Michael Nagler1, Lucas M Bachmann, Lorenzo Alberio, Anne Angelillo-Scherrer, Lars M Asmis, Wolfgang Korte, Adriana Mendez, Guido Reber, Hans Stricker, Dimitrios A Tsakiris, Walter A Wuillemin.
Abstract
BACKGROUND: While the assessment of analytical precision within medical laboratories has received much attention in scientific enquiry, the degree of as well as the sources causing variation between them remains incompletely understood. In this study, we quantified the variance components when performing coagulation tests with identical analytical platforms in different laboratories and computed intraclass correlations coefficients (ICC) for each coagulation test.Entities:
Year: 2013 PMID: 23496906 PMCID: PMC3599351 DOI: 10.1186/1477-9560-11-6
Source DB: PubMed Journal: Thromb J ISSN: 1477-9560
Participating laboratories with coagulometers, assays and reagents used
| A † | Instrumentation Laboratory ACL Top 700 | Clotting test (photometric) | Recombi-plastin® | Clotting test (photometric) | Hemosil® APPT SP | Clauss method | Hemosil® Fibrinogen-C | Kinetic photometric assay | Berichrom® FXIII |
| B | Siemens BCS-XP | Clotting test (photometric) | Innovin® | Clotting test (photometric) | Actin® | Clauss method | Multifibren U | Kinetic photometric assay | Berichrom® FXIII |
| C | Roche STA-R | Clotting test (viscosity) | Innovin® | Clotting test (viscosity) | Actin FS® | Clauss method | STA® Fibrinogen | Latex immunturbidimetry | HEXAMATE® FXIII |
| D | Sysmex CA7000 | Clotting test (photometric) | Innovin® | Clotting test (photometric) | Actin® | Clauss method | Dade® Thrombin | n.a. | n.a. |
| E | Siemens BCS-XP | Clotting test (photometric) | Innovin® | Clotting test (photometric) | Pathromtin® SL | Clauss method | Multifibren U | Kinetic photometric assay | Berichrom® FXIII |
| F | Siemens BCS | Clotting test (photometric) | Innovin® | Clotting test (photometric) | Actin FS® | Clauss method | Multifibren U | Kinetic photometric assay | Berichrom® FXIII |
| G* | Sysmex CA1500 | Clotting test (photometric) | Innovin® | Clotting test (photometric) | Actin FS® | Clauss method | Dade® Thrombin | n.a. | n.a. |
| H | Siemens BCS-XP | Clotting test (photometric) | Innovin® | Clotting test (photometric) | Pathromtin® SL | Clauss method | Multifibren U | Kinetic photometric assay | Berichrom® FXIII |
| I | Roche STA-R | Clotting test (viscosity) | Innovin® | Clotting test (viscosity) | STA® APPT | Clauss method | bovine thrombin, Diagnotec AG | Kinetic photometric assay | Berichrom® FXIII |
¶ Random order, not matching with the alphabetical list mentioned above. * Laboratory was excluded from analysis of variability between laboratories because not all tests were available. † Laboratory was excluded from analysis of variability between laboratories because of a unique platform. ‡ PT, Factors II, V, VII, X. § Factors VIII, IX, XI.
Intraclass correlations of various coagulation tests and intraobserver variability of fibrinogen tests with various platforms
| 1 | 2 | 4 | 2 | |
| 0.85 | 0.37 | 0.65 | 0.66 | |
| Within- laboratory variability ¶ | | 0.02 | 0.03 | 0.04 |
| Between- laboratory variability + | | 0.097 | 0.03 | 0.006 |
| | 0.19 | 0.24 | 0.80 | |
| | 0.44 | 0.04 | | |
| | 0.63 | 0.22 | | |
| | 0.71 | 0.82 | | |
| | 0.93 | 0.61 | | |
| | 0.86 | 0.83 | | |
| | 0.82 | 0.70 | | |
| | 0.90 | 0.65 | | |
| 0.57 | 0.81 |
* laboratory was excluded from analysis of between-laboratory variability because of a unique platform, ‡ analysis was limited to routine coagulation parameters, because one laboratory did not assess all tests, ¶ (Platform + Technician), + Laboratories use the platform in different ways (they interpret the manual differently).