Impact of the Advisa MRI pacing system on the diagnostic quality of cardiac MR images and contraction patterns of cardiac muscle during scans: Advisa MRI randomized clinical multicenter study results.

Juerg Schwitter1, Emanuel Kanal, Matthias Schmitt, Frederic Anselme, Timothy Albert, David L Hayes, David Bello, Attila Tóth, Yanping Chang, Dirk van Osch, Torsten Sommer.   

Abstract

BACKGROUND: The Advisa MRI system is designed to safely undergo magnetic resonance imaging (MRI). Its influence on image quality is not well known.
OBJECTIVE: To evaluate cardiac magnetic resonance (CMR) image quality and to characterize myocardial contraction patterns by using the Advisa MRI system.
METHODS: In this international trial with 35 participating centers, an Advisa MRI system was implanted in 263 patients. Of those, 177 were randomized to the MRI group and 150 underwent MRI scans at the 9-12-week visit. Left ventricular (LV) and right ventricular (RV) cine long-axis steady-state free precession MR images were graded for quality. Signal loss along the implantable pulse generator and leads was measured. The tagging CMR data quality was assessed as the percentage of trackable tagging points on complementary spatial modulation of magnetization acquisitions (n=16) and segmental circumferential fiber shortening was quantified.
RESULTS: Of all cine long-axis steady-state free precession acquisitions, 95% of LV and 98% of RV acquisitions were of diagnostic quality, with 84% and 93%, respectively, being of good or excellent quality. Tagging points were trackable from systole into early diastole (360-648 ms after the R-wave) in all segments. During RV pacing, tagging demonstrated a dyssynchronous contraction pattern, which was not observed in nonpaced (n = 4) and right atrial-paced (n = 8) patients.
CONCLUSIONS: In the Advisa MRI study, high-quality CMR images for the assessment of cardiac anatomy and function were obtained in most patients with an implantable pacing system. In addition, this study demonstrated the feasibility of acquiring tagging data to study the LV function during pacing.
Copyright © 2013 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23434621     DOI: 10.1016/j.hrthm.2013.02.019

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  11 in total

1.  Clinical value of cardiovascular magnetic resonance in patients with MR-conditional pacemakers.

Authors:  Claire E Raphael; Vassilis Vassiliou; Francisco Alpendurada; Sanjay K Prasad; Dudley J Pennell; Raad H Mohiaddin
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2015-11-20       Impact factor: 6.875

Review 2.  Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5 T: influence of pacemaker related artifacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement, short tau inversion recovery and T1-weighted imaging.

Authors:  Oliver Klein-Wiele; Marietta Garmer; Martin Busch; Serban Mateiescu; Rhyan Urbien; Gianluca Barbone; Kaffer Kara; Michael Schulte-Hermes; Frauke Metz; Birgit Hailer; Dietrich Grönemeyer
Journal:  Int J Cardiovasc Imaging       Date:  2016-11-04       Impact factor: 2.357

Review 3.  Current artefacts in cardiac and chest magnetic resonance imaging: tips and tricks.

Authors:  Khalid Alfudhili; Pier G Masci; Jean Delacoste; Jean-B Ledoux; Grégoire Berchier; Vincent Dunet; Salah D Qanadli; Juerg Schwitter; Catherine Beigelman-Aubry
Journal:  Br J Radiol       Date:  2016-03-17       Impact factor: 3.039

4.  Deep Learning to Automate Reference-Free Image Quality Assessment of Whole-Heart MR Images.

Authors:  Davide Piccini; Robin Demesmaeker; John Heerfordt; Jérôme Yerly; Lorenzo Di Sopra; Pier Giorgio Masci; Juerg Schwitter; Dimitri Van De Ville; Jonas Richiardi; Tobias Kober; Matthias Stuber
Journal:  Radiol Artif Intell       Date:  2020-05-27

5.  Feasibility and safety of adenosine cardiovascular magnetic resonance in patients with MR conditional pacemaker systems at 1.5 Tesla.

Authors:  Oliver Klein-Wiele; Marietta Garmer; Rhyan Urbien; Martin Busch; Kaffer Kara; Serban Mateiescu; Dietrich Grönemeyer; Michael Schulte-Hermes; Marc Garbrecht; Birgit Hailer
Journal:  J Cardiovasc Magn Reson       Date:  2015-12-22       Impact factor: 5.364

6.  Deactivation vs. asynchronous pacing - prospective evaluation of a protocol for rhythm management in patients with magnetic resonance conditional pacemakers undergoing adenosine stress cardiovascular magnetic resonance imaging.

Authors:  Oliver Klein-Wiele; Marietta Garmer; Gianluca Barbone; Rhyan Urbien; Martin Busch; Kaffer Kara; Harald Schäfer; Michael Schulte-Hermes; Birgit Hailer; Dietrich Grönemeyer
Journal:  BMC Cardiovasc Disord       Date:  2017-06-02       Impact factor: 2.298

7.  Clinical safety of the ProMRI implantable cardioverter-defibrillator systems during head and lower lumbar magnetic resonance imaging at 3 T: results of the ProMRI 3T ENHANCED Master study.

Authors:  Rainer Zbinden; Christian Wollmann; Johannes Brachmann; Jochen Michaelsen; Clemens Steinwender; Pramesh Kovoor; Sebastian Kelle; Andrew D McGavigan; Chi Keong Ching; Gemma A Figtree; Jan Schmidt; Tobias Timmel; Joachim Lotz
Journal:  Europace       Date:  2019-11-01       Impact factor: 5.214

8.  Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected.

Authors:  Anish N Bhuva; Thomas A Treibel; Andreas Seraphim; Paul Scully; Kristopher D Knott; João B Augusto; Camilla Torlasco; Katia Menacho; Clement Lau; Kush Patel; James C Moon; Peter Kellman; Charlotte H Manisty
Journal:  Front Cardiovasc Med       Date:  2021-01-29

9.  Safe performance of magnetic resonance of the heart in patients with magnetic resonance conditional pacemaker systems: the safety issue of the ESTIMATE study.

Authors:  Christian G Wollmann; Karin Thudt; Bernd Kaiser; Erich Salomonowitz; Harald Mayr; Sebastian Globits
Journal:  J Cardiovasc Magn Reson       Date:  2014-05-06       Impact factor: 5.364

10.  Clinical safety of ProMRI implantable cardioverter-defibrillator systems during head and lower lumbar magnetic resonance imaging at 1.5 Tesla.

Authors:  Wolfgang Rudolf Bauer; Dennis H Lau; Christian Wollmann; Andrew McGavigan; Jacques Mansourati; Theresa Reiter; Simone Frömer; Mark E Ladd; Harald H Quick
Journal:  Sci Rep       Date:  2019-12-03       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.