Literature DB >> 23464311

Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors.

Anna Fredh1, Jonas Bengtsson Scherman, Lotte S Fog, Per Munck af Rosenschöld.   

Abstract

PURPOSE: The purpose of the present study was to investigate the ability of commercial patient quality assurance (QA) systems to detect linear accelerator-related errors.
METHODS: Four measuring systems (Delta(4®), OCTAVIUS(®), COMPASS, and Epiqa™) designed for patient specific quality assurance for rotational radiation therapy were compared by measuring four clinical rotational intensity modulated radiation therapy plans as well as plans with introduced intentional errors. The intentional errors included increasing the number of monitor units, widening of the MLC banks, and rotation of the collimator. The measurements were analyzed using the inherent gamma evaluation with 2% and 2 mm criteria and 3% and 3 mm criteria. When applicable, the plans with intentional errors were compared with the original plans both by 3D gamma evaluation and by inspecting the dose volume histograms produced by the systems.
RESULTS: There was considerable variation in the type of errors that the various systems detected; the failure rate for the plans with errors varied between 0% and 72%. When using 2% and 2 mm criteria and 95% as a pass rate the Delta(4®) detected 15 of 20 errors, OCTAVIUS(®) detected 8 of 20 errors, COMPASS detected 8 of 20 errors, and Epiqa™ detected 20 of 20 errors. It was also found that the calibration and measuring procedure could benefit from improvements for some of the patient QA systems.
CONCLUSIONS: The various systems can detect various errors and the sensitivity to the introduced errors depends on the plan. There was poor correlation between the gamma evaluation pass rates of the QA procedures and the deviations observed in the dose volume histograms.

Entities:  

Mesh:

Year:  2013        PMID: 23464311     DOI: 10.1118/1.4788645

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  29 in total

1.  The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT.

Authors:  J M Park; H-G Wu; J H Kim; J N K Carlson; K Kim
Journal:  Br J Radiol       Date:  2015-03-03       Impact factor: 3.039

Review 2.  Complexity metrics for IMRT and VMAT plans: a review of current literature and applications.

Authors:  Sophie Chiavassa; Igor Bessieres; Magali Edouard; Michel Mathot; Alexandra Moignier
Journal:  Br J Radiol       Date:  2019-07-24       Impact factor: 3.039

3.  On the feasibility of comprehensive high-resolution 3D remote dosimetry.

Authors:  Titania Juang; Ryan Grant; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

4.  Evaluation of the ability of three commercially available dosimeters to detect systematic delivery errors in step-and-shoot IMRT plans.

Authors:  Alison Gray; Omemh Bawazeer; Sankar Arumugam; Philip Vial; Joseph Descallar; David Thwaites; Lois Holloway
Journal:  Rep Pract Oncol Radiother       Date:  2021-09-30

5.  Optimizing the Region for Evaluation of Global Gamma Analysis for Nasopharyngeal Cancer (NPC) Pretreatment IMRT QA by COMPASS: A Retrospective Study.

Authors:  Wenli Lu; Ying Li; Wei Huang; Haixia Cui; Hanyin Zhang; Xin Yi
Journal:  Front Oncol       Date:  2022-06-14       Impact factor: 5.738

6.  Texture analysis on the edge-enhanced fluence of VMAT.

Authors:  So-Yeon Park; Jong Min Park; Wonmo Sung; Il Han Kim; Sung-Joon Ye
Journal:  Radiat Oncol       Date:  2015-04-01       Impact factor: 3.481

7.  Evaluation of 4-Hz log files and secondary Monte Carlo dose calculation as patient-specific quality assurance for VMAT prostate plans.

Authors:  Philipp Szeverinski; Matthias Kowatsch; Thomas Künzler; Marco Meinschad; Patrick Clemens; Alexander F DeVries
Journal:  J Appl Clin Med Phys       Date:  2021-06-20       Impact factor: 2.102

8.  Error sensitivity of a log file analysis tool compared with a helical diode array dosimeter for VMAT delivery quality assurance.

Authors:  Philipp Szeverinski; Matthias Kowatsch; Thomas Künzler; Marco Meinschad; Patrick Clemens; Alexander F DeVries
Journal:  J Appl Clin Med Phys       Date:  2020-10-23       Impact factor: 2.243

9.  IMRT credentialing for prospective trials using institutional virtual phantoms: results of a joint European Organization for the Research and Treatment of Cancer and Radiological Physics Center project.

Authors:  Damien C Weber; Veronique Vallet; Andrea Molineu; Christos Melidis; Vanda Teglas; Suzanne Naudy; Raphael Moeckli; David S Followill; Coen W Hurkmans
Journal:  Radiat Oncol       Date:  2014-05-29       Impact factor: 3.481

10.  The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT.

Authors:  Jung-In Kim; So-Yeon Park; Hak Jae Kim; Jin Ho Kim; Sung-Joon Ye; Jong Min Park
Journal:  Radiat Oncol       Date:  2014-07-28       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.