Radiation safety for the NOMAD portable X-ray system in a temporary morgue setting.

Kenneth P Hermsen1, Stanley S Jaeger, Mackenzie A Jaeger.   

Abstract

The purpose of this study was to determine the radiation levels resulting from leakage and scatter encountered by the forensic dental personnel using the Nomad at St. Gabriel, LA, following Hurricane Katrina. Using a Keithley Radiation Survey Meter and Lucite head phantom, radiation levels were measured at various distances and angles from the Nomad corresponding to the positions occupied by the dental personnel at St. Gabriel. The measurements were used to approximate the maximum total radiation dose from the Nomad to each team member for a 2- and a 4-week deployment. The results show that the maximum scatter radiation dose to any team member was 4.4 microR per X-ray or 0.253 millisieverts (mSv) for a 2-week deployment and 0.506 mSv for a 4-week deployment. Therefore, the leakage and scatter radiation dose from the Nomad was insignificant compared with established radiation safety guidelines of 50 mSv per year for all team members.

Mesh:

Year:  2008        PMID: 18489554     DOI: 10.1111/j.1556-4029.2008.00766.x

Source DB:  PubMed          Journal:  J Forensic Sci        ISSN: 0022-1198            Impact factor:   1.832


  3 in total

1.  The effects of device position on the operator's radiation dose when using a handheld portable X-ray device.

Authors:  Jimmy Makdissi; Ravikiran R Pawar; Ben Johnson; Bun S Chong
Journal:  Dentomaxillofac Radiol       Date:  2016-01-14       Impact factor: 2.419

2.  Evaluation of stray radiation to the operator for five hand-held dental X-ray devices.

Authors:  Richard Smith; Richard Tremblay; Graeme M Wardlaw
Journal:  Dentomaxillofac Radiol       Date:  2019-02-12       Impact factor: 2.419

3.  The reduction methods of operator's radiation dose for portable dental X-ray machines.

Authors:  Jeong-Yeon Cho; Won-Jeong Han
Journal:  Restor Dent Endod       Date:  2012-08-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.