Ricky Hennessy1, Sam L Lim, Mia K Markey, James W Tunnell.
Abstract
We present a Monte Carlo lookup table (MCLUT)-based inverse model for extracting optical properties from tissue-simulating phantoms. This model is valid for close source-detector separation and highly absorbing tissues. The MCLUT is based entirely on Monte Carlo simulation, which was implemented using a graphics processing unit. We used tissue-simulating phantoms to determine the accuracy of the MCLUT inverse model. Our results show strong agreement between extracted and expected optical properties, with errors rate of 1.74% for extracted reduced scattering values, 0.74% for extracted absorption values, and 2.42% for extracted hemoglobin concentration values.Mesh:
Substances:
Year: 2013 PMID: 23455965 PMCID: PMC3584151 DOI: 10.1117/1.JBO.18.3.037003