour in lumbar spinal disc arthroplasty.

Authors:  Thomas M Grupp; James J Yue; Rolando Garcia; Christian Kaddick; Bernhard Fritz; Christoph Schilling; Jens Schwiesau; Wilhelm Blömer
Journal:  Eur Spine J       Date:  2014-06-06       Impact factor: 3.134

Review 10.  Management of metal-on-metal hip implant patients: Who, when and how to revise?

Authors:  Reshid Berber; John A Skinner; Alister J Hart
Journal:  World J Orthop       Date:  2016-05-18